Loads on internal spinal fixators measured in different body positions

Abstract

Telemeterized internal spinal fixation devices were implanted in ten patients. The loads acting on the fixators were compared for different body positions, including standing, sitting, and lying in a supine, prone, and lateral position. Implant loads differed considerably from patient to patient depending, for example, on the indication for surgery and the surgical procedure. They were altered by anterior interbody fusion. Mostly, only small differences in implant loads were found for the various lying positions. Flexion bending moments were significantly higher in upright than in lying body positions. Loads on the fixators were not higher for sitting than for standing. Patients who have undergone mono- or bisegmental spine stabilization should therefore be allowed to sit as soon as they can leave the bed.

This is a preview of subscription content, log in to check access.

Author information

Affiliations

Authors

Additional information

Received: 16 July 1999 Accepted: 20 July 1999

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Rohlmann, A., Bergmann, G. & Graichen, F. Loads on internal spinal fixators measured in different body positions. E Spine J 8, 354–359 (1999). https://doi.org/10.1007/s005860050187

Download citation

  • Key words Internal spinal fixator
  • Load measurement
  • Body position
  • Telemetry
  • Spine