Evaluation of a safe infusion device on reducing occupational exposure of nurses to antineoplastic drugs: a comparative prospective study. Contamoins-1

  • Original Article
  • Published:
International Archives of Occupational and Environmental Health Aims and scope Submit manuscript

Abstract

Purpose

Despite the decreasing of environmental contamination throughout the anticancer drug circuit, the administration of chemotherapies remains at risk of occupational exposure for nurses. Many medical devices aim at securing administration, but none have been scientifically evaluated to verify the actual improvement.

Methods

A monocentric comparative before/after study was carried out in an oncology day hospital to evaluate the efficacy of Safe Infusion Devices in reducing drug exposure compared to usual infusion practices. The rate of nurses’ gloves contamination was estimated. To avoid false negatives and to ensure sampling reproducibility, each sample of gloves was contaminated with a drop of topotecan. Association between contamination and other variables was investigated using a multivariate logistic regression analysis.

Results

The usual practice led to a rate of 58.3% of contaminated samples while Safe Infusion Devices to a rate of 15%: Safe Infusion Devices reduced the risk of gloves contamination by 85% in multivariate analysis (Odds ratio = 0.15; 95% confidence interval = 0.05–0.46; p < 0.001). Topotecan was identified in 100% of the samples. Only one case of cross-contamination has occurred.

Conclusion

Despite the current practice of using neutral solvent-purged infusers, the occupational exposure remains high for nurses and Safe Infusion Devices significantly reduced this risk of exposure. However, glove contamination is only a surrogate endpoint. The results confirmed that the disconnection of empty bags resulted in occupational exposure. Except a contamination due to the leakage of a bag, no cross-contamination was detected. Safe Infusion Devices were highly effective but did not completely eliminate exposure.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

Data availability

The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.

References

Download references

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

FF: Conception and design, interpretation of data, draft, final approval; EB, CM: Conception and design, article revising, final approval; Bastien Raymond: Sample collection, article revising, final approval; SM, AM, MH, WB, XS: Interpretation of data, article revising, final approval; FT: Statistical analysis, interpretation of data, article revising, final approval; JG: Samples analysis, interpretation of data, article revising, final approval.

Corresponding author

Correspondence to Fabien Forges.

Ethics declarations

Conflict of interest

No conflicts of interest to declare.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Forges, F., Blanc, E., Raymond, B. et al. Evaluation of a safe infusion device on reducing occupational exposure of nurses to antineoplastic drugs: a comparative prospective study. Contamoins-1. Int Arch Occup Environ Health 94, 1317–1325 (2021). https://doi.org/10.1007/s00420-021-01679-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00420-021-01679-x

Keywords

Navigation