Abstract
Background
The OTSC clip is used in endoscopic management of gastrointestinal lesions. In rare cases, the removal of the OTSC clip might be desirable. The objective of the study is to investigate feasibility, efficacy, and safety of a novel endoscopic instrument system for removal of the OTSC clip.
Methods
The study series has been conducted in a porcine model. Clip removal is performed with a prototype instrumentation that is designed to locally melt the clip by applying an electrical current pulse onto the clip structure. This system has been evaluated in an animal study (n = 10) in a pig model. A cap prototype with an elongated sleeve has been used for extraction of the OTSC clip fragments.
Results
23 of 24 implanted OTSC clips were successfully opened by applying in a total of 74 current pulses. Superficial mucosal coagulation marks were observed in 14 of the 24 application sites. No other findings such as hemorrhage, deep thermal wall lesions, or perforation were observed.
Conclusions
The study confirms the effectiveness and safety of the proposed endoscopic removal technique. Safe extraction of the clip fragments was feasible with an elongated sleeve at the distal cap. Limitations of the methods are the animal model and the experimental nature of the prototype instrumentation.
Similar content being viewed by others
References
Manta R, Galloro G, Mangiavillano B et al (2013) Over-the-scope clip (OTSC) represents an effective endoscopic treatment for acute GI bleeding after failure of conventional techniques. Surg 27:3162–3164
Weiland T, Fehlker M, Gottwald T, Schurr MO (2013) Performance of the OTSC System in the endoscopic closure of iatrogenic gastrointestinal perforations: a systematic review. Surg Endosc 27:2258–2274
Mennigen R, Colombo-Benkmann M, Senninger N, Laukoetter M (2013) Endoscopic closure of postoperative gastrointestinal leakages and fistulas with the over-the-scope clip (OTSC). J Gastrointest Surg 17:1058–1065
Mönkemüller K, Peter S, Toshniwal J et al (2013) Multipurpose use of the ‘bear claw’ (over-the-scope-clip system) to treat endoluminal gastrointestinal disorders. Dig Endosc 26:350–357
Weiland T, Fehlker M, Gottwald T, Schurr MO (2012) Performance of the OTSC system in the endoscopic closure of gastrointestinal fistulae–a meta-analysis. Minim Invasive Ther Allied Technol 21:249–258
Fähndrich M, Sandmann M, Heike M (2011) Removal of over the scope clips (OTSC) with an Nd:YAG laser. Z Gastroenterol 49:579–583
Mönkemüller K, Toshniwal J, Zabielski M (2012) Endoscopic removal of an over-the-scope-clip („bear claw“). Gastrointest Endosc 76:1077–1078
Neumann H, Diebel H, Mönkemüller K et al (2012) Description of a new, endoscopic technique to remove the over-the-scope-clip in an ex vivo porcine model (with video). Gastrointest Endosc 76:1009–1013
Arezzo A, Bullano A, Fischer H, Morino M (2013) The way to remove an over-the-scope-clip (with video). Gastrointest Endosc 77:974–975
Acknowledgments
The work leading to these results received funding from the German Federal Ministry for Economic Affairs and Energy, contract number EP121473.
Disclosures
Sebastian Schostek, Chi-Nghia Ho, Thomas Gottwald, and Marc Schurr disclose that they hold management positions with Ovesco Endoscopy AG.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Schostek, S., Ho, C.N., Melbert, M. et al. DC current pulses for OTSC clip fragmentation: technology and experimental study. Surg Endosc 29, 2418–2422 (2015). https://doi.org/10.1007/s00464-014-3935-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-014-3935-6