Abstract
Background
Intraoperative tumor localization often is difficult during laparoscopic surgery for early-stage upper gastrointestinal tumors.
Method
This study enrolled 15 patients undergoing laparoscopic gastrectomy. A magnetic marking clip–detecting system was used to apply a marking clip to the tumor site during preoperative gastroscopy, and to detect a marking magnetic body.
Results
In a basic ex vivo study, the mean permeance rate on the gastric wall serosal surface was 0.97 ± 0.01. Magnetic flux densities required for exploration were 52.6 mT on the gastric wall serosal surface and 312.4 mT 10 mm away from the surface. In a clinical study, the mean distance between the detected tumor site and the clip along the longitudinal axis was 8.3 ± 3.2 mm. The mean detection time was 5.7 ± 2.3 min.
Conclusion
The magnetic marking clip–detecting system may be useful for tumor site detection during laparoscopic gastrectomy.
Similar content being viewed by others
References
Allam ME, Mehta D, Zelen J, Fogler R (1998) Posterior wall gastric leiomyoma: endoscopic tattooing facilitates laparoscopic resection. JSLS 2: 83–84
Gamal EM, Altorjay A, Szanto I, Garcia J, Kiss J (1999) Laparoscopic wedge resection of the gastric wall for gastric benign tumour: the collaboration of the laparoscopic surgeon and the endoscopist. Acta Chir Hung 38: 167–168
Hachisu I, Miyazaki S, Hamaguchi K (1989) Endoscopic clip marking of lesions using the newly developed HX-3L clip. Surg Endosc 3: 142–147
Hornig D, Kuhn H, Stadelmann O, Botticher R (1983) Phlegmonous gastritis after Indian ink marking. Endoscopy 15: 266–269
Jensen DM, Machicado GA, Hirabayashi K (2004) Randomized controlled comparison of hemostasis efficacy and retention of three types of hemoclips for bleeding canine ulcers. Am J Gastroenterol 99: 31
Montorsi M, Opocher E, Santambrogio R, Bianchi P, Faranda C, Arcidiacono P, Passoni GR, Cosentino F (1999) Original technique for small colorectal tumor localization during laparoscopic surgery. Dis Colon Rectum 42: 819–822
Ohdaira T, Nagai H, Shibusawa H (2005) Intraoperative localization of early-stage gastrointestinal tumors using a marking clip detector system. Surg Technol Int 14: 79–83
Ohdaira T, Shoji M, Nagai H (2003) Intraoperative localization of colorectal tumors in the early stages using magnetic marking clip detector system (MMCDS). Surg Endosc 17: 692–695
Pillet S, Souhassou M, Lecomte C (2004) Electron density studies of molecular magnetic materials. Acta Crystallogr A 60: 455–464
Snider TE, Goodell WM, Pulitzer DR (1994) Gastrointestinal tattoos. Pathol Lab Med 118: 640–641
Weyman RL, Rao SS (1999) A novel clinical application for endoscopic mucosal clipping. Gastrointest Endosc 49: 522–524
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Ohdaira, T., Nagai, H. Intraoperative localization of early-stage upper gastrointestinal tumors using a magnetic marking clip–detecting system. Surg Endosc 21, 810–815 (2007). https://doi.org/10.1007/s00464-006-9037-3
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-006-9037-3