Skip to main content
Log in

DC current pulses for OTSC clip fragmentation: technology and experimental study

  • New Technology
  • Published:
Surgical Endoscopy Aims and scope Submit manuscript

Abstract

Background

The OTSC clip is used in endoscopic management of gastrointestinal lesions. In rare cases, the removal of the OTSC clip might be desirable. The objective of the study is to investigate feasibility, efficacy, and safety of a novel endoscopic instrument system for removal of the OTSC clip.

Methods

The study series has been conducted in a porcine model. Clip removal is performed with a prototype instrumentation that is designed to locally melt the clip by applying an electrical current pulse onto the clip structure. This system has been evaluated in an animal study (n = 10) in a pig model. A cap prototype with an elongated sleeve has been used for extraction of the OTSC clip fragments.

Results

23 of 24 implanted OTSC clips were successfully opened by applying in a total of 74 current pulses. Superficial mucosal coagulation marks were observed in 14 of the 24 application sites. No other findings such as hemorrhage, deep thermal wall lesions, or perforation were observed.

Conclusions

The study confirms the effectiveness and safety of the proposed endoscopic removal technique. Safe extraction of the clip fragments was feasible with an elongated sleeve at the distal cap. Limitations of the methods are the animal model and the experimental nature of the prototype instrumentation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Manta R, Galloro G, Mangiavillano B et al (2013) Over-the-scope clip (OTSC) represents an effective endoscopic treatment for acute GI bleeding after failure of conventional techniques. Surg 27:3162–3164

    Google Scholar 

  2. Weiland T, Fehlker M, Gottwald T, Schurr MO (2013) Performance of the OTSC System in the endoscopic closure of iatrogenic gastrointestinal perforations: a systematic review. Surg Endosc 27:2258–2274

    Article  PubMed  Google Scholar 

  3. Mennigen R, Colombo-Benkmann M, Senninger N, Laukoetter M (2013) Endoscopic closure of postoperative gastrointestinal leakages and fistulas with the over-the-scope clip (OTSC). J Gastrointest Surg 17:1058–1065

    Article  PubMed  Google Scholar 

  4. Mönkemüller K, Peter S, Toshniwal J et al (2013) Multipurpose use of the ‘bear claw’ (over-the-scope-clip system) to treat endoluminal gastrointestinal disorders. Dig Endosc 26:350–357

    Article  PubMed  Google Scholar 

  5. Weiland T, Fehlker M, Gottwald T, Schurr MO (2012) Performance of the OTSC system in the endoscopic closure of gastrointestinal fistulae–a meta-analysis. Minim Invasive Ther Allied Technol 21:249–258

    Article  PubMed  Google Scholar 

  6. Fähndrich M, Sandmann M, Heike M (2011) Removal of over the scope clips (OTSC) with an Nd:YAG laser. Z Gastroenterol 49:579–583

    Article  PubMed  Google Scholar 

  7. Mönkemüller K, Toshniwal J, Zabielski M (2012) Endoscopic removal of an over-the-scope-clip („bear claw“). Gastrointest Endosc 76:1077–1078

    Article  PubMed  Google Scholar 

  8. Neumann H, Diebel H, Mönkemüller K et al (2012) Description of a new, endoscopic technique to remove the over-the-scope-clip in an ex vivo porcine model (with video). Gastrointest Endosc 76:1009–1013

    Article  PubMed  Google Scholar 

  9. Arezzo A, Bullano A, Fischer H, Morino M (2013) The way to remove an over-the-scope-clip (with video). Gastrointest Endosc 77:974–975

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

The work leading to these results received funding from the German Federal Ministry for Economic Affairs and Energy, contract number EP121473.

Disclosures

Sebastian Schostek, Chi-Nghia Ho, Thomas Gottwald, and Marc Schurr disclose that they hold management positions with Ovesco Endoscopy AG.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Sebastian Schostek.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Schostek, S., Ho, C.N., Melbert, M. et al. DC current pulses for OTSC clip fragmentation: technology and experimental study. Surg Endosc 29, 2418–2422 (2015). https://doi.org/10.1007/s00464-014-3935-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00464-014-3935-6

Keywords

Navigation