Human vs robotic organ retraction during laparoscopic Nissen fundoplication

  • Published:
Surgical Endoscopy Aims and scope Submit manuscript

Abstract

Background: Advances in technique and instrumentation have enabled surgeons to perform an increasing number of complicated procedures through laparoscopy. However, these efforts have often been compromised by the exertion of excessive force when anatomical structures are retracted to create a clear view of the anatomy. Here, we present a comparative study of human and robotic performance in force-controlled organ retraction during laparoscopic Nissen fundoplication (LNF).

Methods: Six female pigs (20–25 kg) were anesthetized, intubated, and placed on mechanical ventilation; pneumoperitoneum (13 mmHg CO2) was established. A force-sensing retractor (FSR) was constructed to record the forces applied in retracting the stomach during dissection of the esophageal hiatus. The FSR was calibrated using known forces and then operated by either human alone or robot under human guidance using the FSR data. The esophageal hiatus was visualized and dissected, and LNF was completed.

Results: Less force was needed for robotic (74.3 ± 10.5 g; mean ± standard deviation) than for human (108.9 ± 34.3 g) retraction (p= 0.007) to obtain an optimal view of the esophageal hiatus. No significant differences were observed for retraction setup time (robot, 14.3 ± 0.8 min; human, 13.7 ± 9.9 min; mean ± SD) or hiatal dissection time (robot, 14.0 ± 3.0 min; human, 14.0 ± 6.1 min; mean ± SD).

Conclusions: These preliminary results illustrate our continuing effort to develop and evaluate an automated surgical assistant for laparoscopy. As more personnel-intensive advanced laparoscopic procedures are performed, robotic retraction is likely to offer a superior alternative to human retraction; it minimizes the forces exerted on the organs while maintaining excellent anatomical view.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 14 April 1998/Accepted: 1 August 1998

Rights and permissions

Reprints and permissions

About this article

Cite this article

Poulose, B., Kutka, M., Mendoza-Sagaon, M. et al. Human vs robotic organ retraction during laparoscopic Nissen fundoplication. Surg Endosc 13, 461–465 (1999). https://doi.org/10.1007/s004649901013

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s004649901013

Navigation