13C-Urea breath test, using a new compact nondispersive isotope-selective infrared spectrophotometer: comparison with mass spectrometry

  • Published:
Journal of Gastroenterology Aims and scope Submit manuscript

Abstract

Background

The 13C-urea breath test (13C-UBT) is the most commonly used noninvasive method of detecting Helicobacter pylori infection. The isotope ratio mass spectrometer (IRMS) is the most commonly used device for this test, but the UBiT-IR300 infrared spectrophotometer, which, by comparison, is a more compact, less expensive, and easier to use analytical device, has now become widely used in the clinical setting in Japan. The objective of this study was to examine the diagnostic performance of the 13C-UBT, using the UBiT-IR300.

Methods

A multicenter open-label study was performed, in which the 13C-UBT was conducted using 100 mg of 13C-urea. Analysis of 13CO2 in the expired breath was performed by infrared spectroscopy and mass spectrometry, and assessment of H. pylori infection was performed by culture, histological examination, and rapid urease test.

Results

In 255 cases of H. pylori infection diagnosed by biopsy methods, the 13C-UBTs, performed with two different 13C-ruea formulations, and using infrared spectroscopy for evaluation, showed a sensitivity of 97.7%, specificity of 98.0%, and accuracy of 97.8% (total number of evaluable cases, n = 505). The rate of agreement in the assessment of H. pylori infection between infrared spectroscopy and mass spectrometry was 100% (n = 505). The regression equation for infrared spectroscopy to mass spectrometry was y = 0.9822x − 0.0809 (n = 2542), with a correlation coefficient of r = 0.99989 (P = 0.0001).

Conclusions

Diagnosis of H. pylori infection can be performed using infrared spectroscopy as well as mass spectrometry.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kato, M., Saito, M., Fukuda, S. et al. 13C-Urea breath test, using a new compact nondispersive isotope-selective infrared spectrophotometer: comparison with mass spectrometry. J Gastroenterol 39, 629–634 (2004). https://doi.org/10.1007/s00535-003-1357-7

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00535-003-1357-7

Key words

Navigation