Background: In minimally invasive surgery intracorporal knot-tying is complicated by a limited field of vision and depth perception.

Methods: The ``knot before loop'' technique aims to reduce intraabdominal movements in number and space needed. A grasping instrument 3 mm in diameter guides a slipfit hollow knot pusher with a notch to hold the thread, when extracorporally forming the knot on the instrument tip and an axial slot. The loop is finished under endoscopic vision, yet a second loop is created along the thread. The knot is tightened and secured by closing the second loop without troublesome instrument change.

Results: The strength of the knot was tested and the feasibility of the instrument set was proven in pigs and 25 cholecystectomies and hernia repairs in humans.

Conclusions: Endoscopic application of a secured slip knot is simplified by the ``knot before loop'' technique. The independent formation of the knot by the assisting personnel allows quick application, equivalent to the use of clips and staples. The benefit in cost saving is high.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Author information

Authors and Affiliations

Authors

Additional information

Received 24 July 1996/Accepted: 12 December 1996

Rights and permissions

Reprints and permissions

About this article

Cite this article

Schaller, G., Paolucci, V., Gutt, C. et al. The ``knot before loop'' technique for the endoscopic ligation and suture-instrument set . Surg Endosc 11 , 785 –788 (1997). https://doi.org/10.1007/s004649900452

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s004649900452

Navigation