Abstract
Purpose
Gas leakage around the cuff of a tracheal tube may frequently occur after tracheal intubation and inflation of the cuff. We assessed if the SmartCuff (Smiths Medical Japan, Tokyo, Japan), an automatic cuff pressure controller, would effectively prevent gas leakage.
Methods
Seventy adult patients were allocated randomly to one of two groups. After induction of general anesthesia and tracheal intubation, in one group (Syringe group), a syringe was used to inflate the cuff, until there was no audible gas leakage, at the airway pressure at 20 cmH2O. In the other group (SmartCuff group), the SmartCuff was used to maintain the cuff pressure to be 20 cmH2O. The mechanical ventilation (tidal volume of 8 ml.kg−1 and 12 breaths per min) was started. The incidence and percentage of gas leakage, and the proportion of adequate seal (defined as gas leakage of < 10%) between the groups were compared.
Results
The incidence of audible gas leakage was significantly higher in the Syringe group (10 of 35 patients (28%)) than in the SmartCuff group (none of 35 patients (0%)) (P = 0.00046, 95%CI for difference: 15–43%), and the proportion of adequate seal was significantly lower in the Syringe group (19 of 35 patients (54%)) than in the Smart cuff group (33 of 35 patients (94%)) (P = 0.0001, 95% CI for difference: 20–58%).
Conclusion
Gas leakage may frequently occur after tracheal intubation, and the use of the SmartCuff can effectively maintain the sealing effect of the cuff.
Similar content being viewed by others
Data availability
The data that support the findings of this study are available from the corresponding author upon reasonable request.
References
Seegobin RD, van Hasselt GL. Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. Br Med J. 1984;288:965–8.
Liu J, Zhang X, Gong W, Li S, Wang F, Fu S, Zhang M, Hang Y. Correlations between controlled endotracheal tube cuff pressure and postprocedural complications: a multicenter study. Anesth Analg. 2010;111:1133–7.
Combes X, Schauvliege F, Peyrouset O, Motamed C, Kirov K, Dhonneur G, Duvaldestin P. Intracuff pressure and tracheal morbidity: influence of filling with saline during nitrous oxide anesthesia. Anesthesiology. 2001;95:1120–4.
Asai T, Shingu K. Leakage of fluid around high-volume, low-pressure cuffs. A comparison of four tracheal tubes. Anaesthesia. 2001;56:39–42.
Rello J, Soñora R, Jubert P, Artigas A, Rué M, Vallés J. Pneumonia in intubated patients: role of respiratory airway care. Am J Resp Crit Care Med. 1996;154:111–5.
Leigh JM, Maynard JP. Pressure on the tracheal mucosa from cuffed tubes. Br Med J. 1979;1(6172):1173–4.
Dullenkopf A, Schmitz A, Frei M, Gerber AC, Weiss M. Air leakage around endotracheal tube cuffs. Eur J Anaesthesiol. 2004;21:448–53.
Braz JR, Volney A, Navarro LH, Braz LG, Nakamura G. Does sealing endotracheal tube cuff pressure diminish the frequency of postoperative laryngotracheal complaints after nitrous oxide anesthesia? J Clin Anesth. 2004;16:320–5.
Tsunoda N, Asai T, Okuda Y. Tracheal tube cuff pressure during anesthesia for robotic-assisted laparoscopic prostatectomy and the efficacy of an automatic cuff pressure controller (SmartCuff). J Anesth. 2023;37:234–41.
Mallampati SR, Gatt SP, Gugino LD, et al. A clinical sign to predict difficult tracheal intubation: a prospective study. Can J Anaesth. 1985;32:429–34.
Samsoon GLT, Young JRB. Difficult tracheal intubation: a retrospective study. Anaesthesia. 1987;42:487–90.
Tsunoda N, Asai T. A double-curved tube for McGrath® MAC videolaryngoscope-guided tracheal intubation. Br J Anaesth. 2022;128:e14–6.
Schober P, Vetter TR. Correct baseline comparisons in a randomized trial. Anesth Analg. 2019;129:639.
Altman DG. Clinical traial. In: Altman DG, editor. Practical statistics for medical research. London: Chapman & Hall; 1991. p. 440–76.
Faul F, Erdfelder E, Lang A-G, Buchner A. G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods. 2007;39:175–91.
Mac Murdo SD, Buffington CW. Brand and size matter when choosing a syringe to relieve pressure in a tracheal tube cuff. Anesth Analg. 2004;99:1445–9.
Bernhard WN, Yost LC, Turndorf H, Cottrell JE, Paegle RD. Physical characteristics of and rates of nitrous oxide diffusion into tracheal tube cuffs. Anesthesiology. 1978;48:413–7.
Funding
None.
Author information
Authors and Affiliations
Contributions
KU: This author helped in the study conception and design, proposal writing, data collection and management, as well as writing the initial draft and approving the final version of the manuscript. TA: This author made the study conception and design, proposal writing, data management, analysis and interpretation, as well as drafting and approving the final version of the manuscript. YO: This author helped in the study design, proposal writing, as well as revising and approving the final version of the manuscript.
Corresponding author
Ethics declarations
Conflict of interests
HA and TS, YO have no conflict of interest; TA is an Associate Editor-in-Chief of the Journal of Anesthesia.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Below is the link to the electronic supplementary material.
About this article
Cite this article
Urabe, K., Asai, T. & Okuda, Y. Usefulness of an automatic cuff pressure controller (SmartCuff) in inhibiting gasleakage around the cuff after tracheal intubation: a randomized controlled study. J Anesth 38, 86–91 (2024). https://doi.org/10.1007/s00540-023-03283-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00540-023-03283-4