Abstract
The study was performed to evaluate the accuracy of the StatStrip (SS) and SureStep Flexx (SF) glucose meters compared to plasma glucose in infants at risk for neonatal hypoglycemia and to determine the effect of bilirubin and hematocrit on the results. A prospective cross-sectional study was conducted on 172 venous blood glucose samples from infants who had initial low point-of-care (POC) glucose tests measured simultaneously by SS and SF. Plasma glucose levels were compared to both POC instruments, and the effect of bilirubin and hematocrit levels on mean glucose differences were analysed. Mean (SD) plasma glucose was 2.12 (0.45) mmol/L; (range, 1.11–3.06 mmol/L). Mean (1.96SD) glucose differences of the SS versus SF were 0.21 (0.70) mmol/L and −0.04 (0.78) mmol/L, respectively. SS sensitivity was 94.7 % with an 86.1 % negative predictive value (NPV) at 2.8 mmol/L, while the SF had a 100 % sensitivity and NPV at the same cut-off level. No correlations were identified between mean glucose differences and either hematocrit or bilirubin levels in both glucose meters. Both the SS and SF glucose meters have limited use when compared to plasma glucose. Hence, they can only be employed as screening tools in at-risk neonates with an appropriate, predetermined cut-off level. Hematocrit and bilirubin levels did not affect the accuracy of both devices.
Similar content being viewed by others
Abbreviations
- SS:
-
StatStrip
- SF:
-
SureStep Flexx
- PG:
-
Plasma glucose
- AAP:
-
American Academy of Pediatrics
- POC:
-
Point-of-care
- ISO:
-
International Organization for Standardization
- ADA:
-
American Diabetes Association
- PPV:
-
Positive predictive value
- NPV:
-
Negative predictive value
References
Adamkin DH (2011) Postnatal glucose homeostasis in late-preterm and term infants. Pediatrics 127:575–579
American Diabetes Association (1994) Self-monitoring of blood glucose. Diabetes Care 17:81–86
Boluyt N, van Kempen A, Offringa M (2006) Neurodevelopment after neonatal hypoglycemia: a systematic review and design of an optimal future study. Pediatrics 117:2231–2243
Boyd R, Leigh B, Stuart P (2005) Capillary versus venous bedside blood glucose estimations. Emerg Med J 22:177–179
Burns CM, Rutherford MA, Boardman JP, Cowan FM (2008) Patterns of cerebral injury and neurodevelopmental outcomes after symptomatic neonatal hypoglycemia. Pediatrics 122:65–74
Cornblath M, Hawdon JM, Williams AF, Aynsley-Green A, Ward-Platt MP, Schwartz R, Kalhan SC (2000) Controversies regarding definition of neonatal hypoglycemia: suggested operational thresholds. Pediatrics 105:1141–1145
Dalgic N, Ergenekon E, Soysal S, Koc E, Atalay Y, Gucuyener K (2002) Transient neonatal hypoglycaemia—long-term effects on neurodevelopmental outcome. J Pediatr Endocrinol Metab 15:319–324
Deshpande S, Ward PM (2005) The investigation and management of neonatal hypoglycaemia. Semin Fetal Neonatal Med 10:351–361
Engle WA, Tomashek KM, Wallman C (2007) "Late-preterm" infants: a population at risk. Pediatrics 120:1390–1401
Girouard J, Forest JC, Masse J, Leroux M, Bradburn NC, Noblet TC, Joynes JO, Baum J (2000) Multicenter evaluation of the Glucometer Elite XL meter, an instrument specifically designed for use with neonates. Diabetes Care 23:1149–1153
Haninger NC, Farley CL (2001) Screening for hypoglycemia in healthy term neonates: effects on breastfeeding. J Midwifery Womens Health 46:292–301
Harris DL, Weston PJ, Harding JE (2012) Incidence of neonatal hypoglycemia in babies identified as at risk. J Pediatr 161:787–791
Hay WW Jr, Raju TN, Higgins RD, Kalhan SC, Devaskar SU (2009) Knowledge gaps and research needs for understanding and treating neonatal hypoglycemia: workshop report from Eunice Kennedy Shriver National Institute of Child Health and Human Development. J Pediatr 155:612–617
Ho HT, Yeung WK, Young BW (2004) Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice. Arch Dis Child Fetal Neonatal Ed 89:F356–F359
Holtzinger C (2008) Evaluation of a new POCT bedside glucose meter and strip with hematocrit and interference corrections. Point Care 7:1–6
ISO 15197 (2003) In vitro diagnostic test systems—requirements for blood–glucose monitoring systems for self-testing in managing diabetes mellitus ISO 15197: 2003(E). International Standard Organization, Geneva
Karon BS, Griesmann L, Scott R, Bryant SC, Dubois JA, Shirey TL, Presti S, Santrach PJ (2008) Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters. Diabetes Technol Ther 10:111–120
Kinnala A, Rikalainen H, Lapinleimu H, Parkkola R, Kormano M, Kero P (1999) Cerebral magnetic resonance imaging and ultrasonography findings after neonatal hypoglycemia. Pediatrics 103:724–729
le Roux CW, Wilkinson SD, Pavitt DV, Muller BR, Alaghband-Zadeh J (2004) A new antiglycolytic agent. Ann Clin Biochem 41:43–46
Makaya T, Memmott A, Bustani P (2012) Point-of-care glucose monitoring on the neonatal unit. J Paediatr Child Health 48:342–346
Mann EA, Salinas J, Pidcoke HF, Wolf SE, Holcomb JB, Wade CE (2008) Error rates resulting from anemia can be corrected in multiple commonly used point-of-care glucometers. J Trauma 64:15–20
Montagnana M, Caputo M, Giavarina D, Lippi G (2009) Overview on self-monitoring of blood glucose. Clin Chim Acta 402:7–13
NCCLS (2002) In: Sacks DB (ed) Point-of-care blood glucose testing in acute and chronic care facilities; approved guideline—second edition. The NCCLS, Wayne, pp C30–A32
Ngerncham S, Piriyanimit S, Kolatat T, Wongsiridej P, Inchgarm L, Kitsommart R, Vutrapongwatana P, Jeerapaet K (2012) Validity of two point of care glucometers in the diagnosis of neonatal hypoglycemia. Indian Pediatr 49:621–625
Nuntnarumit P, Chittamma A, Pongmee P, Tangnoo A, Goonthon S (2011) Clinical performance of the new glucometer in the nursery and neonatal intensive care unit. Pediatr Int 53:218–223
Roth-Kleiner M, Stadelmann Diaw C, Urfer J, Ruffieux C, Werner D (2010) Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting. Eur J Pediatr 169:1387–1395
Straussman S, Levitsky LL (2010) Neonatal hypoglycemia. Curr Opin Endocrinol Diabetes Obes 17:20–24
Vanavanan S, Santanirand P, Chaichanajarernkul U, Chittamma A, Dubois JA, Shirey T, Heinz M (2010) Performance of a new interference-resistant glucose meter. Clin Biochem 43:186–192
Acknowledgments
This study was financially supported by the Siriraj Research Development Fund (Managed by Routine to Research: R2R). The study was investigator initiated, and the sponsors were not involved in the conduct or results of the study. The StatStrip glucose meter and strips were provided by the local distributor in Bangkok, Thailand.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kitsommart, R., Ngerncham, S., Wongsiridej, P. et al. Accuracy of the StatStrip versus SureStep Flexx glucose meter in neonates at risk of hypoglycemia. Eur J Pediatr 172, 1181–1186 (2013). https://doi.org/10.1007/s00431-013-2019-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00431-013-2019-2