False profile view is independently associated with serum metal levels in patients with metal-on-metal hip arthroplasty

  • Original Article
  • Published:
European Journal of Orthopaedic Surgery & Traumatology Aims and scope Submit manuscript

Abstract

Purpose

Elevated levels of serum metal ions can be found in some patients with metal-on-metal (MoM) hip replacements. This study seeks to identify whether there is a significant association between the contact patch to rim distance (CPRD) and the anterior center edge angle (ACEA), respectively, and serum cobalt (Co) and chromium (Cr) levels in patients treated with unilateral MoM hip replacements by using standing anteroposterior and false profile view radiographs.

Methods

This is a retrospective analysis on 53 patients with either unilateral ReCap or M2a-Magnum MoM hip replacements operated in 2006 or 2007. Univariate linear regression and multivariable linear regression (MLR) analyses were performed using the CPRD and ACEA along with risk factors for elevated serum levels of Co and Cr previously suggested in the literature as explanatory variables. Serum Co and Cr were measured using inductively coupled plasma mass spectrometry.

Results

The MLR model showed that the same three variables (gender, CPRD and ACEA) were significantly associated with serum levels of both Co and Cr explaining approximately half of the variation. A univariate analysis showed a polynomial association between both anteversion angle and the ACEA with serum levels of Co and Cr. The vertex of the polynomial function was located at approximately 20° and 40°, respectively.

Conclusion

Gender, CPRD and ACEA are independently associated with serum levels of Co and Cr. Both ACEA and anteversion angles have optimae associated with low serum metal levels which may be useful for post-surgery evaluation of cup positioning.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Shetty V, Shitole B, Shetty G, Thakur H, Bhandari M (2011) Optimal bearing surfaces for total hip replacement in the young patient: a meta-analysis. Int Orthop 35:1281–1287

    Article  PubMed  Google Scholar 

  2. López-López JA, Humphriss RL, Beswick AD et al (2017) Choice of implant combinations in total hip replacement: systematic review and network meta-analysis. BMJ 359:j4651

    Article  PubMed  PubMed Central  Google Scholar 

  3. Smith AJ, Dieppe P, Howard PW, Blom AW (2012) Failure rates of metal-on-metal hip resurfacings: analysis of data from the national joint registry for England and Wales. Lancet 380:1759–1766

    Article  PubMed  Google Scholar 

  4. Al-Hajjar M, Fisher J, Williams S, Tipper JL, Jennings LM (2013) Effect of femoral head size on the wear of metal on metal bearings in total hip replacements under adverse edge-loading conditions. J Biomed Mater Res B Appl Biomater 101:213–222

    Article  PubMed  Google Scholar 

  5. Khan M, Kuiper J-H, Richardson JB (2008) The exercise-related rise in plasma cobalt levels after metal-on-metal hip resurfacing arthroplasty. J Bone Jt Surg Br 90:1152–1157

    Article  CAS  Google Scholar 

  6. Hart AJ, Skinner JA, Henckel J, SampsonGordon BF (2011) Insufficient acetabular version increases blood metal ion levels after metal-on-metal hip resurfacing. Clin Orthop Relat Res 469:2590–2597

    Article  PubMed  PubMed Central  Google Scholar 

  7. Angadji A, Royle M, Collins SN, Shelton JC (2009) Influence of cup orientation on the wear performance of metal-on-metal hip replacements. Proc Inst Mech Eng H 223:449–457

    Article  CAS  PubMed  Google Scholar 

  8. Underwood RJ, Zografos A, Sayles RS, Hart A, Cann P (2012) Edge loading in metal-on-metal hips: low clearance is a new risk factor. Proc Inst Mech Eng H 226:217–226

    Article  PubMed  PubMed Central  Google Scholar 

  9. Matthies AK, Henckel J, Cro S, Suarez A, Noble PC, Skinner J, Hart AJ (2014) Predicting wear and blood metal ion levels in metal-on-metal hip resurfacing. J Orthop Res 32:167–174

    Article  CAS  PubMed  Google Scholar 

  10. Le Duff MJ, Ebramzadeh E, Amstutz HC (2017) Contact patch to rim distance: the quintessential tool for metal-on-metal bearing in vivo performance analysis – a review. Hip Int 27:220–225

    Article  PubMed  Google Scholar 

  11. Yoon Ba JP, Le MJ, Ma D, Johnson Ba AJ, Takamura Ba KM, Ebramzadeh E, Amstutz HC (2013) Contact patch to rim distance predicts metal ion levels in hip resurfacing. Clin Orthop Relat Res 471:1615–1621

    Article  Google Scholar 

  12. Langton DJ, Sprowson AP, Joyce TJ, Reed M, Carluke I, Partington P, Nargol AVF (2009) Blood metal ion concentrations after hip resurfacing arthroplasty: a comparative study of articular surface replacement and Birmingham hip resurfacing arthroplasties. J Bone Jt Surg Br 91:1287–1295

    Article  CAS  Google Scholar 

  13. Haenle M, Mittelmeier W, Barbano R, Wörtler K, Scholz R, Bader R (2010) Accuracy and reliability of different methods to evaluate the acetabular cup version from plain radiographs. Surg Radiol Anat 32:725–730

    Article  PubMed  Google Scholar 

  14. Marx A, von Knoch M, Pförtner J, Wiese M, Saxler G (2006) Misinterpretation of cup anteversion in total hip arthroplasty using planar radiography. Arch Orthop Trauma Surg 126:487–492

    Article  PubMed  Google Scholar 

  15. Manjunath KS, Soruban V, Gopalakrishna KG (2015) Evaluation of radiological methods of assessing cup anteversion in total hip replacement. Eur J Orthop Surg Traumatol 25:1285–1292

    Article  CAS  PubMed  Google Scholar 

  16. de LequesneSeze MS (1961) False profile of the pelvis. a new radiographic incidence for the study of the hip. Its use in dysplasias and different coxopathies. Rev Rhum Mal Osteoartic 28:643–652

    Google Scholar 

  17. Lequesne M (2002) The false profile view of the hip: role, interest, economic considerations. Jt Bone Spine 69:109–113

    Article  Google Scholar 

  18. Zingg PO, Werner CML, Sukthankar A, Zanetti M, Seifert B, Dora C (2009) The anterior center edge angle in Lequesne’s false profile view: interrater correlation, dependence on pelvic tilt and correlation to anterior acetabular coverage in the sagital plane. A cadaver study. Arch Orthop Trauma Surg 129:787–791

    Article  CAS  PubMed  Google Scholar 

  19. Lee JKL, Sardana V, White C, Speirs A, Beaulé PE (2018) Cup version can be accurately measured on the false profile view radiograph. Hip Int 24:347–354

    Article  Google Scholar 

  20. Borgwardt A, Borgwardt L, Zerahn B, Daugaard H, Borgwardt L, Ribel-Madsen S (2018) A randomized 7-year study on performance of the stemmed metal M2a-Magnum and ceramic C2a-Taper, and the resurfacing ReCap hip implants. J Arthroplast 33:1412–1420

    Article  Google Scholar 

  21. Back DL, Young DA, Shimmin AJ (2005) How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing? Clin Orthop Relat Res 438:177–181

    Article  PubMed  Google Scholar 

  22. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR (1978) Dislocations after total hip-replacement arthroplasties. J Bone Joint Surg Am 60:217–220

    Article  CAS  PubMed  Google Scholar 

  23. Lu M, Zhou Y-X, Du H, Zhang J, Liu J (2013) Reliability and validity of measuring acetabular component orientation by plain anteroposterior radiographs. Clin Orthop Relat Res 471:2987–2994

    Article  PubMed  PubMed Central  Google Scholar 

  24. Nishino H, Nakamura S, Arai N, Matsushita T (2013) Accuracy and precision of version angle measurements of the acetabular component after total hip arthroplasty. J Arthroplasty 28:1644–1647

    Article  PubMed  Google Scholar 

  25. Grömping U (2006) Relative importance for linear regression in R: the Package relaimpo. J Stat Softw 17:1–27

    Article  Google Scholar 

  26. Langton DJ, Jameson SS, Joyce TJ, Webb J, Nargol AVF (2008) The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip. J Bone Jt Surg Br 90:1143–1151

    Article  CAS  Google Scholar 

  27. Fisher J (2011) Bioengineering reasons for the failure of metal-on-metal hip prostheses. J Bone Jt Surg Br 93:1001–1004

    Article  CAS  Google Scholar 

  28. Nevelos J, Ingham E, Doyle C, Streicher R, Nevelos A, Walter W, Fisher J, Eng D (2000) Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns. J Arthroplast 15:793–795

    Article  CAS  Google Scholar 

  29. Heisel C, Silva M, Skipor AK, Jacobs JJ, Schmalzried TP (2005) The relationship between activity and ions in patients with metal-on-metal bearing hip prostheses. J Bone Joint Surg Am 87:781–787

    Article  PubMed  Google Scholar 

  30. Daniel J, Ziaee H, Pradhan C, McMinn DJW (2009) Six-year results of a prospective study of metal ion levels in young patients with metal-on-metal hip resurfacings. J Bone Joint Surg Br 91:176–179

    Article  CAS  PubMed  Google Scholar 

Download references

Funding

This research received no specific grant from any funding agency in the public, commercial, or not-for-profit sectors.

Author information

Authors and Affiliations

Authors

Contributions

AB, BZ and SR-M designed the study. AB and SR-M performed the clinical controls and collected data. BML and BZ took the lead in writing the manuscript and performed data processing. BZ performed the statistical calculations. All authors provided critical feedback and helped shape the research, analysis and manuscript.

Corresponding author

Correspondence to Bo Madvig Larsen.

Ethics declarations

Conflicts of interest

The authors declare that they have no conflict of interest.

Ethical Approval

This study was approved by the local ethics committee as protocol KF 01–157/04.

Human or Animal Rights

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.

Consent to participate

Patients gave their written informed consent before participating and the study.

Consent for publication

Patients signed informed consent regarding publishing their data.

Data availability

The data that support the findings of this study are available from the corresponding author, upon reasonable request.

Code availability

Not applicable.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Larsen, B.M., Borgwardt, A., Ribel-Madsen, S. et al. False profile view is independently associated with serum metal levels in patients with metal-on-metal hip arthroplasty. Eur J Orthop Surg Traumatol 31, 1029–1036 (2021). https://doi.org/10.1007/s00590-020-02839-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00590-020-02839-4

Keywords

Navigation