Abstract
Purpose
Intraoperative abnormalities of coagulation function may occur for various reasons. In most scenarios, treatment is directed by laboratory parameters. Unfortunately, standard laboratory testing may take 1–2 h. The purpose of the current study was to evaluate a point-of-care testing device (CoaguChek® XS System) in pediatric patients.
Methods
Patients ranging in age from 2 to 18 years, undergoing posterior spinal fusion (PSF) or cardiac surgery using cardiopulmonary bypass (CPB) were eligible for inclusion. After CPB and/or the surgical procedure, 2.8 ml of blood was obtained and simultaneously tested on both the standard laboratory apparatus and the CoaguChek® XS System.
Results
The study cohort consisted of 100 patients (50 PSF and 50 cardiac cases) with 13 cases excluded, leaving 87 patients (49 PSF and 38 cardiac cases) for analysis. In PSF cases, reference laboratory international normalized ratio (INR) ranged from 0.98 to 1.77 while CoaguChek® XS INR ranged from 1.0 to 1.3. The correlation coefficient was 0.69. The results of the Bland–Altman analysis showed a bias of 0.09, precision of 0.1, and 95% limits of agreement ranging from −0.11 to 0.28. In cardiac cases, reference INR ranged from 1.68 to 14.19, while CoaguChek® XS INR ranged from 1.4 to 7.9. The correlation coefficient was 0.35. The results of the Bland–Altman analysis showed a bias of −1.8, precision of 2.1, and 95% limits of agreement ranging from −6.0 to 2.4.
Conclusions
INR values obtained from CoaguChek® XS showed a moderate correlation with reference laboratory values within the normal range. However, in the presence of coagulopathy, the discrepancy was significantly greater, thereby making the CoaguChek® XS clinically unreliable.
Similar content being viewed by others
References
Spielmann N, Mauch JY, Madjdpour C, Schmugge M, Albisetti M, Weiss M, Haas T. Comparison of point-of-care testing (POCT): i-STAT® international normalized ratio (INR) vs reference laboratory INR in pediatric patients undergoing major surgery. Pediatr Anesth. 2011;21:1041–5.
Horlocker TT, Nuttall GA, Dekutoski MB, Bryant SC. The accuracy of coagulation tests during spinal fusion and instrumentation. Anesth Analg. 2001;93:33–8.
Levy JH, Dutton RP, Hemphill JC 3rd, Shander A, Cooper D, Paidas MJ, Kessler CM, Holcomb JB. Lawson JH; Hemostasis Summit Participants. Hemostasis Summit Participants. Multidisciplinary approach to the challenge of hemostasis. Anesth Analg. 2010;110:354–64.
American Society of Anesthesiologists Task Force on Perioperative Blood Management. Practice guidelines for perioperative blood management: an updated report by the American Society of Anesthesiologists Task Force on Perioperative Blood Management. Anesthesiology. 2015;122:241–75.
Meneghelo ZM, Barroso CM, Liporace IL, Cora AP. Comparison of the international normalized ratio levels obtained by portable coagulometer and laboratory in a clinic specializing in oral anticoagulation. Int J Lab Hematol. 2015;37:536–43.
Biedermann JS, Leebeek FW, Buhre PN, de Lathouder S, Barends JP, de Maat MP, van der Meer FJ, Kruip MJ. Agreement between Coaguchek XS and STA-R Evolution (Hepato Quick) INR results depends on the level of INR. Thromb Res. 2015;136:652–7.
Acknowledgements
This research was funded from intramural clinical research grant from Nationwide Children’s Hospital, the OSU CCTS, the CTSA Grant (UL1TR001070).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
No author has a conflict of interest with regard to this study.
About this article
Cite this article
Kako, H., Raman, V.T., Tumin, D. et al. Point-of-care testing for coagulation function: CoaguChek® XS System versus standard laboratory testing in pediatric patients with normal and abnormal coagulation function. J Anesth 31, 345–350 (2017). https://doi.org/10.1007/s00540-017-2317-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00540-017-2317-3