Can we trust intraocular pressure measurements in eyes with intracameral air?

  • Glaucoma
  • Published:
Graefe's Archive for Clinical and Experimental Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To evaluate the effect of intracameral air on intraocular pressure (IOP) measurements using Goldmann applanation tonometry (GAT) and applanation resonance tonometry (ART) in an in-vitro porcine eye model.

Methods

IOP was measured on thirteen freshly enucleated eyes at three reference pressures: 20, 30, and 40 mmHg. Six measurements/method were performed in a standardized order with GAT and ART respectively. Air was injected intracamerally in the same manner as during Descemet’s stripping endothelial keratoplasty (DSEK) and Descemet’s membrane endothelial keratoplasty (DMEK), and the measurements were repeated.

Results

Measured IOP increased significantly for both tonometry methods after air injection: 0.7 ± 2.1 mmHg for GAT and 10.6 ± 4.9 mmHg for ART. This difference was significant at each reference pressure for ART but not for GAT.

Conclusions

Although slightly affected, this study suggests that we can trust GAT IOP-measurements in eyes with intracameral air, such as after DSEK/DMEK operations. Ultrasound-based methods such as ART should not be used.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

  1. Price MO, Price FW (2007) Descemet’s stripping endothelial keratoplasty. Curr Opin Ophthalmol 18:290–294. doi:10.1097/ICU.0b013e3281a4775b00055735-200707000-00004

    Article  PubMed  Google Scholar 

  2. Dapena I, Moutsouris K, Droutsas K, Ham L, van Dijk K, Melles GR (2011) Standardized “no-touch” technique for descemet membrane endothelial keratoplasty. Arch Ophthalmol 129:88–94. doi:10.1001/archophthalmol.2010.334

    Article  PubMed  Google Scholar 

  3. Lee JS, Desai NR, Schmidt GW, Jun AS, Schein OD, Stark WJ, Eghrari AO, Gottsch JD (2009) Secondary angle closure caused by air migrating behind the pupil in descemet stripping endothelial keratoplasty. Cornea 28:652–656. doi:10.1097/ICO.0b013e318191b883

    Article  PubMed  CAS  Google Scholar 

  4. Lockington D, Luo X, Wang H, Hill NA, Ramaesh K (2012) Mathematical and computer simulation modelling of intracameral forces causing pupil block due to air bubble use in Descemet’s Stripping Endothelial Keratoplasty: the mechanics of iris buckling. Clin Experiment Ophthalmol 40:182–186. doi:10.1111/j.1442-9071.2011.02653.x

    Article  PubMed  Google Scholar 

  5. Naveiras M, Dirisamer M, Parker J, Ham L, van Dijk K, Dapena I, Melles GR (2012) Causes of glaucoma after descemet membrane endothelial keratoplasty. Am J Ophthalmol 153(958–966):e951. doi:10.1016/j.ajo.2011.10.003

    Google Scholar 

  6. Nieuwendaal CP, van der Meulen IJ, Lapid-Gortzak R, Mourits MP (2013) Intraocular pressure after descemet stripping endothelial keratoplasty (DSEK). Int Ophthalmol 33:147–151. doi:10.1007/s10792-012-9665-7

    Article  PubMed  Google Scholar 

  7. Johannesson G, Hallberg P, Eklund A, Linden C (2008) Pascal, ICare and Goldmann applanation tonometry—a comparative study. Acta Ophthalmol 86:614–621

    Article  PubMed  Google Scholar 

  8. Kniestedt C, Punjabi O, Lin S, Stamper RL (2008) Tonometry through the ages. Surv Ophthalmol 53:568–591. doi:10.1016/j.survophthal.2008.08.024

    Article  PubMed  Google Scholar 

  9. Muller L, Kohnen T (2009) Influence of residual corneal bed thickness after myopic LASIK on intraocular pressure measurements : Goldmann applanation tonometry and dynamic contour tonometry. Ophthalmologe 106:21–28

    Article  PubMed  CAS  Google Scholar 

  10. Whitacre MM, Stein R (1993) Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 38:1–30

    Article  PubMed  CAS  Google Scholar 

  11. Eklund A, Hallberg P, Linden C, Lindahl OA (2003) An applanation resonator sensor for measuring intraocular pressure using combined continuous force and area measurement. Invest Ophthalmol Vis Sci 44:3017–3024

    Article  PubMed  Google Scholar 

  12. Hallberg P, Eklund A, Santala K, Koskela T, Lindahl O, Linden C (2006) Underestimation of intraocular pressure after photorefractive keratectomy: a biomechanical analysis. Med Biol Eng Comput 44:609–618. doi:10.1007/s11517-006-0093-6

    Article  PubMed  Google Scholar 

  13. Hallberg P, Santala K, Linden C, Lindahl OA, Eklund A (2006) Comparison of Goldmann applanation and applanation resonance tonometry in a biomicroscope-based in vitro porcine eye model. J Med Eng Technol 30:345–352

    Article  PubMed  CAS  Google Scholar 

  14. Eisenberg DL, Sherman BG, McKeown CA, Schuman JS (1998) Tonometry in adults and children. A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo. Ophthalmology 105:1173–1181. doi:10.1016/S0161-6420(98)97016-6

    Article  PubMed  CAS  Google Scholar 

  15. Goldmann H (1957) Applanation tonometry. Josiah Macy, Jr. Foundation, New York

  16. Hallberg P, Eklund A, Backlund T, Linden C (2007) Clinical evaluation of applanation resonance tonometry: a comparison with Goldmann applanation tonometry. J Glaucoma 16:88–93

    Article  PubMed  Google Scholar 

  17. Johannesson G, Hallberg P, Eklund A, Koskela T, Linden C (2012) Change in intraocular pressure measurement after myopic LASEK: a study evaluating Goldmann, Pascal and applanation resonance tonometry. J Glaucoma 21:255–259. doi:10.1097/IJG.0b013e31820719c8

    Article  PubMed  Google Scholar 

  18. Szabo T (2004) Diagnostic ultrasound imaging — inside out. Elsevier Academic Press, New York

    Google Scholar 

  19. Doughty MJ, Zaman ML (2000) Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 44:367–408, DOI S0039-6257(00)00110-7

    Article  PubMed  CAS  Google Scholar 

  20. Jonsson M, Markstrom K, Behndig A (2006) Slit-scan tomography evaluation of the anterior chamber and corneal configurations at different ages. Acta Ophthalmol Scand 84:116–120. doi:10.1111/j.1600-0420.2005.00577.x

    Article  PubMed  Google Scholar 

  21. Kim JM, Park KH, Kim SH, Kang JH, Cho SW (2010) The relationship between the cornea and the optic disc. Eye (Lond) 24:1653–1657. doi:10.1038/eye.2010.98

    Article  CAS  Google Scholar 

Download references

Acknowledgments

The authors thank technical engineers Göran Mannberg and Kenji Claesson for valuable technical assistance and contribution. This study was supported by grants from the Kronprinsessan Margaretas Arbetsnämnd and the Insamlingsstiftelsen för medicinsk forskning at Umeå University

Conflicts of interest

The financial disclosures of the authors are the following: Gauti Jóhannesson – none; Christina Lindén – none; Anders Behndig – none; Anders Eklund – equity owner (CMTF Futv AB Sweden); Per Hallberg – equity owner (CMTF Futv AB Sweden). The sponsor or funding organization had no role in the design or conduct of this research.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Gauti Jóhannesson.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Jóhannesson, G., Lindén, C., Eklund, A. et al. Can we trust intraocular pressure measurements in eyes with intracameral air?. Graefes Arch Clin Exp Ophthalmol 252, 1607–1610 (2014). https://doi.org/10.1007/s00417-014-2720-5

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00417-014-2720-5

Keywords

Navigation