Abstract
Purpose
To compare visual and contrast sensitivity (CS) outcomes, reading skills, and spectacle independence in patients implanted with two models of trifocal intraocular lenses (IOLs) or an extended range of vision (ERV) IOL.
Methods
This non-randomized prospective series of cases included 120 eyes of 60 patients undergoing cataract surgery with bilateral implantation of three different IOLs: the ERV IOL Tecnis Symfony (40 eyes) (Abbott Medical Optics), the trifocal IOLs PanOptix IQ (40 eyes) (Alcon), and AT LISA tri 839MP (40 eyes) (Carl Zeiss Meditec). Visual results, photopic and mesopic CS, binocular reading skills (MNREAD charts), and patient satisfaction were evaluated 3 months after surgery.
Results
There was no statistically significant difference between groups regarding uncorrected and corrected distance visual acuity. The Tecnis Symfony IOL showed better mesopic intermediate visual outcomes than the two trifocal IOLs (p < 0.05 vs AT LISA). Under photopic conditions, AT LISA tri 839MP and PanOptix IQ showed better near visual outcomes compared with the ERV IOL (p < 0.05 to p < 0.001). The Tecnis Symfony IOL provided significantly better photopic and mesopic CS outcomes than the other IOL models (p < 0.001). Reading skills were not significantly different between the three IOL models (p > 0.05). Less patients implanted with the two trifocal IOLs required a near addition than patients with the ERV IOL.
Conclusions
All the tested IOLs provided good visual outcome, reading performance, and spectacle independence after cataract surgery. While trifocal IOLs gave better near visual acuity results, the ERV IOL provided better contrast sensitivity.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.References
Calladine D, Evans JR, Shah S, Leyland M (2012) Multifocal versus monofocal intraocular lenses after cataract extraction. Cochrane Database Syst Rev 12(9):CD003169
Alió JL, Plaza-Puche AB, Piñero DP, Amparo F, Rodríguez-Prats JL, Ayala MJ (2011) Quality of life evaluation after implantation of 2 multifocal intraocular lens models and a monofocal model. J Cataract Refract Surg 37:638–648
Percival SP (1989) Prospective study of the new diffractive bifocal intraocular lens. Eye 3:571–575
Shen Z, Lin Y, Zhu Y, Liu X, Yan J, Yao K (2017) Clinical comparison of patient outcomes following implantation of trifocal or bifocal intraocular lenses: a systematic review and meta-analysis. Sci Rep 7:45337
Mendicute J, Kapp A, Lévy P, Krommes G, Arias-Puente A, Tomalla M, Barraquer E, Rozot P, Bouchut P (2016) Evaluation of visual outcomes and patient satisfaction after implantation of a diffractive trifocal intraocular lens. J Cataract Refract Surg 42:203–210
Kohnen T (2015) First implantation of a diffractive quadrifocal (trifocal) intraocular lens. J Cataract Refract Surg 41:2330–2332
Pedrotti E, Bruni E, Bonacci E, Badalamenti R, Mastropasqua R, Marchini G (2016) Comparative analysis of the clinical outcomes with a monofocal and an extended range of vision intraocular lens. J Refract Surg 32:436–442
Hohberger B, Laemmer R, Adler W, Juenemann AGM, Horn FK (2007) Measuring contrast sensitivity in normal subjects with OPTEC® 6500: influence of age and glare. Graefes Arch Clin Exp Ophthalmol 245:1805–1814
Radner W (2016) Near vision examination in presbyopia patients: do we need good homologated near vision charts. Eye Vis 3:29
Cillino G, Casuccio A, Pasti M, Bono V, Mencucci R, Cillino S (2014) Working-age cataract patients: visual results, reading performance, and quality of life with three diffractive multifocal intraocular lenses. Ophthalmology 121:34–44
Martínez de la Casa JM, Carballo-Álvarez J, García-Bella J, Vázquez-Molini JM, Morales L, Sanz-Fernández JC, Polo V, García-Feijoo J (2016) Photopic and mesopic performance of 2 different trifocal diffractive intraocular lenses. Eur J Ophthalmol 27(1):26–30
Mojzis P, Majerova K, Hrckova L, Piñero DP (2015) Implantation of a diffractive trifocal intraocular lens: one-year follow-up. J Cataract Refract Surg 41:1623–1630
Kohnen T, Titke C, Böhm M (2016) Trifocal intraocular lens implantation to treat visual demands in various distances following lens removal. Am J Ophthalmol 161:71–77.e1
Bilbao-Calabuig R, Llovet-Rausell A, Ortega-Usobiaga J, Martínez-Del-Pozo M, Mayordomo-Cerdá F, Segura-Albentosa C, Baviera J, Llovet-Osuna F (2017) Visual outcomes following bilateral lmplantation of two diffractive trifocal intraocular lenses in 10 084 eyes. Am J Ophthalmol 179:55–66
Plaza-Puche AB, Alio JL (2016) Analysis of defocus curves of different modern multifocal intraocular lenses. Eur J Ophthalmol 26(5):412–417
Lee S, Choi M, Xu Z, Zhao Z, Alexander E, Liu Y (2016) Optical bench performance of a novel trifocal intraocular lens compared with a multifocal intraocular lens. Clin Ophthalmol 10:1031–1038
Gundersen KG, Potvin R (2017) Trifocal intraocular lenses: a comparison of the visual performance and quality of vision provided by two different lens designs. Clin Ophthalmol 11:1081–1087
Lawless M, Hodge C, Reich J, Levitz L, Bhatt UK, McAlinden C, Roberts K, Roberts TV (2017) Visual and refractive outcomes following implantation of a new trifocal intraocular lens. Eye Vis (Lond) 4:10
Kohnen T, Herzog M, Hemkeppler E, Schönbrunn S, De Lorenzo N, Petermann K, Böhm M (2017) Visual performance of a quadrifocal (trifocal) intraocular lens following removal of the crystalline lens. Am J Ophthalmol. https://doi.org/10.1016/j.ajo.2017.09.016
Cochener B, Concerto Study Group (2016) Clinical outcomes of a new extended range of vision intraocular lens: International Multicenter Concerto Study. J Cataract Refract Surg 42:1268–1275
Monaco G, Gari M, Di Censo F, Poscia A, Ruggi G, Scialdone A (2017) Visual performance after bilateral implantation of 2 new presbyopia-correcting intraocular lenses: trifocal versus extended range of vision. J Cataract Refract Surg 43(6):737–747
Ruiz-Mesa R, Abengózar-Vela A, Ruiz-Santos M (2017) A comparative study of the visual outcomes between a new trifocal and an extended depth of focus intraocular lens. Eur J Ophthalmol. https://doi.org/10.5301/ejo.5001029
Ruiz-Mesa R, Abengózar-Vela A, Aramburu A, Ruiz-Santos M (2017) Comparison of visual outcomes after bilateral implantation of extended range of vision and trifocal intraocular lenses. Eur J Ophthalmol 27(4):460–465
Artal P, Manzanera S, Piers P, Weeber H (2010) Visual effect of the combined correction of spherical and longitudinal chromatic aberrations. Opt Express 18:1637–1648
Domínguez-Vicent A, Esteve-Taboada JJ, Del Águila-Carrasco AJ, Ferrer-Blasco T, Montés-Micó R (2016) In vitro optical quality comparison between the Mini WELL Ready progressive multifocal and the TECNIS Symfony. Graefes Arch Clin Exp Ophthalmol 254:1387–1397
Gatinel D, Loicq J (2016) Clinically relevant optical properties of bifocal, trifocal, and extended depth of focus intraocular lenses. J Refract Surg 32:273–280
Esteve-Taboada JJ, Domínguez-Vicent A, Del Águila-Carrasco AJ, Ferrer-Blasco T, Montés-Micó R (2015) Effect of large apertures on the optical quality of three multifocal lenses. J Refract Surg 31:666–676
Chang DH, Rocha KM (2016) Intraocular lens optics and aberrations. Curr Opin Ophthalmol 27:298–303
Attia MS, Auffarth GU, Khoramnia R, Linz K, Kretz FT (2015) Near and intermediate reading performance of a diffractive trifocal intraocular lens using a reading desk. J Cataract Refract Surg 41:2707–2714
Rasp M, Bachernegg A, Seyeddain O, Ruckhofer J, Emesz M, Stoiber J, Grabner G, Dexl AK (2012) Bilateral reading performance of 4 multifocal intraocular lens models and a monofocal intraocular lens under bright lighting conditions. J Cataract Refract Surg 38:1950–1961
Alió JL, Plaza-Puche AB, Piñero DP, Amparo F, Jiménez R, Rodríguez-Prats JL, Javaloy J, Pongo V (2011) Optical analysis, reading performance, and quality-of-life evaluation after implantation of a diffractive multifocal intraocular lens. J Cataract Refract Surg 37:27–37
Whittaker SG, Lovie-Kitchin J (1993) Visual requirements for reading. Optom Vis Sci 70(1):54–65
Acknowledgements
We would also thank prof. Gianni Virgili for additional statistical assistance.
Funding
No public or private financial support was received for the study. The authors received third-party statistical support facilitated by Abbott.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Rights and permissions
About this article
Cite this article
Mencucci, R., Favuzza, E., Caporossi, O. et al. Comparative analysis of visual outcomes, reading skills, contrast sensitivity, and patient satisfaction with two models of trifocal diffractive intraocular lenses and an extended range of vision intraocular lens. Graefes Arch Clin Exp Ophthalmol 256, 1913–1922 (2018). https://doi.org/10.1007/s00417-018-4052-3
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00417-018-4052-3