Skin wound closure with a novel shape-memory alloy fixator

  • Published:
Surgical Endoscopy And Other Interventional Techniques Aims and scope Submit manuscript

Abstract

Background

Previously, we have described novel, thermally deployable tissue fixators based on the shape-memory properties of nickel titanium. The present study reports on the in vivo wound closure with these fixators as a preliminary to evaluating their use for bowel anastomoses.

Methods

Twenty adult Sprague-Dawley rats had 2.0-cm dorsal skin incisions approximated either with shape-memory alloy (SMA) fixators or conventional skin staples on a random basis. Electrical resistance heating was used to transform and deploy the SMA fixators into the wound. The rats were killed on day 14 or day 28 and wound specimens were harvested for force distraction studies and histologic examination.

Results

There was no incidence of wound dehiscence. 14- and 28-day wounds from both groups showed no significant difference in breaking force or energy. Histology revealed appropriate stages of wound healing for both SMA-closed and control wounds.

Conclusions

The results confirm the efficacy and safety of tissue-edge approximation with SMA fixators.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. ASTM F86: (2001) Surface preparation and marking of metallic surgical implants. Annual book of ASTM standards. ASTM Book of Standards Section 13 (Volumes 13.01 & 13.02) 2004, ISBN: 0803137583

  2. Athanasoulis CA, Kaufman JA, Halpern EF, Waltman AC, Geller SC, Fan CM (2000) Inferior vena caval filters: review of a 26 year single center clinical experience. Radiology 216(1): 54–66

    CAS  PubMed  Google Scholar 

  3. Brown S, Frank T, El Shallaly G, Cuschieri A (2003) Comparison of conventional and gaze down imaging in laparoscopic task performance. Surg Endosc 17: 586–590

    CAS  PubMed  Google Scholar 

  4. Campbell PA, Song C, Frank TG, Cuschieri A (2001) Dynamic thermography of shape memory alloy surgical fixators. Thermol Int 11(1): 5–10

    Google Scholar 

  5. Dai KR, Hou XK, Sun YH, Tang RG, Qiu SJ, Ni C (1993) Treatment of intra-articular fractures with shape memory compression staples. Injury 24(10): 651–655

    Article  CAS  PubMed  Google Scholar 

  6. Frank TG, Xu W, Stockham G, Cuschieri A (1999) Development of SMA fixation devices for minimal access surgery. SMST-99, Proceedings of the First European Conference on Shape Memory and Superelastic Technologies, Antwerp, Belgium 1999

  7. Frank TG, Xu W, Cuschieri A (2000) Instruments based on shape memory alloy properties for minimal access surgery, interventional radiology and flexible endoscopy. Min Invas Ther Allied Technol 9(2): 89–98

    Article  Google Scholar 

  8. Hanna GB, Shimi S, Cuschieri A (1997) Influence of direction of view, target-to-endoscope distance and manipulation angle on endoscopic knot tying. Br J Surg 84(10): 1460–1464

    CAS  PubMed  Google Scholar 

  9. Ikeuchi D, Onodera H, Aung T, Kan S, Kawamoto K, Imamura M, Maetani S (1999) Correlation of tensile strength with bursting pressure in the evaluation of intestinal anastomosis. Dig Surg 16: 478–485

    Article  CAS  PubMed  Google Scholar 

  10. Levenson SM, Geever EF, Crowley LV, Oates JF 3rd, Berard CW, Rosen H (1965) The healing of rat skin wounds. Ann Surg 161(2): 293–308

    CAS  PubMed  Google Scholar 

  11. Ng Y, Song C, Frank T, McLean D, Shimi S, Cuschieri A, Campbell P (2003) Optimised deployment of heat-activated surgical staples using thermography. Appl Phys Lett 83(9): 1884–1886

    CAS  Google Scholar 

  12. Ng Y, Shimi S, Frank T, Campbell P, Kernohan N, Cuschieri A (2004) European Association of Endoscopic Surgeons, Shape memory NiTi fixator, Barcelona, Spain, June 9–12, 2004

  13. Oxlund H, Christensen H, Seyer-Hansen M, Anddreassen T (1996) Collagen deposition and mechanical strength of colon anastomoses and skin incisional wounds of rats. J Surg Res 66: 25–30

    Article  CAS  PubMed  Google Scholar 

  14. Shabalovskaya S (2002) Surface, corrosion and biocompatibility aspects of Nitinol as an implant material. Biomed Mater Eng 12: 69–109

    CAS  PubMed  Google Scholar 

  15. Simon M, Kaplow R, Salzman E, Freiman D (1977) A vena cava filter using thermal shape memory alloy. Radiology 125: 89–94

    Google Scholar 

  16. Song C, Campbell PA, Frank TG, Cuschieri A (2002) Thermal modelling of shape memory alloy fixator for medical application. Smart Mater Struct 11: 1–5

    Article  Google Scholar 

  17. Tutein Nolthenius RP, van Herwaarden JA, van den Berg JC, van Marrewijk C, Teijink JA, Moll FL (2001) Three year single centre experience with the AneuRx aortic stent graft. Eur J Vasc Endovasc Surg 22: 257–64

    Article  CAS  PubMed  Google Scholar 

  18. Watters DA, Smith AN, Eastwood MA, Anderson KC, Elton RA, Mugerwa JW (1985) Mechanical properties of the colon: comparison of the mechanical properties of African and European colon in vitro. Gut 26(4): 384–392

    CAS  PubMed  Google Scholar 

  19. Xu W, Frank TG, Stockham G, Cuschieri A (1999) Shape memory alloy fixator system for suturing tissue in minimal access surgery. Ann Biomed Eng 27: 663–669

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgments

We thank the technicians of the Tissue Bank at Ninewells Hospital for invaluable assistance with preparation of specimens, particularly George Thomson and Andy Grant. This work was funded by Wellcome Trust Grant 057735.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to A. Cuschieri.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ng, Y., Shimi, S.M., Kernohan, N. et al. Skin wound closure with a novel shape-memory alloy fixator. Surg Endosc 20, 311–315 (2006). https://doi.org/10.1007/s00464-004-0001-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00464-004-0001-9

Keywords

Navigation