Inter-method agreement in retinal blood vessels diameter analysis between Dynamic Vessel Analyzer and optical coherence tomography

  • Lucia Benatti
  • Federico Corvi
  • Livia Tomasso
  • Stefano Mercuri
  • Lea Querques
  • Fulvio Ricceri
  • Francesco Bandello
  • Giuseppe QuerquesEmail author
Retinal Disorders

Abstract

Purpose

To analyze the inter-methods agreement in arteriovenous ratio (AVR) evaluation between spectral-domain optical coherence tomography (SD-OCT) and Dynamic Vessel Analyzer (DVA).

Methods

Healthy volunteers underwent DVA and SD-OCT examination. AVR was measured by SD-OCT using the four external lines of the optic nerve head-centered 7-line cube and by DVA using an automated AVR estimation. The mean AVR was calculated, twice, separately by two independent readers for each tool.

Results

Twenty-two eyes of 11 healthy subjects (five women and six men, mean age 35) were included. AVR analysis by DVA showed high inter-observer agreement between reader 1 and 2, and high intra-observer agreement for both reader 1 and reader 2. With regard to AVR analysis on SD-OCT, we found high inter-observer agreement between reader 1 and 2, and low intra-observer agreement for reader 2 but high intra-observer agreement for reader 1. Overall, the mean AVR measured on SD-OCT turned out to be significantly higher than mean AVR measured through DVA (reader 1, 0.9023 ± 0.06 vs 0.8036 ± 0.08; p < 0.001, and reader 2, 0.9067 ± 0.06 vs 0.8083 ± 0.05; p= 0.003).

Conclusions

No inter-method agreement in AVR could be detected in the present study due to bias in measurements (shift between DVA and SD-OCT). We found significant difference in the two noninvasive methods for AVR measurement, with a tendency for SD-OCT to overestimate retinal vascular caliber in comparison to DVA. This may be useful for achieving greater accuracy in the evaluation of retinal vessel in ocular as well as systemic diseases.

Keywords

Arteriovenous ratio Dynamic vessel analyzer Inter-methods agreement Reproducibility Semi-automated analysis Spectral-domain optical coherence tomography 

Notes

Compliance with ethical standards

The authors have no proprietary, funding, or conflicts of interest to disclose.

Informed consent was obtained from all subjects in agreement with the Declaration of Helsinki for research involving human subjects.

No animals were involved in the research.

Funding

No funding was received for this research.

Informed consent

Informed consent was obtained from all individual participants included in the study.

Conflict of interest

All authors certify that they have no affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers’ bureaus; membership, employment, consultancies, stock ownership, or other equity interest; and expert testimony or patent-licensing arrangements), or non-financial interest (such as personal or professional relationships, affiliations, knowledge, or beliefs) in the subject matter or materials discussed in this manuscript.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.

References

  1. 1.
    Knudston JM, Lee EK, Hubbard LD et al (2003) Revised formulas for summarizing retinal vessel diameter. Curr Eye Res 27:143–149CrossRefGoogle Scholar
  2. 2.
    Klein R, Myers CE, Lee KE, Gangnon R, Klein BE (2012) Changes in retinal vessel diameter and incidence and progression of diabetic retinopathy. Arch Ophthalmol 130:749–755PubMedPubMedCentralGoogle Scholar
  3. 3.
    Wang JJ, Liew G, Klein R et al (2007) Retinal vessel diameter and cardiovascular mortality: pooled data analysis from two older populations. Eur Heart J 28:1984–1992CrossRefPubMedGoogle Scholar
  4. 4.
    Seoung-Bock L, Ki Bang U, Chul H (1998) Retinal vessel diameter in normal and primary open-angle glaucoma. Korean J Ophthalmol 12:51–59CrossRefGoogle Scholar
  5. 5.
    Wong TY, Knudtson MD, Klein R et al (2004) Computer-assisted measurement of retinal vessel diameter in the Beaver Dam Eye Study: methodology, correlation between eyes, and effect of refractive errors. Ophthalmology 111:1183–1190CrossRefPubMedGoogle Scholar
  6. 6.
    Goldenberg D, Shahar J, Loewenstein A et al (2013) Diameters of retinal blood vessels in a healthy cohort as measured by spectral domain optical coherence tomography. Retina 33:1888–1894CrossRefPubMedGoogle Scholar
  7. 7.
    Garhofer G, Bek T, Boehm AG et al (2010) Use of the retinal vessel analyzer in ocular blood flow research. Acta Ophthalmol 88:717–722CrossRefPubMedGoogle Scholar
  8. 8.
    Brueckmann A, Seeliger C, Schwefer M et al (2013) The arterio-venous ratio of retinal vessels in the first trimester as a predictor for preeclampsia. Pregnancy Hypertens 3:84–85Google Scholar
  9. 9.
    Terai N, Haustein M, Siegel A et al (2014) Diameter of retinal vessels in patients with diabetic macular edema is not altered by intravitreal ranibizumab (Lucentis). Retina 34:1466–1472CrossRefPubMedGoogle Scholar
  10. 10.
    Heitmar R, Kalitzeos AA, Patel SR et al (2015) Comparison on subjective and objective methods to determine the retinal arterio-venous ratio using fundus photography. J Ophthalm 8:252–257Google Scholar
  11. 11.
    Jürgens C, Ittermann T, Vötze H, Tost F (2014) Comparison of two non-mydriatic fundus cameras to obtain retinal arterio-venous ratio. Ophthalmic Epidemiol 21:333–338CrossRefPubMedGoogle Scholar
  12. 12.
    Shoukri MM (2010) Measures of interobserver agreement and reliability, 2nd edn. CRC Press, Boca RatonCrossRefGoogle Scholar
  13. 13.
    Wong T, Klein R, Sharett A et al (2003) The prevalence and risk factors of microvascular abnormalities in older people: the cardiovascular health study. Ophthalmology 110:658–666CrossRefPubMedGoogle Scholar
  14. 14.
    Mitchell P, Wang JJ, Wong TY, Sith W, Klein R, Leeder SR (2005) Retinal microvascular sings and risk of stroke and stroke mortality. Neurology 65:1005–1009CrossRefPubMedGoogle Scholar
  15. 15.
    Yang K, Zhan SY, Liang YB et al (2012) Association of dilated retinal arteriolar calibre with early age-related macular degeneration: the Handan Eye Study. Graefes Arch Clin Exp Ophthalmol 250:741–749CrossRefPubMedGoogle Scholar
  16. 16.
    Kawasaki R, Wang JJ, Rochtchina E, Lee AJ, Wong TY, Michell P (2013) Retinal vessel calibre is associated with the 10-year incidence of glaucoma: the Blue Mountains Eye Study. Ophthalmology 120:84–90CrossRefPubMedGoogle Scholar
  17. 17.
    Corvi F, Querques G, La Spina C, Lattanzio R, Bandello F (2015) Dynamic and static retinal vessel analyses in patients with macular edema secondary to retinal vein occlusion. Retina 35(10):2052–2059CrossRefPubMedGoogle Scholar

Copyright information

© Springer-Verlag Berlin Heidelberg 2017

Authors and Affiliations

  • Lucia Benatti
    • 1
  • Federico Corvi
    • 1
  • Livia Tomasso
    • 1
  • Stefano Mercuri
    • 1
  • Lea Querques
    • 1
  • Fulvio Ricceri
    • 1
  • Francesco Bandello
    • 1
  • Giuseppe Querques
    • 1
    Email author
  1. 1.Department of OphthalmologyUniversity Vita-Salute, IRCCS Ospedale San RaffaeleMilanItaly

Personalised recommendations