Real-time display of the stomach slow wave and its parameters in a newly designed electrogastrographic system

  • Published:
Journal of Gastroenterology Aims and scope Submit manuscript

Abstract:

We designed a new three-channel electrogastrographic (EGG) system, which was easily operated on the Windows 95 platform and could automatically provide slow wave parameters. The purpose of the present study was to test its reliability and accuracy in clinical recording. The system included a signal acquisition device assembled on a printed circuit board. Recorded myoelectrical signals were filtered, amplified, digitized, and transmitted via this device into a notebook personal computer (PC). Based on the short-term Fourier transform the software could transfer the time domain of the signal into the frequency domain. Real-time displayed slow wave parameters, including dominant frequency/power, percent of normal frequency (2–4 cpm), instability coefficient in frequency/power, and power ratio, were automatically renewed every 64 s. Twenty healthy subjects (M/F, 12/8; age, 23–51 years) were enrolled to measure both fast and postprandial myoelectrical activities for each 30-min recording. Our results indicated that meal ingestion significantly increased dominant frequency (3.15 ± 0.20 vs 3.23 ± 0.23 cpm; P < 0.05) and power (26.1 ± 3.8 vs 28.4 ± 3.9 dB; P < 0.05). The power ratio of the meal effect was 2.02 ± 2.07. Other parameters, including instability coefficient and percent of normal frequency, remained similar despite food ingestion. This newly designed EGG system is acceptable for clinically measuring gastric myoelectrical activity; the real-time display of many EGG parameters is an advantage with this new system.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: May 1, 2000 / Accepted: July 28, 2000

Rights and permissions

Reprints and permissions

About this article

Cite this article

Chang, FY., Lu, CL., Chen, CY. et al. Real-time display of the stomach slow wave and its parameters in a newly designed electrogastrographic system. J Gastroenterol 36, 10–17 (2001). https://doi.org/10.1007/s005350170148

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s005350170148

Navigation