Skip to main content

Advertisement

Log in

Metal artifact reduction using virtual monochromatic images for patients with pedicle screws implants on CT

  • Original Article
  • Published:
European Spine Journal Aims and scope Submit manuscript

Abstract

Purpose

To evaluate the effect of spectral CT for metal artifact reduction in patients with pedicle screw.

Methods

45 patients with 119 pairs of pedicle screws underwent spectral CT examination. One set of conventional (140 kVp) polychromatic image and nine sets of virtual monochromatic images (60–140 keV) were obtained. The standard deviation (SD) of CT number in 12 locations around the implant and 1 on homogenous fat was measured to generate artifact index (AI). Objective assessment including AI, CT number and SD value was performed with independent t test and paired sample t test. Two radiologists independently reviewed the image quality, and Wilcoxon signed-rank test and kappa analysis were used for the subjective scores of image quality.

Results

The artifact index (AI) of all the regions decreased as keV increased. AIs of 100–140 keV were lower than that of 140 kVp images. At 120 keV there was no significant difference in CT numbers of psoas major muscle and vertebral canal between pedicle screw level and pedicle level, but a significant difference in SD value was determined between the two levels. The subjective scores at 100–140 keV were higher than the images at 140 kVp, and the highest subjective score of two observers and excellent interobserver agreement were found at 120 keV (κ = 0.889).

Conclusions

Virtual monochromatic images at high-energy levels have a well-concordant effect of removing metal artifacts, and 120 keV monochromatic images provided an accurate CT number and good subjective score.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

Abbreviations

AI:

Artifact index

GSI:

Gemstone spectral imaging

MARs:

Metal artifacts reduction software

MSCT:

Multi-slice computed tomography

ROI:

Region of interest

SD:

Standard deviation

References

  1. Tian NF, Huang QS, Zhou P, Zhou Y, Wu RK, Lou Y et al (2011) Pedicle screw insertion accuracy with different assisted methods: a systematic review and meta-analysis of comparative studies. Eur Spine J 20:846–859

    Article  PubMed  PubMed Central  Google Scholar 

  2. Amato V, Giannachi L, Irace C, Corona C (2010) Accuracy of pedicle screw placement in the lumbosacral spine using conventional technique: computed tomography postoperative assessment in 102 consecutive patients. J Neurosurg Spine 12:306–313

    Article  PubMed  Google Scholar 

  3. Sarlak AY, Tosun B, Atmaca H, Sarisoy HT, Buluç L (2009) Evaluation of thoracic pedicle screw placement in adolescent idiopathic scoliosis. Eur Spine J 18:1892–1897

    Article  PubMed  PubMed Central  Google Scholar 

  4. Amiot LP, Lang K, Putzier M, Zippel H, Labelle H (2000) Comparative results between conventional and computer-assisted pedicle screw installation in the thoracic, lumbar, and sacral spine. Spine 25:606–614

    Article  CAS  PubMed  Google Scholar 

  5. Murtagh RD, Quencer RM, Castellvi AE, Yue JJ (2011) New techniques in lumbar spinal instrumentation: what the radiologist needs to know. Radiology 260:317–330

    Article  PubMed  Google Scholar 

  6. Douglas-Akinwande AC, Buckwalter KA, Rydberg J, Rankin JL, Choplin RH (2006) Multichannel CT: evaluating the spine in postoperative patients with orthopedic hardware. Radiographics 26:S97–S110

    Article  PubMed  Google Scholar 

  7. Lee YH, Park KK, Song HT, Kim S, Suh JS (2012) Metal artefact reduction in gemstone spectral imaging dual-energy CT with and without metal artefact reduction software. Eur Radiol 22:1331–1340

    Article  PubMed  Google Scholar 

  8. Wang Y, Qian B, Li B, Qin G, Zhou Z, Qiu Y et al (2013) Metal artifacts reduction using monochromatic images from spectral CT: evaluation of pedicle screws in patients with scoliosis. Eur J Radiol 82:360–366

    Article  Google Scholar 

  9. Srinivasan A, Hoeffner E, Ibrahim M, Shah GV, LaMarca F, Mukherji SK (2013) Utility of dual-energy CT virtual keV monochromatic series for the assessment of spinal transpedicular hardware-bone interface. AJR Am J Roentgenol 201:878–883

    Article  PubMed  Google Scholar 

  10. Xia D, Roeske JC, Yu L, Pelizzari CA, Mundt AJ, Pan X (2005) A hybrid approach to reducing computed tomography metal artifacts in intracavitary brachytherapy. Brachytherapy 4:18–23

    Article  CAS  PubMed  Google Scholar 

  11. Boas FE, Fleischmann D (2011) Evaluation of two iterative techniques for reducing metal artifacts in computed tomography. Radiology 259:894–902

    Article  PubMed  Google Scholar 

  12. Pessis E, Campagna R, Sverzut JM, Bach F, Rodallec M, Guerini H et al (2013) Virtual monochromatic spectral imaging with fast kilovoltage switching: reduction of metal artifacts at CT. Radiographics 33:573–583

    Article  PubMed  Google Scholar 

  13. Barrett JF, Keat N (2004) Artifacts in CT: recognition and avoidance. Radiographics 24:1679–1691

    Article  PubMed  Google Scholar 

  14. Matsumoto K, Jinzaki M, Tanami Y, Ueno A, Yamada M, Kuribayashi S (2011) Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that obtained with conventional 120-kVp CT. Radiology 259:257–262

    Article  PubMed  Google Scholar 

  15. Lee M-J, Kim S, Lee S-A, Song HT, Huh YM, Kim DH et al (2007) Overcoming artifacts from metallic orthopedic implants at high-field-strength MR imaging and multi-detector CT. Radiographics 27:791–803

    Article  PubMed  Google Scholar 

  16. Brook OR, Gourtsoyianni S, Brook A, Mahadevan A, Wilcox C, Raptopoulos V (2012) Spectral CT with metal artifacts reduction software for improvement of tumor visibility in the vicinity of gold fiducial markers. Radiology 263:696–705

    Article  PubMed  Google Scholar 

  17. Matsuda I, Akahane M, Sato J, Katsura M, Kiryu S, Yoshioka N et al (2012) Precision of measurement of CT numbers: comparison of dual-energy CT spectral imaging with fast kVp switching and conventional CT with phantoms. Jpn J Radiol 30:34–39

    Article  PubMed  Google Scholar 

  18. Hui P, Wang XJ, Cui ZP, Sun H, Li TW, Yao HX et al (2011) The application of spectral CT imaging in reducing artifacts caused by metallic implants. Chin J Radiol 45:740–742

    Google Scholar 

  19. Guggenberger R, Winklhofer S, Osterhoff G, Wanner GA, Fortunati M, Andreisek G et al (2012) Metallic artefact reduction with monoenergetic dual-energy CT: systematic ex vivo evaluation of posterior spinal fusion implants from various vendors and different spine levels. Eur Radiol 22:2357–2364

    Article  CAS  PubMed  Google Scholar 

  20. Zhou C, Zhao YE, Luo S, Shi H, Li L, Zheng L, Zhang LJ et al (2011) Monoenergetic imaging of dual-energy CT reduces artifacts from implanted metal orthopedic devices in patients with factures. Acad Radiol 18:1252–1257

    Article  PubMed  Google Scholar 

  21. Zhang QH, Gao Y, Li KC, Li Y, Du XY, Yang J et al (2012) A comparative study of spectral CT GSI vs. MARS in removing artifacts associated with spinal metal implant in vertebral canal. J Clin Radiol 31:1338–1342

    Google Scholar 

  22. Zhao XS, Zhang HW, Li LX, Liu JG, Zhou MY, Yue KT et al (2011) Gemstone spectral imaging in removing metal artifacts produced by 125I radioactive particles. Chin J Med Imaging Technol 21:2344–2348

    Google Scholar 

Download references

Conflict of interest

We declared that we have no financial and personal relationships with other people or organizations that can inappropriately influence our work; there is no professional or other personal interest of any nature or kind in any product, service and/or company that could be construed as influencing the position presented in the manuscript entitled.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Yue Dong.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Dong, Y., Shi, A.J., Wu, J.L. et al. Metal artifact reduction using virtual monochromatic images for patients with pedicle screws implants on CT. Eur Spine J 25, 1754–1763 (2016). https://doi.org/10.1007/s00586-015-4053-4

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00586-015-4053-4

Keywords

Navigation