Study design
To study the effect of respirable and ultrafine particles on epithelial function, we aimed to examine altogether 60 furnace workers from three metal smelters in Norway with Endo-PAT. The Endo-PAT2000 device (Itamar Medical, Israel) is a simple, non-invasive, operator-independent method for assessing both BPA and reactive hyperemic peripheral arterial tone (Axtell et al. 2010). Each furnace worker was to be examined twice; after a full-day shift at work (WD), and on a day off (DO), thereby making each participant their own control. Exposure measurements were performed on the same WD as the health examination was carried out. No air samples were collected on the DO, assuming a zero exposure on DO.
Two manganese smelters (plant A) and one silicon metal smelter (plant B) were contacted and invited to participate in the study, which all plants accepted. All plants practiced a full-shift work schedule, with five teams working in permanent rotating shifts. The recruitment of participants was organized by the health and safety management of each plant. They asked four workers at each shift team if they would voluntarily participate at both time points during the study, giving 20 participants from each plant. If a worker refused, or was unable to participate in the examinations, another worker was asked to take his or her place in the study. The main functions of the furnace workers were to control the tapping process (tappers) and to operate the cranes (crane operators). Tappers were monitoring the tapping process, either close (3–5 m) to the tapping hole, or nearby in a pressurized cabin. Crane operators performed their tasks mainly after the end of the tapping process, and 5–10 m further away from the tapping hole. Each tapping/crane driving process is performed three-to-four times during the shift, and takes 45–75 min each time. When they were not working at the tap floor, both tappers and crane drivers were resting in a control area, not exposed to PM {Berlinger, 2015 #870}. At two of the smelters, the furnace workers had either a tapper or crane operator function during their shift, whereas at the third plant, all workers performed both tapping and operating the cranes.
Each participant carried air sampling equipment during the whole WD shift. Health examinations were performed immediately after the 8-h day shift, starting around 14:00. The participants were re-examined at about the same time of the day (around 12:00) on a DO, after at least 2-day off work.
At inclusion, 65 potential participants were registered. One subject was excluded from the study due to serious health problems discovered during the first testing. Three participants were excluded from the final analyses, because they had worked an extra, unscheduled, night shift on the night before the DO testing. In addition, two participants had sickness absence from their second testing. Thus, the study comprises 59 subjects.
Clinical examinations and Endo-PAT procedure
Demographics, height, and weight were recorded once. Blood pressure was measured once, in a seated position, to determine the required blood pressure cuff inflation during the Endo-PAT examination. A blood sample was drawn from the right arm at both examinations, before the Endo-PAT examination.
An Endo-PAT fingertip probe was placed on the index finger of each hand, and a blood pressure cuff was applied to the left overarm. The participants rested comfortably, while a physician interviewed them about current and previous work history, medical conditions, and potential confounders.
After a 10 min rest, the Endo-PAT procedure was initiated with fingertip probe inflation and a baseline recording of pulse amplitude for 7 min. Then, the blood pressure cuff was inflated until occlusion of the brachial artery (200 mm Hg or 60 mm Hg above systolic blood pressure, whichever was highest). The Endo-PAT recording continued with total occlusion of the pulse of the left index finger. After exactly 5 min, the blood pressure cuff was deflated, and the recording continued for another 7 min.
The Endo-PAT software automatically calculates the Reactive Hyperemia Index (RHI), which is the post-to-pre occlusion Peripheral Arterial Tone (PAT) signal ratio in the occluded arm, relative to the ratio in the control arm, corrected for baseline vascular tone of the occluded arm (Axtell et al. 2010). After the automatic analysis of the results, all recordings were manually checked. In four cases, the automated analyses gave no valid results because of artefacts in the registration. A manual setting of the time points for occlusion gave a valid analysis in two of these cases, whereas in the two other cases, no valid results could be achieved. When the automated analyses gave an occlusion duration outside the range 4.5–5.5 min (in nine cases), the borders of occlusion were set manually to 5 min. After reset, the Endo–PAT software recalculated RHI. The software also calculated the mean PAT signal amplitude in the baseline region of interest (BPA), both for the occluded and the control arm. Before statistical analyses, BPA was calculated as the mean between the two arms.
Blood sampling and analyses
Blood was collected from the cubital vein in 8.5 mL vacutainer tubes (BD Vacutainer, Belliver Industrial Estate, Plymouth UK) and separated by centrifugation at 1900g for 15 min. All samples were pipetted into 1.0 mL NUNC® polypropylene cryotubes (Sigma-Aldrich, St. Louis, Missouri, US) and stored in a freezer (− 18 °C) at the respective plants up to 5 weeks until transported to the National Institute of Occupational Health, Oslo, Norway (STAMI), where the samples were stored at − 80 °C until analysis.
Sample preparation and analyses of serum (S-) nicotine, S-cotinine, and S-caffeine were performed at STAMI, as previously described (Bast-Pettersen et al. 2017; Ellingsen et al. 2006). Detection limits (DL) were 31, 1.9, and 2.1 µg/L, for S-nicotine, S-cotinine, and S-caffeine, respectively. Before statistical analyses, results below DL were replaced by a value representing DL/2.
Exposure measurements
Full-shift (8 h) respirable aerosol samples were collected by personal sampling with 37-mm respirable cyclones (JS Holdings, Stevenage, UK), using in-house built PS103 model personal sampling pumps (STAMI), at an airflow rate of 2.2 L/min. Sioutas cascade impactors (SKC, Eighty Four, PA, USA), using Leland Legacy model high flow personal sampling pumps (SKC, Eighty Four, PA, USA) operated at an airflow rate of 9 L/min, were mounted in parallel with a second 37-mm respirable cyclone, for the collection of aerosol fractions. The Sioutas cascade impactor separates PM on the impactor stages from the top to the bottom in the following aerodynamic particle diameter ranges (in μm): 10–2.5, 2.5–1.0, 1.0–0.5, and 0.5–0.25. Particles below the 0.25 µm cut-point are collected on an after-filter. Further details on the aerosol sampling have been published (Berlinger et al. 2015).
The collected PM was determined gravimetrically by a six-place Sartorius Micro model MC5 balance (Sartorius AG, Göttingen, Germany) in a climate-controlled weighing room dedicated to low filter mass measurements, at relative humidity 40 ± 2% and temperature 20 ± 1°C. DL were below 0.01 mg for all kinds of substrates and filters used in the study (Berlinger et al. 2015).
The finest aerosol fraction measured in this study was PM250, and we chose to perform the analyses of associations between exposure concentrations and Endo-PAT results using this fraction as a surrogate for ultrafine particles. In addition, we performed the same analyses using the respirable aerosol fraction (PMResp), as this is a standard aerosol fraction used in many occupational studies.
Statistical analyses
Variables with skewed distribution were log-transformed. Mixed model linear regressions were performed to study the ratio between geometric mean (GM) RHI and GM BPA between WD and DO. A fixed-effects linear regression model (Gunasekara et al. 2014) was constructed, using the ΔRHI or ΔBPA as outcome variables, and the individual exposure to PM250 or PMResp as exposure variables. Stepwise multiple linear regression analysis (backward procedure) was used to identify covariates for inclusion in the mixed- and fixed-effects models. The level of significance for inclusion was set to 0.10 (two-tailed), and the following covariates were included: age (dichotomized by the median: 19–33 vs 34–64 years) ΔS-caffeine, and Δ(time since last meal). Because of missing data in the covariates, the number of participants in the final analyses was somewhat reduced. Due to differences in PM air concentrations and participant age between the manganese plants and the silicon metal plant, stratification by metal production (into Plant A and Plant B) was also considered relevant in the further analyses. The level of significance in the mixed- and fixed-effects analyses was set to 0.05 (two-tailed).
All data analyses were performed using STATA 15 (StataCorp LP, Texas, USA).