Abstract
Methods
Based on the results from 42 IVF patients on the day of oocyte aspiration, our study tested the hypothesis that 3-D Doppler ultrasound of the endometrium supported by a contrast agent possesses advantages over a conventional 3-D Doppler ultrasound examination. We also determined whether the subendometrial blood flow indices measured under continuous infusion of Levovist (Schering, Berlin, Germany) were an indicator of a successful implantation.
Results
It was demonstrated that, after the administration of Levovist, all subendometrial blood flow parameters (e.g., vascularization index [VI: 8.82±16.64], flow index [FI: 35.67±9.33], and vascularization flow index [VFI: 3.99±8.80]) were significantly higher than without the contrast agent (p<0.001; VI: 3.09±7.11; FI: 29.13±14.26; VFI: 1.35±3.31). However, no significant differences between blood flow with contrast agent and pregnancy rate were found.
Conclusion
In summary, quantitative 3-D Doppler ultrasound employing Levovist (Schering, Berlin, Germany) is significantly more sensitive when compared to conventional 3-D Doppler examination. For the first time this was demonstrated on 42 IVF patients on the day of oocyte aspiration. However, no correlation with outcome of the IVF-treatment could be found.
Similar content being viewed by others
References
Ascenti G, Zimbaro G, Mazziotti S, Gaeta M, Lamberto S, Scribano E (2001) Contrast-enhanced power Doppler US in the diagnosis of renal pseudotumors. Eur Radiol 11:2496–2499
Ascenti G, Zimbaro G, Mazziotti S, Gaeta M, Settineri N, Scribano E (2001) Usefulness of power Doppler and contrast-enhanced sonography in the differentiation of hyperechoic renal masses. Abdom Imaging 26:654–660
Ascenti G, Zimbaro G, Mazziotti S, Visalli C, Lamberto S, Scribano E, Gaeta M (2001) Intrahepatic portal vein aneurysm: three-dimensional power Doppler demonstration in four cases. Abdom Imaging 26:520–523
Aytoz A, Ubaldi F, Tournaye H, Nagy ZP, Van Steirteghem A, Devroey P (1997) The predictive value of uterine artery blood flow measurements for uterine receptivity in an intracytoplasmic sperm injection program. Fertil Steril 68:935–937
Bauer A, Bogdahn U, Haase A, Schlief R (1998) [3-dimensional echo-enhanced transcranial Doppler ultrasound diagnosis]. Radiologe 38:394–398
Becher H, Tiemann K, Schlief R, Luderitz B, Nanda NC (1997) Harmonic power Doppler contrast echocardiography: preliminary clinical results. Echocardiography 14:637
Bloechle M, Schreiner T, Kuchler I, Schurenkampe P, Lisse K (1997) Color Doppler assessment of ascendant uterine artery perfusion in an in-vitro fertilization-embryo transfer programme after pituitary desensitization and ovarian stimulation with human recombinant follicle stimulating hormone. Hum Reprod 12:1772–1777
Bogers HA, Sedelaar JP, Beerlage HP, de la Rosette JJ, Debruyne FM, Wijkstra H, Aarnink RG (1999) Contrast-enhanced three-dimensional power Doppler angiography of the human prostate: correlation with biopsy outcome. Urology 54:97–104
Chien LW, Au HK, Chen PL, Xiao J, Tzeng CR (2002) Assessment of uterine receptivity by the endometrial-subendometrial blood flow distribution pattern in women undergoing in vitro fertilization-embryo transfer. Fertil Steril 78:245–251
Contart P, Baruffi RL, Coelho J, Mauri AL, Petersen C, Franco JG Jr (2000) Power Doppler endometrial evaluation as a method for the prognosis of embryo implantation in an ICSI program. J Assist Reprod Genet 17:329–334
Delcker A, Haussermann P, Weimar C (1999) Effect of echo contrast media on the visualization of transverse sinus thrombosis with transcranial 3-D duplex sonography. Ultrasound Med Biol 25:1063–1068
Doria AS, Guarniero R, Molnar LJ, Modena M, Cunha FG, de Godoy Jr RM, Cerri GG (2000) Three-dimensional (3D) contrast-enhanced power Doppler imaging in Legg-Calve-Perthes disease. Pediatr Radiol 30:871–874
Droste DW, Jurgens R, Weber S, Tietje R, Ringelstein EB (2000) Benefit of echocontrast-enhanced transcranial color-coded duplex ultrasound in the assessment of intracranial collateral pathways. Stroke 31:920–923
Droste DW, Kaps M, Navabi DG, Ringelstein EB (2000) Ultrasound contrast enhancing agents in neurosonology: principles, methods, future possibilities. Acta Neurol Scand 102:1–10
Eichler C, Krampl E, Reichel V, Zegermacher G, Obruca A, Strohmer H, Feldner-Busztin M, Feichtinger W (1993) The relevance of endometrial thickness and echo patterns for the success of in vitro fertilization evaluated in 148 patients. J Assist Reprod Genet 10:223–227
Feichtinger W (1998) Follicle aspiration with interactive three-dimensional digital imaging (Voluson): a step toward real-time puncturing under three-dimensional ultrasound control. Fertil Steril 70:374–377
Ferrara K, DeAngelis G (1997) Color flow mapping. Ultrasound Med Biol 23:321–345
Forsberg F, Rawool NM, Merton DA, Liu JB, Goldberg BB (2002) Contrast enhanced vascular three-dimensional ultrasound imaging. Ultrasonics 40:117–122
Friedler S, Schenker JG, Herman A, Lewin A (1996) The role of ultrasonography in the evaluation of endometrial receptivity following assisted reproductive treatments: a critical review. Hum Reprod Update 2:323–335
Kupesic S, Kurjak A (2000) Contrast-enhanced, three-dimensional power Doppler sonography for differentiation of adnexal masses. Obstet Gynecol 96:452–458
Kupesic S, Bekavac I, Bjelos D, Kurjak A (2001) Assessment of endometrial receptivity by transvaginal color Doppler and three-dimensional power Doppler ultrasonography in patients undergoing in vitro fertilization procedures. J Ultrasound Med 20:125–134
Kurjak A, Kupesic S, Sparac V, Bekavac I (2001) Preoperative evaluation of pelvic tumors by Doppler and three-dimensional sonography. J Ultrasound Med 20:829–840
Kyei-Mensah A, Maconochie N, Zaidi J, Pittrof R, Campbell S, Tan SL (1996) Transvaginal three-dimensional ultrasound: reproducibility of ovarian and endometrial volume measurements. Fertil Steril 66:718–722
Kyei-Mensah A, Zaidi J, Pittrof R, Shaker A, Campbell S, Tan SL (1996) Transvaginal three-dimensional ultrasound: accuracy of follicular volume measurements. Fertil Steril 65:371–376
Pan H A, Wu MH, Cheng YC, Li CH, Chang FM (2002) Quantification of Doppler signal in polycystic ovary syndrome using three-dimensional power Doppler ultrasonography: a possible new marker for diagnosis. Hum Reprod 17:201–206
Pellicer A, Ardiles G, Neuspiller F, Remohi J, Simon C, Bonilla-Musoles F (1998) Evaluation of the ovarian reserve in young low responders with normal basal levels of follicle-stimulating hormone using three-dimensional ultrasonography. Fertil Steril 70:671–675
Salle B, Bied-Damon V, Benchaib M, Desperes S, Gaucherand P, Rudigoz RC (1998) Preliminary report of an ultrasonography and color Doppler uterine score to predict uterine receptivity in an in-vitro fertilization programme. Hum Reprod 13:1669–1673
Schild RL, Indefrei D, Eschweiler S, Van der Ven H, Fimmers R, Hansmann M (1999) Three-dimensional endometrial volume calculation and pregnancy rate in an in-vitro fertilization programme. Hum Reprod 14:1255–1258
Schild RL, Holthaus S, d'Alquen J, Fimmers R, Dorn C, van Der Ven H, Hansmann M (2000) Quantitative assessment of subendometrial blood flow by three-dimensional ultrasound is an important predictive factor of implantation in an in-vitro fertilization programme. Hum Reprod 15:89–94
Schild RL, Knobloch C, Dorn C, Fimmers R, van der Ven H, Hansmann M (2001) Endometrial receptivity in an in vitro fertilization program as assessed by spiral artery blood flow, endometrial thickness, endometrial volume, and uterine artery blood flow. Fertil Steril 75:361–366
Schild RL, Knobloch C, Dorn C, Fimmers R, van der Ven H, Hansmann M (2001) The role of ovarian volume in an in vitro fertilization programme as assessed by 3D ultrasound. Arch Gynecol Obstet 265:67–72
Schlief R (1991) Ultrasound contrast agents. Curr Opin Radiol 3:198–207
Schlief R, Deichert U (1991) Hysterosalpingo-contrast sonography of the uterus and fallopian tubes: results of a clinical trial of a new contrast medium in 120 patients. Radiology 178:213–215
Schwartz LB, Chiu AS, Courtney M, Krey L, Schmidt-Sarosi C (1997) The embryo versus endometrium controversy revisited as it relates to predicting pregnancy outcome in in-vitro fertilization-embryo transfer cycles. Hum Reprod 12:45–50
Sedelaar JP, Aarnink RG, van Leenders GJ, Beerlage HP, Debruyne FM, Wijkstra H, de La Rosette JJ (2000) The application of three-dimensional contrast-enhanced ultrasound to measure volume of affected tissue after HIFU treatment for localized prostate cancer. Eur Urol 37:59–68
Sparac V, Kupesic S, Kurjak A (2000) What do contrast media add to three-dimensional power Doppler evaluation of adnexal masses? Croat Med J 41:257–261
Ugolini P, Delouche A, Herment A, Diebold B (2001) In vitro quantification of flow using continuous infusion of Levovist and pairs of harmonic power Doppler images. Ultrasound Med Biol 27:637–642
Wu CH, Shih JC, Chang YL, Lee SY, Hsieh FJ (1998) Two-dimensional and three-dimensional power Doppler sonographic classification of vascular patterns in cervical lymphadenopathies. J Ultrasound Med 17:459–464
Yao J, Takeuchi M, Teupe C, Sheahan M, Connolly R, Walovitch RC, Fetterman RC, Church CC, Udelson JE, Pandian NG (2002) Evaluation of a new ultrasound contrast agent (AI-700) using two-dimensional and three-dimensional imaging during acute ischemia. J Am Soc Echocardiogr 15:686–694
Yuval Y, Lipitz S, Dor J, Achiron R (1999) The relationships between endometrial thickness, and blood flow and pregnancy rates in in-vitro fertilization. Hum Reprod 14:1067–1071
Zaidi J, Campbell S, Pittrof R, Tan SL (1995) Endometrial thickness, morphology, vascular penetration and velocimetry in predicting implantation in an in vitro fertilization program. Ultrasound Obstet Gynecol 6:191–198
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Dorn, C., Reinsberg, J., Willeke, C. et al. Three-dimensional power Doppler ultrasound of the subendometrial blood flow under the administration of a contrast agent (Levovist). Arch Gynecol Obstet 270, 94–98 (2004). https://doi.org/10.1007/s00404-003-0491-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00404-003-0491-z