A case example and literature review of catastrophic wear before catastrophic failure: identification of trunnionosis and metallosis in metal-on-polyethylene hip arthroplasty prior to frank failure or fracture

  • Up-to date Review and Case Report • HIP - ARTHROPLASTY
  • Published:
European Journal of Orthopaedic Surgery & Traumatology Aims and scope Submit manuscript

Abstract

Though rare, there are documented failures of femoral prosthesis due to corrosion of the head–neck interface in total hip arthroplasty (THA), a phenomenon known as trunnionosis. This wear can result in metallosis, whereby metal debris scatters the surrounding soft tissues. We present on a 58-year-old female who presented with increase in hip and back pain 10 years following right THA using a metal-on-polyethylene construct with a large femoral head (44 mm). Aspiration withdrew metallic fluid, and intraoperative findings showed corrosion of the head–neck taper with surrounding metallosis and pseudocapsule formation. Despite advances in THA design, corrosion and wear between components still exists and may be cause for failure. We present on both the subtle clinical findings and the recommended workup when suspicion is high for trunnionosis, metallosis, or wear, ideally with identification prior to catastrophic failure such as component dislocation or fracture as previously reported.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Banerjee S, Cherian JJ, Bono JV, Kurtz SM, Geesink R, Meneghini RM (2015) Gross trunnion failure after primary total hip arthroplasty. J Arthroplasty 30(4):641–648. https://doi.org/10.1016/j.arth.2014.11.023

    Article  PubMed  Google Scholar 

  2. Cooper HJ, Urban RM, Wixson RL, Meneghini RM, Jacobs JJ (2013) Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck. J Bone Joint Surg Am 95(10):865–872. https://doi.org/10.2106/JBJS.L.01042

    Article  PubMed  PubMed Central  Google Scholar 

  3. Pourzal R, Lundberg HJ, Hall DJ, Jacobs JJ (2018) What factors drive taper corrosion? J Arthroplasty 1:1–2. https://doi.org/10.1016/j.arth.2018.03.055

    Article  Google Scholar 

  4. Tan SC, Teeter MG, Del Balso C, Howard JL, Lanting BA (2015) Effect of taper design on trunnionosis in metal on polyethylene total hip arthroplasty. J Arthroplasty 30(7):1269–1272. https://doi.org/10.1016/j.arth.2015.02.031

    Article  PubMed  Google Scholar 

  5. Hussey DK, McGrory BJ (2017) Ten-year cross-sectional study of mechanically assisted crevice corrosion in 1352 consecutive patients with metal-on-polyethylene total hip arthroplasty. J Arthroplasty 32(8):2546–2551. https://doi.org/10.1016/j.arth.2017.03.020

    Article  PubMed  Google Scholar 

  6. Cooper HJ, Della Valle CJ, Berger RA, Tetreault M, Paprosky WG, Sporer SM (2012) Corrosion at the head-neck taper as a cause for adverse local tissue reactions after total hip arthroplasty. J Bone Joint Surg Am 94(18):1655–1661. https://doi.org/10.2106/JBJS.K.01352

    Article  PubMed  Google Scholar 

  7. Morlock M, Bunte D, Guhrs J, Bishop N (2017) Corrosion of the head-stem taper junction-are we on the verge of an epidemic? Review article. HSS J 13(1):42–49. https://doi.org/10.1007/s11420-016-9526-4

    Article  PubMed  Google Scholar 

  8. Morlock MM (2015) The taper disaster—how could it happen? Hip Int 25(4):339–346. https://doi.org/10.5301/hipint.5000269

    Article  PubMed  Google Scholar 

  9. Matsen Ko L, Chen AF, Deirmengian GK, Hozack WJ, Sharkey PF (2016) Catastrophic femoral head-stem trunnion dissociation secondary to corrosion. J Bone Joint Surg Am 98(16):1400–1404. https://doi.org/10.2106/JBJS.15.00914

    Article  PubMed  Google Scholar 

  10. Arvinte D, Kiran M, Sood M (2015) Femoral head-trunnion dissociation secondary to corrosion of a modular total hip replacement: a case report. JBJS Case Connect 5(2):e4–e35. https://doi.org/10.2106/jbjs.cc.n.00146

    Article  Google Scholar 

  11. Mariconda M, Galasso O, Costa GG, Recano P, Cerbasi S (2011) Quality of life and functionality after total hip arthroplasty: a long-term follow-up study. BMC Musculoskelet Disord 12:222. https://doi.org/10.1186/1471-2474-12-222

    Article  PubMed  PubMed Central  Google Scholar 

  12. Svensson O, Mathiesen EB, Reinholt FP, Blomgren G (1988) Formation of a fulminant soft-tissue pseudotumor after uncemented hip arthroplasty. A case report. J Bone Joint Surg Am 70(8):1238–1242

    Article  CAS  PubMed  Google Scholar 

  13. Goldberg JR, Gilbert JL (2003) In vitro corrosion testing of modular hip tapers. J Biomed Mater Res B Appl Biomater 64(2):78–93. https://doi.org/10.1002/jbm.b.10526

    Article  CAS  PubMed  Google Scholar 

  14. Walker P, Campbell D, Della Torre P, Brazil D, McTighe T (2016) Trunnion Corrosion and early failure in monolithic metal-on-polyethylene TMZF femoral components: a case series. Reconstr Rev 6(3):13–18. https://doi.org/10.15438/rr.6.3.161

    Article  Google Scholar 

  15. Fokter SK, Molicnik A, Kavalar R, Pelicon P, Rudolf R, Gubeljak N (2017) Why do some titanium-alloy total hip arthroplasty modular necks fail? J Mech Behav Biomed Mater 69:107–114. https://doi.org/10.1016/j.jmbbm.2016.12.012

    Article  CAS  PubMed  Google Scholar 

  16. Fokter SK, Repse-Fokter A, Takac I (2009) Case report: femoral neuropathy secondary to total hip arthroplasty wear debris. Clin Orthop Relat Res 467(11):3032–3035. https://doi.org/10.1007/s11999-009-0894-7

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

We acknowledge Dr. Ian Martin, DO, for his assistance in the presentation and description of our pathology results and provided photomicrograph. We acknowledge Dr. Wesley Stroud, MD, for the illustrative intraoperative images and medical record review.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Allison M. Hunter.

Ethics declarations

Conflict of interest

The authors whose names are listed immediately below certify that they have no affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers’ bureaus; membership, employment, consultancies, stock ownership, or other equity interest; and expert testimony or patent-licensing arrangements) or non-financial interest (such as personal or professional relationships, affiliations, knowledge, or beliefs) in the subject matter or materials discussed in this manuscript.

Statement of informed consent

Each author certifies that his or her institution approved the reporting of this case report, that all investigations were conducted in conformity with ethical principles of research, and that informed consent for participation in the study was obtained.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hunter, A.M., Hsu, A., Shah, A. et al. A case example and literature review of catastrophic wear before catastrophic failure: identification of trunnionosis and metallosis in metal-on-polyethylene hip arthroplasty prior to frank failure or fracture. Eur J Orthop Surg Traumatol 29, 711–715 (2019). https://doi.org/10.1007/s00590-018-2333-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00590-018-2333-5

Keywords

Navigation