Abstract
Background
Narrow band imaging (NBI) can emphasize images of the surface microvasculature of lesions, because the central wavelengths of the NBI filter are 415 and 540 nm and these wavelengths are well absorbed by hemoglobin. Flexible spectral imaging color enhancement (FICE) increases the contrast in depictions of mucosal lesions. However, quantitative evaluation of the image enhancement shown by NBI and FICE has not been reported. The aim of this study was to measure and compare the degrees of image enhancement in NBI and FICE.
Methods
We compared the visibility of human blood diluted with distilled water between that shown by white-light imaging (WLI) and that shown by NBI or FICE. One milliliter of human blood was plated onto a 12-well transparent plastic plate to set up doubling dilutions, from 1/2 to 1/223. High-definition endoscopes were used for each imaging method. A total of 11 endoscopists independently evaluated the visibility of the diluted blood. The median dilution was defined as the limit of visibility in each image.
Results
NBI enabled clearer visualization of the presence of blood compared with conventional WLI. NBI recognized blood contamination up to a 1/214 dilution, whereas conventional WLI recognized blood contamination up to a 1/211 dilution. In contrast, FICE did not improve the visualization of diluted blood and recognized blood contamination up to a 1/210 dilution.
Conclusions
NBI more effectively enhanced images of diluted blood compared to conventional WLI, while FICE did not improve the visualization of the diluted blood. These data suggest the usefulness of NBI for the early detection of gastrointestinal neoplasia, which is accompanied by abundant neovascularization.
Similar content being viewed by others
References
Kaltenbach T, Sano Y, Friedland S, Soetikno R. American Gastroenterological Association (AGA) Institute technology assessment on image-enhanced endoscopy. Gastroenterology. 2009;134:327–40.
Gono K, Yamazaki K, Doguchi N, Nonami T, Obi T, Yamaguchi M, et al. Endoscopic observation of tissue by narrow band illumination. Opt Rev. 2003;10:211–5.
Gono K, Obi T, Yamaguchi M, Ohyama N, Machida H, Sano Y, et al. Appearance of enhanced tissue feature in narrow-band endoscopic imaging. J Biomed Opt. 2004;9:568–77.
Muto M, Katada C, Sano Y, Yishida S. Narrowband imaging: a new diagnostic approach to visualize angiogenesis in the superficial neoplasm. Clin Gastroenterol Hepatol. 2005;3:S16–20.
Muto M, Nakane M, Katada C, Sano Y, Ohtsu A, Esumi H, et al. Squamous cell carcinoma in situ at oropharyngeal and hypopharyngeal mucosal sites. Cancer. 2004;101:1375–81.
Muto M, Ugumori T, Sano Y, Ohtsu A, Yoshida S. Narrow band imaging combined with magnified endoscopy for the cancer at the head and neck region. Dig Endosc. 2005;17:S23–4.
Yoshida T, Inoue H, Usui S, Satodate H, Fukami N, Kudo SE. Narrow-band imaging system with magnifying endoscopy for superficial esophageal lesions. Gastrointest Endosc. 2004;59:288–95.
Watanabe A, Taniguchi M, Tsujie H, Hosokawa M, Fujita M, Sasaki S. The value of narrow band imaging endoscope for early head and neck cancers. Otolaryngol Head Neck Surg. 2008;138:446–51.
Ugumori T, Muto M, Hayashi R, Hayashi T, Kishimoto S. Prospective study of early detection of pharyngeal superficial carcinoma with the narrowband imaging laryngoscope. Head Neck. 2009;31:189–94.
Muto M, Minashi K, Yano T, Saito Y, Oda I, Nonaka S, et al. Early detection of superficial squamous cell carcinoma in the head and neck region and esophagus by narrow band imaging: a multicenter randomized controlled trial. J Clin Oncol. 2010;28:1566–72.
Pohl J, May A, Rabenstein T, Pech O, Ell C. Computed virtual chromoendoscopy: a new tool for enhancing tissue surface structures. Endoscopy. 2007;39:80–3.
Nakayoshi T, Tajiri H, Matsuda K, Kaise M, Ikegami M, Sasaki H. Magnifying endoscopy combined with narrow band imaging system for early gastric cancer: correlation of vascular pattern with histopathology. Endoscopy. 2004;36:1080–4.
Sumiyama K, Kaise M, Nakayoshi T, Kato M, Mashiko T, Uchiyama Y, et al. Combined use of a magnifying endoscope with a narrow band imaging system and a multibending endoscope for en bloc EMR of early stage gastric cancer. Gastrointest Endosc. 2004;60:79–84.
Machida H, Sano Y, Hamamoto Y, Muto M, Kozu T, Tajiri H, et al. Narrow-band imaging in the diagnosis of colorectal lesions: a pilot study. Endoscopy. 2004;36:1094–8.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Muto, M., Higuchi, H., Ezoe, Y. et al. Differences of image enhancement in image-enhanced endoscopy: narrow band imaging versus flexible spectral imaging color enhancement. J Gastroenterol 46, 998–1002 (2011). https://doi.org/10.1007/s00535-011-0419-5
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00535-011-0419-5