Abstract
Pre-clinical wear testing of intervertebral disc prostheses is commonly carried out according to ISO 18192-1. Ten million multiaxial loading cycles are applied at a frequency of 1 Hz. At this frequency, testing takes about 4 months. Testing at higher frequencies would therefore be desirable. ISO 18192-1 also offers testing at 2 Hz; however, it says the impact on the implant material behaviour as well as on the accuracy of the test machine shall be investigated by the user. Since such data are not available so far, the aim of this study was to carry out comparative wear tests at 1 and 2 Hz. Seven Prodisc-L lumbar disc prostheses were tested. After a pre-soak period, the implants were placed in specimen cups filled with calf serum, mounted to a Spine Wear Simulator and loaded according to ISO 18192-1. Testing was carried out at a temperature of 37 ± 2°C. Four million loading cycles were applied at 1 Hz and eight million at 2 Hz in an alternating sequence. Each time after 12 days of testing the implants were removed to measure the weight and the height of the polyethylene cores. Then, the test serum was exchanged and the implants were remounted to the testing machine. The mean wear rate was 5.6 ± 2.3 mg per million cycles at 1 Hz and 7.7 ± 1.6 mg per million cycles at 2 Hz during the first six million loading cycles (p < 0.05) and 2.0 ± 0.6 and 4.1 ± 0.7 mg per million cycles during the second six million cycles (p < 0.05). Similarly, the mean heightloss was also smaller at 1 Hz than at 2 Hz (p < 0.05) with −0.02 ± 0.02 mm versus −0.04 ± 0.02 mm per million cycles during the first half of testing and −0.01 ± 0.01 versus −0.02 ± 0.01 mm per million cycles during the second half. The accuracy of the test machine was within the limits described by ISO 18192-1 at both frequencies. The results showed that the wear rate was higher at the beginning than at the end of testing. Also, the results indicated that testing at 2 Hz increases the wear rate compared with 1 Hz in case of a polyethylene-on-metal implant design. In the absence of retrieval studies it is difficult to decide which rate results in a more physiological wear pattern. However, a loading frequency of 1 Hz is probably closer to physiology than 2 Hz since the loading amplitudes prescribed by ISO 18192-1 are high. They rather represent movements like tying shoes or standing up from a chair than walking or sitting. The authors therefore suggest testing at 1 Hz.
Similar content being viewed by others
References
Berg S, Tullberg T, Branth B, Olerud C, Tropp H (2009) Total disc replacement compared to lumbar fusion: a randomised controlled trial with 2-year follow-up. Eur Spine J 18(10):1512–1519
Bishop NE, Waldow F, Morlock MM (2008) Friction moment of large metal-on-metal hip joint bearings and other modern designs. Med Eng Phys 30(8):1057–1064
Bushelow M, Nechtow W, Walker J (2008) Metal-on-Metal Lumbar TDR Design: a theoretical analysis of the affect of diameter clearance and surface finish on lubrication. In: 23rd Annual Meeting of the North American Spine Society P8
Fisher J, Chan KL, Hailey JL, Shaw D, Stone M (1995) Preliminary study of the effect of aging following irradiation on the wear of ultrahigh-molecular-weight polyethylene. J Arthroplasty 10(5):689–692
Godmann S (2005) Wear particulate and osteolysis. Orthop Clin North Am 36(1):41–48
Grupp TM, Yue JJ, Garcia R, Basson J, Schwiesau J, Fritz B, Blömer W (2009) Biotribological evaluation of artifical disc arthroplasty devices: influence of loading and kinematic patterns during in vitro wear simulation. Eur Spine J 18:98–108
Ingham E, Fisher J (2005) The role of macrophages in osteolysis of total joint replacement. Biomaterials 26(11):1271–1286
Kurtz SM, Peloza J, Siskey R, Villarraga ML (2005) Analysis of a retrieved polyethylene total disc replacement component. Spine J 5(3):344–350
Kurtz SM, van Ooij A, Ross R, de Waal Malefijt J, Peloza J, Ciccarelli L, Villarraga ML (2007) Polyethylene wear and rim fracture in total disc arthroplasty. Spine J 7(1):12–21
Liao YS, McKellop H, Lu Z, Campbell P, Benya P (2003) The effect of frictional heating and forced cooling on the serum lubricant and wear of UHMW polyethylene cups against cobalt–chromium and zirconia balls. Biomaterials 24(18):3047–3059
Nechtow W, Hintner M, Bushelow M, Kaddick C (2006) IVD replacement mechanical performance depends strongly on input parameters. In: 52nd Annual Meeting of the Orthopaedic Research Society Paper No. 0118
Nechtow W, Long W, Coombs D, Bushelow M, Hintner M, Ochs A, Kaddick C (2008) Vertebral motions segment dynamics influence Prodisc-L wear performance. In: 54th Annual Meeting of the Orthopaedic Research Society Paper No. 1928
Pare PE, Chan FW, Bhattacharya S, Goel VK (2009) Surface slide track mapping of implants for total disc arthroplasty. J Biomech 42:131–139
Punt IM, Cleutjens JP, de Bruin T, Willems PC, Kurtz SM, van Rhijn LW, Schurink GW, van Ooij A (2009) Periprosthetic tissue reactions observed at revision of total intervertebral disc arthroplasty. Biomaterials 30(11):2079–2084
Rosen C, Kiester PD, Lee TQ (2009) Lumbar disk replacement failures: review of 29 patients and rationale for revision. Orthopedics 32(8). doi:10.3928/01477447-20090624-08
van Ooij A, Kurtz SM, Stessels F, Noten H, van Rhijn L (2007) Polyethylene wear debris and long-term clinical failure of the Charite disc prosthesis: a study of 4 patients. Spine 32(2):223–229
van Ooij A, Oner FC, Verbout AJ (2003) Complications of artificial disc replacement: a report of 27 patients with the SB Charite disc. J Spinal Disord Tech 2003(16):4
Wilke HJ, Neef P, Caimi M, Hoogland T, Claes L (1999) New in vivo measurements of pressures in the intervertebral disc in daily life. Spine 24:755–762
Conflict of interest
None.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kettler, A., Bushelow, M. & Wilke, HJ. Influence of the loading frequency on the wear rate of a polyethylene-on-metal lumbar intervertebral disc replacement. Eur Spine J 21 (Suppl 5), 709–716 (2012). https://doi.org/10.1007/s00586-010-1582-8
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00586-010-1582-8