Abstract
Data on the effects of a prolonged inflation time during the resuscitation of very prematurely born infants are limited; one study showed no effect, and in another, although lower bronchopulmonary dysplasia (BPD) rates were seen, that effect could have been due to the prolonged inflation time, the positive end expiratory pressure applied or the combination of the two. The aims of our study were to assess the length of inflation times used during face mask and t-piece resuscitation of prematurely born infants in the labour suite and determine whether prolonged inflations led to longer inflation flow times. A respiration monitor (NM3 respiratory profile monitor) was used to record flow, airway pressure and tidal volume changes. The first five inflations for each baby were analysed. Forty prematurely born infants (median gestational age 30, range 26–32 weeks) were examined. Their median inflation pressure was 17.6 (range 12.2–27.4) cm H2O, inflation time 0.89 (range 0.33–2.92) s, expiratory tidal volume 1.01 (range 0.02–11.41) ml/kg and inflation flow time 0.11 (range 0.04–0.54) s. There was no significant relationship between the inflation time and the inflation flow time, but there was a significant relationship between the inflation pressure and the inflation flow time (p = 0.024). Conclusion: These results suggest that prolonging inflation times during face mask resuscitation of prematurely born infants would not improve ventilation as prolonged inflation did not lead to longer inflation flow times.
Similar content being viewed by others
References
Harling AE, Beresford MW, Vince GS, Bates M, Yoxall CW (2005) Does sustained lung inflation at resuscitation reduce lung injury in the preterm infant? Arch Dis Child Fetal Neonatal Ed 90:F406–F410
Klingenberg C, Dawson JA, Gerber A, Kamlin C, Pg D, Morley CJ (2011) Sustained inflations: comparing three neonatal resuscitation devices. Neonatology 100:78–84
McHale S, Thomas M, Hayden E, Bergin K, McCallion N, Molloy EJ (2008) Variation in inspiratory time and tidal volume with T-piece neonatal resuscitator: assocation with operator experience and distraction. Resuscitation 79:230–233
Murthy V, Dattani N, Fox G, Campbell M, Milner AD, Greenough A (2011) The first five inflations during resuscitation of prematurely born infants. Arch Dis Child (in press)
Resuscitation Council (UK) (2010) Resuscitation guidelines (Ed J P Nolan) www.resus.org.uk
Schmolzer GM, Dawson J, Kamlin CO, O’Donnell CP, Morley CJ, Davis PG (2011) Airway obstruction and gas leak during mask ventilation of preterm infants in the delivery room. Arch Dis Child Fetal Neonatal Ed 96:F254–F257
Schmolzer GM, Kamlin OC, O’Donnell CP, Dawson JA, Morley CJ, Davis PG (2010) Assessment of tidal volume and gas leak during mask ventilation of preterm infants in the delivery room. Arch Dis Child Fetal Neonatal Ed 95:F393–F397
te Pas AB, Walther FJ (2007) A randomised, controlled trial of delivery room respiratory management in very preterm infants. Pediatrics 120:322–329
Upton CJ, Milner AD (1991) Endotracheal resuscitation of neonates using a rebreathing bag. Arch Dis Child 66:39–42
Vyas H, Milner AD, Hopkin IE, Boon AW (1981) Physiologic responses to prolonged and slow rise inflation in the resuscitation of the asphyxiated newborn infant. J Pediatr 99:635–669
Acknowledgements
This study was funded by the Guy’s & St Thomas’ Charitable Trust. We would like to thank Mrs. Deirdre Gibbons for the secretarial assistance.
Competing interests
There are no conflicts of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Murthy, V., Creagh, N., Peacock, J.L. et al. Inflation times during resuscitation of preterm infants. Eur J Pediatr 171, 843–846 (2012). https://doi.org/10.1007/s00431-011-1661-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00431-011-1661-9