Abstract
Bismuth film electrodes (BiFEs) have a potential to replace toxic mercury used most frequently for determination of heavy metals (Cd, Pb, Zn) by anodic stripping voltammetry. We prepared a graphite disc electrode (0.5 mm in diameter) from a pencil-lead rod and developed a nitrogen doped diamond-like carbon (NDLC) microelectrode array consisting of 50,625 microdiscs with 3 μm in diameter and interelectrode distances of 20 μm on a highly conductive silicon substrate as a support for BiFEs. The disc graphite BiFE was used for simultaneous determination of Pb(II), Cd(II) and Zn(II) by square wave voltammetry (SWV) in an aqueous solution. We found the optimum bismuth-to-metal concentration ratio in the solution to be 20. The dependence of the stripping responses on the concentration of target metals was linear in the range from 1 × 10−8 to 1.2 × 10−7 mol/L. Detection limits 2.4 × 10−9 mol/L for Pb(II), 2.9 × 10−9 mol/L for Cd(II) and 1.2 × 10−8 mol/L for Zn(II) were estimated. A bismuth-plated NDLC microelectrode array was used for Pb(II) determination by differential pulse voltammetry (DPV) in an aqueous solution. We found that the stripping current for bismuth-plated NDLC array was linear in the concentration range of Pb(II) from 2 × 10−8 to 1.2 × 10−7 mol/L. The detection limit 2.2 × 10−8 mol/L was estimated from a calibration plot.
Similar content being viewed by others
References
Baldrianova L, Svancara I, Vlcek M, Economou A, Sotiropoulos S (2006a) Effect of Bi(III) concentration on the stripping voltammetric response of in situ bismuth-coated carbon paste and gold electrodes. Electrochim Acta 52:481–490
Baldrianova L, Svancara I, Economou A, Sotiropoulos S (2006b) Anodic stripping voltammetry at in situ bismuth-plated carbon and gold microdisc electrodes in variable electrolyte content unstirred solutions. Anal Chim Acta 580:24–31
Charalambous A, Economou A (2005) A study on the utility of bismuth-film electrodes for the determination of In(III) in the presence of Pb(II) and Cd(II) by square wave anodic stripping voltammetry. Anal Chim Acta 547:53–58
Demetriades D, Economou A, Voulgaropoulos A (2004) A study of pencil-lead bismuth-film electrodes for the determination of trace metals by anodic stripping voltammetry. Anal Chim Acta 519:167–172
Economou A (2005) Bismuth-film electrodes: recent developments and potentialities for electroanalysis. Trends Anal Chem 24:334–340
Hutton EA, Hocevar SB, Ogorevc B (2005) Ex situ praparation of bismuth film microelectrode for use in electrochemical stripping microanalysis. Anal Chim Acta 537:285–292
Kefala G, Economou A, Voulgaropoulos A, Sofoniou M (2003) A study of bismuth-film electrodes for the detection of trace metals by anodic stripping voltammetry and their application to the determination of Pb and Zn in tapwater and human hair. Talanta 61:603–610
Liu LX, Liu E (2005) Nitrogenated diamond-like carbon films for metal tracing. Surf Coatings Technol 198:189–193
Manivannan A, Kawasaki R, Tryk DA, Fujishima A (2004) Interaction of Pb and Cd during anodic stripping voltammetric analysis at boron-doped diamond electrodes. Electrochim Acta 49:3313–3318
Svancara I, Vytras K (2006) Elektroanalyza s bismutovymi elektrodami. Chem Listy 100:90–113
Wang J, Lu J, Kirgoz UA, Hocevar SB, Ogorevc B (2001a) Insights into the anodic stripping voltammetric behavior of bismuth film electrodes. Anal Chim Acta 434:29–34
Wang J, Kirgoz UA, Lu J (2001b) Stripping voltammetry with the electrode material acting as a ‘built-in’ internal standard. Electrochem Commun 3:703–706
Wang J, Lu J, Hocevar SB, Farias PAM (2000) Bismuth-coated carbon electrodes for anodic stripping voltammetry. Anal Chem 72:3218–3222
Acknowledgments
This work was supported by the grant of the Slovak Grant Agency Vega, Project 1/3095/06 and by Science and Technology Assistance Agency under the contract No. APVT-20-021004.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Rehacek, V., Hotovy, I., Vojs, M. et al. Bismuth film electrodes for heavy metals determination. Microsyst Technol 14, 491–498 (2008). https://doi.org/10.1007/s00542-007-0432-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00542-007-0432-z