Abstract
Background
Gasketless laparoscopic insufflator systems are marketed for the ability to prevent desufflation of pneumoperitoneum during laparoscopy. However, surgeons raised concern for possible introduction of non-absorbable room air, including oxygen (O2), with these systems. A community-university collaborative was created to test this hypothesis.
Methods
An artificial abdomen, calibrated to equivalent compliance and volume of an average abdomen, was connected to a flow meter, oxygen concentration sensor, and commercially available laparoscopic gasketless cannula system. A commercially available gasketed cannula system was utilized as a control. Intra-abdominal concentration of oxygen was measured at 0–60 L per minute (L/min) of insufflated carbon dioxide (CO2) aspiration, as would occur during laparoscopic suctioning. For reference, a 5-mm laparoscopic suction device has an aspiration rate of approx. 42 L per minute. At the test facility, room air was 20.5% O2 at 50% humidity. Descriptive and univariate statistics were calculated with p < 0.05 considered significant.
Results
At 0 L/min CO2 aspiration, there was minimal (< 0.5%) oxygen detected intra-abdominally. However, with increasing rates of aspiration of pneumoperitoneum, increasing amounts of room air were detected intraabdominally in the gasketless versus gasketed cannula systems (mean ± standard deviation): 14.7 ± 1.2% versus 1.2 ± 0.5%, p < 0.0001 at 5 L/min aspiration, 18.1 ± 0.69% versus 1.1 ± 0.02%, p < 0.0001 at 10 L/min, 50.4 ± 2.19% vs 1.01 ± 0.003%, p < 0.0001 at 20 L/min. Above 25 L/min aspiration, the standard gasketed cannula systems experienced desufflation, but the gasketless system continued to entrain air to maintain insufflation: 64% room air at 30 L/min aspiration, 71% at 40 L/min aspiration, 77% at 50 L/min aspiration, and 84% at 60 L/min aspiration.
Conclusions
Gasketless cannula insufflation systems maintain abdominal insufflation by entraining non-medical room air. Especially at high aspiration rates, the majority of absorbable CO2 was replaced by non-medical room air, increasing potential for gas embolism with poorly absorbed oxygen and nitrogen. Authors have reported these experimental findings to the FDA and companies marketing these devices.
Similar content being viewed by others
References
Hong J, Kim W, Kil H (2010) Detection of subclinical CO2 embolism by transesophageal echocardiography during laparoscopic radical prostatectomy. Urology 75:581–584
Burcharth J, Burgdorf S, Lolle I, Rosenberg J (2012) Successful resuscitation after carbon dioxide embolism during laparoscopy. Surg Laparosc Endosc Percutan Tech 22:164–167
Sharma KC, Kabinoff G, Ducheine Y, Tierney J, Brandstetter RD (1997) Laparoscopic surgery and its potential for medical complications. Hear Lung J Acute Crit Care 26:52–67
Richter S, Matthes C, Ploenes T, Aksakal D, Wowra T, Hückstädt T, Schier F, Kampmann C (2013) Air in the insufflation tube may cause fatal embolizations in laparoscopic surgery: an animal study. Surg Endosc Other Interv Tech 27:1791–1797
Toung TJK, Rossberg M, Hutchins G (2016) Volume of air in a lethal venous air embolism. Anesthesiology 94:360–361
Menes T, Spivak H (2000) Laparoscopy: searching for the proper insufflation gas. Surg Endosc 14:1050–1056
Mirski MA, Lele AV, Fitzsimmons L, Toung TJ, Warltier DC, Mirski MA, Lele AV (2007) Diagnosis and treatment of vascular air embolism. Anesthesiology 106:164–177
Park EY, Kwon JY, Kim KJ (2012) Carbon dioxide embolism during laparoscopic surgery. Yonsei Med J 53:459–466
Fors D, Eiriksson K, Arvidsson D, Rubertsson S (2010) Gas embolism during laparoscopic liver resection in a pig model: frequency and severity. Br J Anaesth 105:282–288
Otsuka Y, Katagiri T, Ishii J, Maeda T, Kubota Y, Tamura A, Tsuchiya M, Kaneko H (2013) Gas embolism in laparoscopic hepatectomy: what is the optimal pneumoperitoneal pressure for laparoscopic major hepatectomy? J Hepatobiliary Pancreat Sci 20:137–140
Schmandra TC, Mierdl S, Bauer H, Gutt C, Hanisch E (2002) Transoesophageal echocardiography shows high risk of gas embolism during laparoscopic hepatic resection under carbon dioxide pneumoperitoneum. Br J Surg 89:870–876
Kobayashi S, Honda G, Kurata M, Tadano S, Sakamoto K, Okuda Y, Abe K (2016) An experimental study on the relationship among airway pressure, pneumoperitoneum pressure, and central venous pressure in pure laparoscopic hepatectomy. Ann Surg 263:1159–1163
Emilie Herrmann MD, Schaeffer-Reiss C, Clémentine PhD, Venot MD, Noll E, Girish MD, Joshi P, Pierre MBA,MD, Diemunsch A, PD MD (2017) CO2 pneumoperitoneum contamination with air during laparoscopic surgery: valveless versus conventional insufflation systems. Annu Meet Am Soc Anesthesiol 126:1425–1426
Muth CM, Shank ES (2000) Gas embolism. N Engl J Med 342:476–482
Hagen PT, Scholz DG, Edwards WD (1984) Incidence and size of patent foramen ovale during the first 10 decades of life: an autopsy study of 965 normal hearts. Mayo Clin Proc 59:17–20
Gutt CN, Oniu T, Mehrabi A, Schemmer P, Kashfi A, Kraus T, Büchler MW (2004) Circulatory and respiratory complications of carbon dioxide insufflation. Dig Surg 21:95–105
Funding
AirSeal InteliFlow insufflation system and cannulas and Stryker PneumoSure insufflation pumps were provided as a donation by Stryker Corporation and Lexion Medical, but these companies did not have any involvement with study design, data analysis, interpretation of results, or decision to publish. No external funding was provided for the study nor to the authors for their work.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Disclosures
Dr. Ciara Huntington, Mr. Prince, Ms. Hazelbaker, Mr. Lopes, Mr. Webb, Dr LeMaster, and Dr. Thomas Huntington declare that there are no conflicts of interest or financial ties to disclose with any products or corporations relevant to this publication.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Huntington, C.R., Prince, J., Hazelbaker, K. et al. Safety first: significant risk of air embolism in laparoscopic gasketless insufflation systems. Surg Endosc 33, 3964–3969 (2019). https://doi.org/10.1007/s00464-019-06683-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-019-06683-4