The effectiveness of the proximal femoral stem centraliser on the stem alignment in total hip replacement

  • Almunir YousefEmail author
  • Angus Wallace
Original Article
  • 93 Downloads

Abstract

We studied a new polymethylmethacrylate proximal femoral centraliser designed for use along with the distal femoral centraliser. Using cadaver femora we conducted an in vitro study to measure the influence of the proximal centraliser on the cement mantle and femoral stem alignment in a total hip replacement (THR). The cadaver femora were randomised into two groups. The Royal National Orthopaedic Hospital (RNOH) femoral stems were inserted to each femur using third-generation cementing technique .The first group was inserted using proximal and distal centralisers. The second group was inserted using distal centralisers only. All the hip specimens were X-rayed using high-tech scanner and computer software. The stem alignment and cement mantle were measured. We concluded that the proximal femoral centraliser improved the cement mantle distribution and femoral stem alignment in THR. This improvement will reflect on the long-term survival of the THR. We also showed an excellent and accurate way of measuring the femoral stem alignment in THR.

Keywords

Total hip replacement Femoral stem Cementing Centraliser 

Efficience du centreur fémoral proximal de la tige des PTH

Résumé

Nous avons étudié un nouveau centreur proximal en polyméthylméthacrilate (PMMA) prévu pour être utilisé avec le centralisateur distal. Grâce à des fémurs de cadavres, nous avons étudié in vitro l’influence du centreur proximal sur le manteau de ciment et sur le positionnement de la tige fémorale des prothèses totales de hanche (PTH). Les fémurs de cadavres ont été randomisés en deux groupes. Des tiges de type « Royal National Orthopaedic Hospital » (RNOH) ont été implantés dans chaque fémur en suivant la technique de cimentage de troisième génération : le premier groupe avec les deux centreurs, proximal et distal ; le second groupe en n’utilisant que le seul centreur distal. Toutes les hanches ont été radiographiées avec un scanner de haute technologie et une gestion par ordinateur. L’alignement des tiges et la qualité du manteau de ciment ont été mesurés. Nous avons pu conclure que l’utilisation du centreur proximal améliore le positionnement de la tige prothétique ainsi que la qualité du manteau de ciment dans les PTH. Cette amélioration se répercutera sur la durée de vie de la PTH. Nous avons également démontré une excellente et fiable façon de mesurer le positionnement de la tige fémorale des PTH.

Mots clés

Prothèse totale de hanche Tige fémorale Cementage Centreur 

References

  1. 1.
    Berger RA, Steel MJ, Wood K, Evans R, D Antonino J, Rubash HE (1997) Effect of centralizing devices on cement mantle deficiencies and initial prosthetic alignment in total hip arthroplasty. J Arthroplasty 12(4):434–443CrossRefPubMedGoogle Scholar
  2. 2.
    Charles Gozzand, Subine Gheduzzi, Anthony W.Miles, Ian D. Learmonth (2003) An in-vitro investigation of the CPS-Plus femoral stem. Influence of the centralizer on cement pressurization during stem insertion. Acta Ortho Scand 74(2):154–158CrossRefGoogle Scholar
  3. 3.
    Charnley J (1970) Acrylic cement in orthopedic surgery. Longman group, EdinburghGoogle Scholar
  4. 4.
    Crowninshield RD, Tolbert JR (1983) Cement strain measurement surrounding loose and well fixed femoral component stem. J Biomed Master Res 17:819CrossRefGoogle Scholar
  5. 5.
    Edward Ebramzadeh, Augsto sarmiento, Harry A, McKellop, Adolfo llinas, William GoGan (1994) The cement mantle in total hip arthroplasty. JBJS 76A:77–87Google Scholar
  6. 6.
    Goldburg BA, Al Habbal G, Noble PC, Paravic, Liebs TR, Tullo HS (1998) Proximal and distal femoral centralizers in modern cemented hip arthroplasty. Clin Orthop 349:163–173PubMedCrossRefGoogle Scholar
  7. 7.
    Huddlestone HD (1988) Femoral lysis after cemented hip arthroplasty. J Arthroplasty 3:285–297PubMedCrossRefGoogle Scholar
  8. 8.
    Huiskes R (1980) Some fundamental aspects of human joint replacement, analysis of stress and heat conduction in bone prosthesis structure. Acta Ortho Scand suppl 185:109–200Google Scholar
  9. 9.
    James E, Dowd, Charles W. Cha, Sunil Trakru, Sin-Yoon Kim, Ick- Hawan Yang, Harry E, Rubash (1998) Failure of total hip arthroplasty with precoated prosthesis, Clin Ortho Relat Res 355:123–136CrossRefGoogle Scholar
  10. 10.
    Kadakia NR, Noble PC, Beardsely CL et al (1997) Do cement voids cause premature failure of cement mantles in total hip replacement. Orth Tras 21:106–299Google Scholar
  11. 11.
    Malony WJ, Jasty M, Burke DW, O Connor DO, Zalauski EB, Bragdon C, Harris WH (1989) Biomechanical and histological investigation of cemented total hip arthroplasties, a study of autopsy retrieved femora after in vivo cycling. Clin Ortho 249:129–140Google Scholar
  12. 12.
    Noble PC, Collier MB, Maltry JA, Kamaric E, Tullos HS (1998) Pressurization and centralization enhance the quality and reproducibility of cement mantle. Clin Orth 355:77–89CrossRefGoogle Scholar
  13. 13.
    Noble PC, Tulles HS, London G (1991) The optimum cement mantle for total hip replacement theory and practice. In instructional course lectures. Am Acad Orthop Surg 40:145–150Google Scholar
  14. 14.
    Ramaniraka NA, Rakotomanana LR, Leyvraz PF (2000) The fixation of cemented femoral component effect of stem stiffness, cement thickness and roughness of the cement-bone surface. J Bone Joint Surg (Br) 82:297–303CrossRefGoogle Scholar

Copyright information

© Springer-Verlag 2005

Authors and Affiliations

  1. 1.Orthopaedic DepartmentKing’s Mill HospitalNottsUK
  2. 2.Orthopaedic DepartmentQueen’s Medical CentreNottinghamUK

Personalised recommendations