Abstract
Introduction
In 1959, Harmon was the first to implant a vitallium sphere into the intervertebral disc space in order to preserve mobility and maintain disc height in patients with back pain secondary to degenerative disc disease. Fernstrom reported good results after implantation of stainless steel spheres. However, a reduction in disc height frequently occurred secondary to sphere subsidence into the endplates. This often led to revision surgery, and the concept was abandoned.
Materials and methods
We report on four patients who underwent cobalt–chrome sphere implantation, and later presented to us with symptoms that were a direct result of sphere subsidence. All four patients presented with low back pain and/or lower extremity pain, and some with weakness. Imaging demonstrated that all patients had a loss of disc space height with sphere subsidence. Three patients underwent sphere removal, anterior interbody fusion using femoral ring allograft and posterior pedicle or facet screw fixation. In the fourth patient, the sphere was subsided into both the L5 and S1 endplates prohibiting removal. The patient underwent a posterior L5–S1 posterolateral fusion and facet screw fixation. The length of time from sphere implantation to the revision surgery ranges from 8 to 41 months.
Results
Ten months to 3 years after the revision surgery, all four patients reported an improvement in outcome.
Conclusion
To the best of our knowledge, this is the first report describing the complications resulting from cobalt chrome sphere implantation and revision strategies for managing the recently resurrected cobalt chrome sphere.
Similar content being viewed by others
References
Wang C, Auerbach J, Witschey W, Balderston R, Reddy R, Borthakur A (2007) Advances in magnetic resonance imaging for the assessment of degenerative disc disease of the lumbar spine. Semin Spine Surg. 19(2):65–71
Fujiwara A, Tamai K, An H et al (2000) The relationship between disc degeneration, facet joint osteoarthritis, and stability of the degenerative lumbar spine. J Spinal Disord 13(5):444–450
Freemont A, Peacock T, Goupille P, Hoyland J, O’Brien J, Jayson M (1997) Nerve ingrowth into diseased intervertebral disc in chronic back pain. Lancet 350(9072):178–181
Guyer R, McAfee P, Banco R et al (2009) Prospective, randomized, multicenter Food and Drug Administration investigational device exemption study of lumbar total disc replacement with the CHARITE artificial disc versus lumbar fusion: five-year follow-up. Spine J 9(5):374–86 Epub 2008 Sep 19
Harmon P (1963) Anterior excision and vertebral body fusion operation for intervertebral disk syndromes of the lower lumbar spine: three-to five-year results in 244 cases. Clin Orthop Relat Res 26:107–127
Fernstrom U (1966) Arthroplasty with intercorporal endoprosthesis in herniated disc and in painful disc. Acta Chir Scand 357:154–159
Reitz H, Joubert M (1964) Intractable headache and cervico-brachialgia treated by complete replacement of cervical intervertebral discs with a metal prosthesis. S Afr Med J 38:881–884
TS ItoM, Kaneda K (1993) A biomechanical definition of spinal segmental instability taking personal and disc level differences into account. Spine 18(15):2295–2304
Conflict of interest
None.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Siemionow, K.B., Hu, X. & Lieberman, I.H. The Fernstrom ball revisited. Eur Spine J 21, 443–448 (2012). https://doi.org/10.1007/s00586-011-2040-y
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00586-011-2040-y