Background

Using a conventional Doppler probe, it is difficult to identify blood vessels precisely in lymph node dissection during laparoscopic surgery.

Methods

For 26 patients undergoing laparoscopic colectomy and 26 patients undergoing laparoscopic subtotal gastrectomy, we used conventional and vertical scanning Doppler probes to compare their ability to identify invisible blood vessels through differences in power output voltage.

Results

The vertical scanning Doppler probe was 100% successful in identifying the target blood vessels regardless of the patient’s body mass index (BMI), and also was superior to the conventional Doppler probe in detection sensitivity. The vertical scanning Doppler probe was useful for image training of recently qualified surgeons through real-time monitoring of the probe position and response on the monitor. Furthermore, findings showed that the vertical scanning Doppler probe was applicable to lymph node dissection as a forceps for laparoscopic usage.

Conclusion

The vertical scanning Doppler could identify the target blood vessels regardless of the patient’s BMI. It was verified to be useful as a surgical assistive device apart from being a sensor and may serve in the education of recently qualified surgeons.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1.
Fig. 2.
Fig. 3.
Fig. 4.
Fig. 5.

Similar content being viewed by others

References

  1. Fowler DL, White SA (1992) The use of a Doppler probe for identifying the cystic artery during a laparoscopic cholecystectomy: a pilot study. Surg Laparosc Endosc 2: 117–119

    Article  PubMed  CAS  Google Scholar 

  2. Loughlin KR, Brooks DC (1992) The use of a Doppler probe in laparoscopic surgery. J Laparoendosc Surg 2: 191–194

    PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to T. Ohdaira.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ohdaira, T., Nagai, H. A vertical scanning Doppler probe identifies blood vessels during laparoscopic surgery. Surg Endosc 21, 782–784 (2007). https://doi.org/10.1007/s00464-006-9094-7

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00464-006-9094-7

Keywords

Navigation