Abstract
Purpose
The ValveClamp system is a novel edge-to-edge mitral valve repair system designed for the ease of operation. We report the outcomes of our initial experience of treating functional mitral regurgitation (MR) with the ValveClamp system.
Methods
The subjects of this study were patients with symptomatic functional MR despite standard medical therapy, who were treated with transapical ValveClamp implantation. The patients were divided into an atrial functional mitral regurgitation (AFMR) group and a ventricular functional mitral regurgitation (VFMR) group. Clinical and echocardiographic outcomes were evaluated at baseline and then at the 3-month follow up.
Results
Twelve patients, with a median age of 71 years (range 65–78 years), were assigned to the AFMR group (n = 5) or the VFMR group (n = 7). The device implantation rate was 100%, and 10 (83.3%) patients required implantation of only one clamp. The catheter time was less than 10 min in half of the patients, the fastest time being 5 min. There were no procedure-related complications. At the 3-month follow up, all patients were free from all-cause mortality, surgery, and rehospitalization. MR improved to ≤ 2 + in all 12 patients with MR grade 3 + or 4 + at baseline, (100%) and to ≤ 1 + in 9 of these patients (75%), with a low-pressure gradient. The left atrial diameter and the left ventricular end diastolic diameter decreased significantly in both the AFMR and VFMR groups. The left ventricular eject fraction at the 3-month follow up showed a rising trend in both the AFMR and VFMR groups, whereas PASP decreased remarkably. All 12 patients with baseline NYHA functional class III/IV (100%) showed improvement of at least 1 class, and 2 of these patients (16.7%) showed improvement of at least 2 classes.
Conclusions
The ValveClamp system is simple and effective for transapical transcatheter edge to edge repair in patients with functional MR.
Similar content being viewed by others
References
Dziadzko V, Clavel MA, Dziadzko M, Medina-Inojosa JR, Michelena H, Maalouf J, et al. Outcome and undertreatment of mitral regurgitation: a community cohort study. Lancet. 2018;391(10124):960–9.
Dziadzko V, Dziadzko M, Medina-Inojosa JR, Benfari G, Michelena HI, Crestanello JA, et al. Causes and mechanisms of isolated mitral regurgitation in the community: clinical context and outcome. Eur Heart J. 2019;40(27):2194–202.
Otto CM, Nishimura RA, Bonow RO, Carabello BA, Erwin JP, Gentile F, et al. 2020 ACC/AHA guideline for the management of patients with valvular heart disease: executive summary: a report of the American College of Cardiology/American Heart Association Joint Committee on clinical practice guidelines. J Am Coll Cardiol. 2021;77(4):450–500.
Vahanian A, Beyersdorf F, Praz F, Milojevic M, Baldus S, Bauersachs J, et al. 2021 ESC/EACTS Guidelines for the management of valvular heart disease. Eur Heart J. 2021;62:209.
Carabello BA. MitraClip and tertiary mitral regurgitation-mitral regurgitation gets curiouser and curiouser. JAMA Cardiol. 2019;4(4):307–8.
Sorajja P, Vemulapalli S, Feldman T, Mack M, Holmes DR, Stebbins A, et al. Outcomes with transcatheter mitral valve repair in the united states: an STS/ACC TVT registry report. J Am Coll Cardiol. 2017;70(19):2315–27.
Marquetand C, Reil JC, Saad M, Langer HF, Patzelt J. Treatment of recurrent MR by MitraClip XTR in a patient with a PASCAL device. JACC Cardiovasc Interv. 2019;12(24):e219–21.
Raghunathan D, Garcia-Sayan E, Dhoble A, Smalling R. First-in-human report of MitraClip G4 implantation for severe degenerative mitral regurgitation. Catheter Cardiovasc Interv. 2020;96(3):E395–7.
Praz F, Braun D, Unterhuber M, Spirito A, Orban M, Brugger N, et al. Edge-to-edge mitral valve repair with extended clip arms: early experience from a multicenter observational study. JACC Cardiovasc Interv. 2019;12(14):1356–65.
Pan W, Pan C, Jilaihawi H, Wei L, Tang Y, Zhou D, et al. A novel user-friendly transcatheter edge-to-edge mitral valve repair device in a porcine model. Catheter Cardiovasc Interv. 2019;93(7):1354–60.
Long Y, Pan W, Pan C, Li W, Zhou D, Ge J. Real-time echocardiographic guidance for transcatheter repair noncentral mitral regurgitation using the novel ValveClamp system. J Card Surg. 2021;36(11):4378–80.
Ge Z, Pan W, Zhou D, Li W, Wei L, Liu X, et al. Effect of a novel transcatheter edge-to-edge repair device on the three-dimensional geometry of mitral valve in degenerative mitral regurgitation. Catheter Cardiovasc Interv. 2021;97(1):177–85.
Ge Z, Pan C, Li W, Zhou D, Pan W, Wei L, et al. Real-time monitoring and step-by-step guidance for transapical mitral valve edge-to-edge repair using transesophageal echocardiography. J Interv Cardiol. 2021;2021:6659261.
Pan W, Zhou D, Wu Y, Guo Y, Pan X, Pan C, et al. First-in-human results of a novel user-friendly transcatheter edge-to-edge mitral valve repair device. JACC Cardiovasc Interv. 2019;12(23):2441–3.
Long Y, Pan W, Pan C, Li W, Wei L, Wu Y, et al. Transcatheter mitral valve edge-to-edge repair with a novel system. Int Heart J. 2022;63(1):23–9.
Asch FM, Grayburn PA, Siegel RJ, Kar S, Lim DS, Zaroff JG, et al. Echocardiographic outcomes after transcatheter leaflet approximation in patients with secondary mitral regurgitation: the COAPT trial. J Am Coll Cardiol. 2019;74(24):2969–79.
Stone GW, Vahanian AS, Adams DH, Abraham WT, Borer JS, Bax JJ, et al. Clinical trial design principles and endpoint definitions for transcatheter mitral valve repair and replacement: part 1: clinical trial design principles: a consensus document from the mitral valve academic research consortium. J Am Coll Cardiol. 2015;66(3):278–307.
Garcia-Sayan E, Raghunathan D, Li FM, Dhoble A, Sheu RD, Jelacic S, et al. Initial experience with the fourth generation MitraClip™: outcomes, procedural aspects, and considerations for device selection. Catheter Cardiovasc Interv. 2021;98(4):E626–36.
Lim DS, Kar S, Spargias K, Kipperman RM, O’Neill WW, Ng MKC, et al. Transcatheter valve repair for patients with mitral regurgitation: 30-day results of the CLASP study. JACC Cardiovasc Interv. 2019;12(14):1369–78.
Chakravarty T, Makar M, Patel D, Oakley L, Yoon SH, Stegic J, et al. Transcatheter edge-to-edge mitral valve repair with the MitraClip G4 system. JACC Cardiovasc Interv. 2020;13(20):2402–14.
Kaur S, Sadana D, Patel J, Gad M, Sankaramangalam K, Krishnaswamy A, et al. Atrial fibrillation and transcatheter repair of functional mitral regurgitation: evidence from a meta-regression. JACC Cardiovasc Interv. 2020;13(20):2374–84.
Godino C, Sisinni A, Pivato CA, Adamo M, Taramasso M, Parlati A. Prognostic value of pre-operative atrial fibrillation in patients with secondary mitral regurgitation undergoing MitraClip implantation. Am J Cardiol. 2021;143:51–9.
Gertz ZM, Herrmann HC, Lim DS, Kar S, Kapadia SR, Reed GW, et al. Implications of atrial fibrillation on the mechanisms of mitral regurgitation and response to MitraClip in the COAPT trial. Circ Cardiovasc Interv. 2021;14(4): e010300.
Yoshida J, Ikenaga H, Nagaura T, Utsunomiya H, Kawai M, Makar M, et al. Impact of percutaneous edge-to-edge repair in patients with atrial functional mitral regurgitation. Circ J. 2021;85(7):1001–10.
Funding
This study was supported by National Key R&D Program of China (2020YFC2008100).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
Dr. Zhou and Dr. Pan are consultants for Hanyu Medical Technology. The other authors have no conflicts of interest to declare.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
Supplementary file1 (MP4 7060 KB)
Supplementary file2 (MOV 14268 KB)
Supplementary file3 (MOV 13242 KB)
Supplementary file4 (MOV 11130 KB)
Supplementary file5 (MOV 14510 KB)
Supplementary file6 (MOV 8547 KB)
Supplementary file7 (MOV 3705 KB)
Rights and permissions
Springer Nature or its licensor holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Li, W., Long, Y., Pan, W. et al. Transcatheter edge to edge repair using the ease-of-use valve clamp system for functional mitral regurgitation: a primary report. Surg Today 53, 90–97 (2023). https://doi.org/10.1007/s00595-022-02559-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00595-022-02559-8