Skip to main content
Log in

Usefulness of an automatic cuff pressure controller (SmartCuff) in inhibiting gasleakage around the cuff after tracheal intubation: a randomized controlled study

  • Original Article
  • Published:
Journal of Anesthesia Aims and scope Submit manuscript

Abstract

Purpose

Gas leakage around the cuff of a tracheal tube may frequently occur after tracheal intubation and inflation of the cuff. We assessed if the SmartCuff (Smiths Medical Japan, Tokyo, Japan), an automatic cuff pressure controller, would effectively prevent gas leakage.

Methods

Seventy adult patients were allocated randomly to one of two groups. After induction of general anesthesia and tracheal intubation, in one group (Syringe group), a syringe was used to inflate the cuff, until there was no audible gas leakage, at the airway pressure at 20 cmH2O. In the other group (SmartCuff group), the SmartCuff was used to maintain the cuff pressure to be 20 cmH2O. The mechanical ventilation (tidal volume of 8 ml.kg−1 and 12 breaths per min) was started. The incidence and percentage of gas leakage, and the proportion of adequate seal (defined as gas leakage of < 10%) between the groups were compared.

Results

The incidence of audible gas leakage was significantly higher in the Syringe group (10 of 35 patients (28%)) than in the SmartCuff group (none of 35 patients (0%)) (P = 0.00046, 95%CI for difference: 15–43%), and the proportion of adequate seal was significantly lower in the Syringe group (19 of 35 patients (54%)) than in the Smart cuff group (33 of 35 patients (94%)) (P = 0.0001, 95% CI for difference: 20–58%).

Conclusion

Gas leakage may frequently occur after tracheal intubation, and the use of the SmartCuff can effectively maintain the sealing effect of the cuff.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

Data availability

The data that support the findings of this study are available from the corresponding author upon reasonable request.

References

  1. Seegobin RD, van Hasselt GL. Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs. Br Med J. 1984;288:965–8.

    Article  CAS  Google Scholar 

  2. Liu J, Zhang X, Gong W, Li S, Wang F, Fu S, Zhang M, Hang Y. Correlations between controlled endotracheal tube cuff pressure and postprocedural complications: a multicenter study. Anesth Analg. 2010;111:1133–7.

    Article  PubMed  Google Scholar 

  3. Combes X, Schauvliege F, Peyrouset O, Motamed C, Kirov K, Dhonneur G, Duvaldestin P. Intracuff pressure and tracheal morbidity: influence of filling with saline during nitrous oxide anesthesia. Anesthesiology. 2001;95:1120–4.

    Article  CAS  PubMed  Google Scholar 

  4. Asai T, Shingu K. Leakage of fluid around high-volume, low-pressure cuffs. A comparison of four tracheal tubes. Anaesthesia. 2001;56:39–42.

    Article  Google Scholar 

  5. Rello J, Soñora R, Jubert P, Artigas A, Rué M, Vallés J. Pneumonia in intubated patients: role of respiratory airway care. Am J Resp Crit Care Med. 1996;154:111–5.

    Article  CAS  PubMed  Google Scholar 

  6. Leigh JM, Maynard JP. Pressure on the tracheal mucosa from cuffed tubes. Br Med J. 1979;1(6172):1173–4.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  7. Dullenkopf A, Schmitz A, Frei M, Gerber AC, Weiss M. Air leakage around endotracheal tube cuffs. Eur J Anaesthesiol. 2004;21:448–53.

    Article  CAS  PubMed  Google Scholar 

  8. Braz JR, Volney A, Navarro LH, Braz LG, Nakamura G. Does sealing endotracheal tube cuff pressure diminish the frequency of postoperative laryngotracheal complaints after nitrous oxide anesthesia? J Clin Anesth. 2004;16:320–5.

    Article  PubMed  Google Scholar 

  9. Tsunoda N, Asai T, Okuda Y. Tracheal tube cuff pressure during anesthesia for robotic-assisted laparoscopic prostatectomy and the efficacy of an automatic cuff pressure controller (SmartCuff). J Anesth. 2023;37:234–41.

    Article  PubMed  Google Scholar 

  10. Mallampati SR, Gatt SP, Gugino LD, et al. A clinical sign to predict difficult tracheal intubation: a prospective study. Can J Anaesth. 1985;32:429–34.

    Article  CAS  Google Scholar 

  11. Samsoon GLT, Young JRB. Difficult tracheal intubation: a retrospective study. Anaesthesia. 1987;42:487–90.

    Article  CAS  PubMed  Google Scholar 

  12. Tsunoda N, Asai T. A double-curved tube for McGrath® MAC videolaryngoscope-guided tracheal intubation. Br J Anaesth. 2022;128:e14–6.

    Article  PubMed  Google Scholar 

  13. Schober P, Vetter TR. Correct baseline comparisons in a randomized trial. Anesth Analg. 2019;129:639.

    Article  PubMed  Google Scholar 

  14. Altman DG. Clinical traial. In: Altman DG, editor. Practical statistics for medical research. London: Chapman & Hall; 1991. p. 440–76.

    Google Scholar 

  15. Faul F, Erdfelder E, Lang A-G, Buchner A. G*Power 3: A flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods. 2007;39:175–91.

    Article  PubMed  Google Scholar 

  16. Mac Murdo SD, Buffington CW. Brand and size matter when choosing a syringe to relieve pressure in a tracheal tube cuff. Anesth Analg. 2004;99:1445–9.

    Article  PubMed  Google Scholar 

  17. Bernhard WN, Yost LC, Turndorf H, Cottrell JE, Paegle RD. Physical characteristics of and rates of nitrous oxide diffusion into tracheal tube cuffs. Anesthesiology. 1978;48:413–7.

    Article  CAS  PubMed  Google Scholar 

Download references

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

KU: This author helped in the study conception and design, proposal writing, data collection and management, as well as writing the initial draft and approving the final version of the manuscript. TA: This author made the study conception and design, proposal writing, data management, analysis and interpretation, as well as drafting and approving the final version of the manuscript. YO: This author helped in the study design, proposal writing, as well as revising and approving the final version of the manuscript.

Corresponding author

Correspondence to Takashi Asai.

Ethics declarations

Conflict of interests

HA and TS, YO have no conflict of interest; TA is an Associate Editor-in-Chief of the Journal of Anesthesia.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Below is the link to the electronic supplementary material.

Supplementary file1 (DOCX 35 kb)

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Urabe, K., Asai, T. & Okuda, Y. Usefulness of an automatic cuff pressure controller (SmartCuff) in inhibiting gasleakage around the cuff after tracheal intubation: a randomized controlled study. J Anesth 38, 86–91 (2024). https://doi.org/10.1007/s00540-023-03283-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00540-023-03283-4

Keywords

Navigation