Experimental effect of injectable calcium phosphate cement on osteoporotic distal radius: resistance in a fall provoking a Colles’ fracture

  • C. Noirfalisse
  • G. Poumarat
  • P. LiverneauxEmail author
Original Article
  • 79 Downloads

Abstract

Stating background

The purpose of this work was to determine a possible change, after an injection of calcium phosphate cement, in the mechanical resistance of the distal radius in fractures often seen in osteoporotic patients.

Methods

We reproduced a Colles’ fracture on two anatomical subjects. Only one wrist per cadaver had been injected of calcium phosphate cement Cementek LV®. In this way, we compared the mechanical strength of a wrist injected and a wrist not injected.

Results

This study shows that an injection of calcium phosphate cement to replace the trabecular bone in the radius is mechanically interesting in preventing fractures in osteoporotic patients if the cement completely fills the bone cavity.

Conclusions

The confined study permits us to show that the cement has mechanical resistance in compression much better than those found in the trabecular bone.

Keywords

Bone substitute Mechanical resistance Osteoporosis Radius Calcium phosphate cement 

Effet expérimental de l’injection d’un ciment phosphocalcique dans un radius distal ostéoporotique: résistance à une chute provoquant une fracture de Colles

Résumé

Introduction

Le but de ce travail était de déterminer une possible modification, après injection de ciment phosphocalcique, de la résistance mécanique du radius distal lors des fractures habituellement rencontrées chez les patients ostéoporotiques.

Méthodes

Nous avons reproduit une fracture de Colles sur deux sujets anatomiques. Un seul poignet par cadavre a été injecté par le ciment phosphocalcique Cementek LV®. De cette façon, nous avons comparé la résistance mécanique d’un poignet injecté et celle d’un poignet non injecté.

Résultats

Cette étude montre qu’une injection de ciment phosphocalcique pour remplacer l’os trabéculaire du radius distal est mécaniquement capable d’empêcher les fractures chez les patients ostéoporotiques, à condition que le ciment remplisse complètement la cavité osseuse.

Conclusion

Cette étude certes limitée nous permet de montrer que le ciment possède une résistance mécanique à la compression meilleure que celle de l’os trabéculaire.

Mots clés

Substitut osseux Résistance mécanique Ostéoporose Radius Ciment phosphocalcique 

References

  1. 1.
    Allieu Y (1998) Mécanismes et classifications des fractures du radius distal. In: Allieu Y (ed) Fractures du radius distal de l’adulte. Cahiers d'enseignement de la SOFCOT. Expansion scientifique, Paris, pp 28–48Google Scholar
  2. 2.
    Bonnel F (1993) Os cortical et os trabéculaire: un point de vue mécanique finaliste (de l’anatomie microscopique à la microscopie). Masson, Paris, pp 1–13Google Scholar
  3. 3.
    Breart G, Cormier C, Le Gales C, De Vernejoule MC, Dupuy C (2000) L’ostéoporose, ou comment prévenir une maladie du vieillissement—Fondation pour la recherche médicale. http://www.frm.org
  4. 4.
    Castaing J et Le Club des Dix (1964) Les fractures récentes de l’extrémité inférieure du radius chez l’adulte. Rev Chir Orthop 50:581–696Google Scholar
  5. 5.
    Consensus Development Conference (1993) Diagnosis, prophylaxis and treatment of osteoporosis. Am J Med 94:646–650CrossRefGoogle Scholar
  6. 6.
    Cortet B, Cotten A, Boutry N, Flipo RM, Duquesnoy B, Chastanet P, Delcambre B (1999) Percutaneous vertebroplasty in the treatment of osteoporotic vertebral compression fractures: an open prospective study. J Rheumatol 26:2222–2228PubMedGoogle Scholar
  7. 7.
    Hillmeier J, Grafe I, Da Fonseca K, Meeder PJ, Noldge G, Libicher M, Kock HJ, Haag M, Kasperk C (2004) The evaluation of balloon kyphoplasty for osteoporotic vertebral fractures: an interdisciplinary concept. Orthopade 33:893–904PubMedGoogle Scholar
  8. 8.
    Kapandji IA (2001) Physiologie articulaire du membre supérieur cinquième édition. Maloine, Paris, pp 138–170Google Scholar
  9. 9.
    Lewis RM (1950) Colles’s fracture: causative mechanism. Surgery 27:427–435Google Scholar
  10. 10.
    Liverneaux P (2004) Experimental increase in resistance of the osteoporotic distal radius with calcium phosphate cement. Chir Main 23:37–44PubMedCrossRefGoogle Scholar
  11. 11.
    Matsuyama Y, Goto M, Yoshihara H, Tsuji T, Sakai Y, Nakamura H, Sato K, Kamiya M, Ishiguro N (2004) Vertebral reconstruction with biodegradable calcium phosphate cement in the treatment of osteoporotic vertebral compression fracture using instrumentation. J Spinal Disord Tech 17:291–296PubMedCrossRefGoogle Scholar
  12. 12.
    Mehbod A, Aunoble S, Le Huec JC (2003) Vertebroplasty for osteoporotic spine fracture: prevention and treatment. Eur Spine J 12:S155–S162PubMedCrossRefGoogle Scholar
  13. 13.
    Pistoia W, Van Rietbergen B, Lochmuller EM, Lill CA, Eckstein F, Ruegsegger P (2002) Estimation of distal radius failure load with micro-finite element analysis models based on three-dimensional peripheral quantitative computed tomography images. Bone 30:842–848PubMedCrossRefGoogle Scholar
  14. 14.
    Tanaka A, Ishii Y (2002) Mechanical properties of the femur after injection of calcium phosphate cement containing arbekacin sulfate and polylactic acid in a rat model of experimental osteoporosis. J Orthop Sci 7:370–378PubMedCrossRefGoogle Scholar

Copyright information

© Springer-Verlag 2006

Authors and Affiliations

  1. 1.Laboratoire de Biomécanique AnatomieERIM EA 3295, Faculté de MédecineClermont-Ferrand CedexFrance
  2. 2.SOS main Strasbourg-Hôpitaux-UniversitairesIllkirchFrance
  3. 3.Hand Surgery DepartmentStrasbourg University HospitalsIllkirchFrance

Personalised recommendations