Abstract
Purpose
Using a new Goldmann contact lens associated ophthalmodynamometric device, it was the purpose of the present study to determine the central retinal vein collapse pressure in eyes with retinal vein occlusions or retinal venous stasis.
Methods
The prospective clinical non-interventional comparative study included 19 patients with central retinal vein occlusion (n=8), branch retinal vein occlusion (n=4), or retinal venous stasis (n=7) and 42 subjects of a control group. With topical anesthesia, a Goldmann contact lens fitted with a pressure sensor was put onto the cornea. Pressure was exerted on the globe by pressing the contact lens, and the pressure value at the time when the central retinal vein started pulsating was noted.
Results
Central retinal vein collapse pressure measured 103.6±25.4 arbitrary units (AU) in eyes with central retinal vein occlusion what was significantly higher than in the eyes with retinal venous stasis (58.1±37.5 AU; p=0.02) and the eyes with branch retinal vein occlusion (43.8±25.5 AU; p=0.004). In the latter two groups, the measurements of the central retinal vein collapse pressure were significantly (p<0.001) higher than the measurements in the eyes of the control group (4.2±7.8 AU).
Conclusion
As measured by a new ophthalmodynamometer with direct biomicroscopic visualization of the central retinal vessels during examination, central retinal vein collapse pressure is significantly higher in eyes with central retinal vein occlusion, followed by eyes with branch retinal vein occlusion, eyes with retinal venous stasis and, finally, normal eyes. These findings may have diagnostic and therapeutic implications.
This is a preview of subscription content,
to check access.Similar content being viewed by others
References
Bettelheim H (1969) [The clinical significance of ophthalmodynamometry and ophthalmodynamography] Klin Monatsbl Augenheilkd 155:769–791
Draeger J, Rumberger E, Hechler B (1999) Intracranial pressure in microgravity conditions: non-invasive assessment by ophthalmodynamometry. Aviat Space Environ Med 70:1227–1229
Entenmann B, Robert YC, Pirani P, Kanngiesser H, Dekker PW (1997) Contact lens tonometry—application in humans. Invest Ophthalmol Vis Sci 38:2447–2451
Firsching R, Schutze M, Motschmann M, Behrens-Baumann W (2000) Venous opthalmodynamometry: a noninvasive method for assessment of intracranial pressure. J Neurosurg 93:33–36
Galin MA, Baras I, Dodick JM (1969) Semiautomated suction ophthalmodynamometry. Am J Ophthalmol 68:237–240
Hedges TR, Weinstein JD, Kassell NF, et al (1965) Correlation of ophthalmodynamometry with ophthalmic artery pressure in the rhesus monkey. Am J Ophthalmol 60:1098–1101
Hitchings RA, Spaeth GL (1976) Chronic retinal vein occlusion in glaucoma. Br J Ophthalmol 60:694–699
Jessen K, Weigelin EH-K (1980) Müller's ophthalmodynamometer—standardization and calibration. Klin Monatsbl Augenheilkd 176:993–997
Jonas JB (2003) Reproducibility of ophthalmodynamometric measurements of the central retinal artery and vein collapse pressure. Br J Ophthalmol (in press)
Jonas JB (2003) Central retinal artery and vein pressure in patients with chronic open-angle glaucoma. Br J Ophthalmol (in press)
Jonas JB, Harder B (2002) Ophthalmodynamometric estimation of cerebrospinal fluid pressure in pseudotumor cerebri. Br J Ophthalmol 87:361–362
Jonas JB, Niessen A (2002) Ophthalmodynamometric diagnosis of unilateral ischemic ophthalmopathy. Am J Ophthalmol 134:911–912
Krieglstein GK, da Silva FA (1979) Comparative measurements of the ophthalmic arterial pressure using the Mikuni dynamometer and the Stepanik arteriotonograph. Albrecht Von Graefes Arch Klin Exp Ophthalmol 212:77–91
Machleder HI, Barker WF (1977) Noninvasive methods for evaluation of extracranial cerebrovascular disease. A comparison. Arch Surg 112:944–946
Meyer-Schwickerath R, Kleinwachter T, Firsching R, Papenfuss HD (1995) Central retinal venous outflow pressure. Graefes Arch Clin Exp Ophthalmol 233:783–788
Morgan WH, Yu DY, Cooper RL, et al (1997) Retinal artery and vein pressures in the dog and their relationship to aortic, intraocular and cerebrospinal fluid pressures. Microvasc Res 53:211–221
Van der Werff TJ (1972) The pressure measured in ophthalmodynamometry. Arch Ophthalmol 87:290–292
Wunsh SE (1969) Ophthalmodynamometry. N Engl J Med 281:446
Author information
Authors and Affiliations
Corresponding author
Additional information
Proprietary interest: none
Rights and permissions
About this article
Cite this article
Jonas, J.B. Ophthalmodynamometric assessment of the central retinal vein collapse pressure in eyes with retinal vein stasis or occlusion. Graefe's Arch Clin Exp Ophthalmol 241, 367–370 (2003). https://doi.org/10.1007/s00417-003-0643-7