Three-dimensional power Doppler ultrasound of the subendometrial blood flow under the administration of a contrast agent (Levovist)

  • Original Article
  • Published:
Archives of Gynecology and Obstetrics Aims and scope Submit manuscript

Abstract

Methods

Based on the results from 42 IVF patients on the day of oocyte aspiration, our study tested the hypothesis that 3-D Doppler ultrasound of the endometrium supported by a contrast agent possesses advantages over a conventional 3-D Doppler ultrasound examination. We also determined whether the subendometrial blood flow indices measured under continuous infusion of Levovist (Schering, Berlin, Germany) were an indicator of a successful implantation.

Results

It was demonstrated that, after the administration of Levovist, all subendometrial blood flow parameters (e.g., vascularization index [VI: 8.82±16.64], flow index [FI: 35.67±9.33], and vascularization flow index [VFI: 3.99±8.80]) were significantly higher than without the contrast agent (p<0.001; VI: 3.09±7.11; FI: 29.13±14.26; VFI: 1.35±3.31). However, no significant differences between blood flow with contrast agent and pregnancy rate were found.

Conclusion

In summary, quantitative 3-D Doppler ultrasound employing Levovist (Schering, Berlin, Germany) is significantly more sensitive when compared to conventional 3-D Doppler examination. For the first time this was demonstrated on 42 IVF patients on the day of oocyte aspiration. However, no correlation with outcome of the IVF-treatment could be found.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Ascenti G, Zimbaro G, Mazziotti S, Gaeta M, Lamberto S, Scribano E (2001) Contrast-enhanced power Doppler US in the diagnosis of renal pseudotumors. Eur Radiol 11:2496–2499

    Article  CAS  PubMed  Google Scholar 

  2. Ascenti G, Zimbaro G, Mazziotti S, Gaeta M, Settineri N, Scribano E (2001) Usefulness of power Doppler and contrast-enhanced sonography in the differentiation of hyperechoic renal masses. Abdom Imaging 26:654–660

    Article  CAS  PubMed  Google Scholar 

  3. Ascenti G, Zimbaro G, Mazziotti S, Visalli C, Lamberto S, Scribano E, Gaeta M (2001) Intrahepatic portal vein aneurysm: three-dimensional power Doppler demonstration in four cases. Abdom Imaging 26:520–523

    Article  CAS  PubMed  Google Scholar 

  4. Aytoz A, Ubaldi F, Tournaye H, Nagy ZP, Van Steirteghem A, Devroey P (1997) The predictive value of uterine artery blood flow measurements for uterine receptivity in an intracytoplasmic sperm injection program. Fertil Steril 68:935–937

    Article  CAS  PubMed  Google Scholar 

  5. Bauer A, Bogdahn U, Haase A, Schlief R (1998) [3-dimensional echo-enhanced transcranial Doppler ultrasound diagnosis]. Radiologe 38:394–398

    Article  CAS  PubMed  Google Scholar 

  6. Becher H, Tiemann K, Schlief R, Luderitz B, Nanda NC (1997) Harmonic power Doppler contrast echocardiography: preliminary clinical results. Echocardiography 14:637

    PubMed  Google Scholar 

  7. Bloechle M, Schreiner T, Kuchler I, Schurenkampe P, Lisse K (1997) Color Doppler assessment of ascendant uterine artery perfusion in an in-vitro fertilization-embryo transfer programme after pituitary desensitization and ovarian stimulation with human recombinant follicle stimulating hormone. Hum Reprod 12:1772–1777

    Article  CAS  PubMed  Google Scholar 

  8. Bogers HA, Sedelaar JP, Beerlage HP, de la Rosette JJ, Debruyne FM, Wijkstra H, Aarnink RG (1999) Contrast-enhanced three-dimensional power Doppler angiography of the human prostate: correlation with biopsy outcome. Urology 54:97–104

    Article  PubMed  Google Scholar 

  9. Chien LW, Au HK, Chen PL, Xiao J, Tzeng CR (2002) Assessment of uterine receptivity by the endometrial-subendometrial blood flow distribution pattern in women undergoing in vitro fertilization-embryo transfer. Fertil Steril 78:245–251

    Article  PubMed  Google Scholar 

  10. Contart P, Baruffi RL, Coelho J, Mauri AL, Petersen C, Franco JG Jr (2000) Power Doppler endometrial evaluation as a method for the prognosis of embryo implantation in an ICSI program. J Assist Reprod Genet 17:329–334

    Article  CAS  PubMed  Google Scholar 

  11. Delcker A, Haussermann P, Weimar C (1999) Effect of echo contrast media on the visualization of transverse sinus thrombosis with transcranial 3-D duplex sonography. Ultrasound Med Biol 25:1063–1068

    Article  CAS  PubMed  Google Scholar 

  12. Doria AS, Guarniero R, Molnar LJ, Modena M, Cunha FG, de Godoy Jr RM, Cerri GG (2000) Three-dimensional (3D) contrast-enhanced power Doppler imaging in Legg-Calve-Perthes disease. Pediatr Radiol 30:871–874

    Article  CAS  PubMed  Google Scholar 

  13. Droste DW, Jurgens R, Weber S, Tietje R, Ringelstein EB (2000) Benefit of echocontrast-enhanced transcranial color-coded duplex ultrasound in the assessment of intracranial collateral pathways. Stroke 31:920–923

    CAS  PubMed  Google Scholar 

  14. Droste DW, Kaps M, Navabi DG, Ringelstein EB (2000) Ultrasound contrast enhancing agents in neurosonology: principles, methods, future possibilities. Acta Neurol Scand 102:1–10

    Article  CAS  PubMed  Google Scholar 

  15. Eichler C, Krampl E, Reichel V, Zegermacher G, Obruca A, Strohmer H, Feldner-Busztin M, Feichtinger W (1993) The relevance of endometrial thickness and echo patterns for the success of in vitro fertilization evaluated in 148 patients. J Assist Reprod Genet 10:223–227

    CAS  PubMed  Google Scholar 

  16. Feichtinger W (1998) Follicle aspiration with interactive three-dimensional digital imaging (Voluson): a step toward real-time puncturing under three-dimensional ultrasound control. Fertil Steril 70:374–377

    Google Scholar 

  17. Ferrara K, DeAngelis G (1997) Color flow mapping. Ultrasound Med Biol 23:321–345

    Article  CAS  PubMed  Google Scholar 

  18. Forsberg F, Rawool NM, Merton DA, Liu JB, Goldberg BB (2002) Contrast enhanced vascular three-dimensional ultrasound imaging. Ultrasonics 40:117–122

    Article  CAS  PubMed  Google Scholar 

  19. Friedler S, Schenker JG, Herman A, Lewin A (1996) The role of ultrasonography in the evaluation of endometrial receptivity following assisted reproductive treatments: a critical review. Hum Reprod Update 2:323–335

    CAS  PubMed  Google Scholar 

  20. Kupesic S, Kurjak A (2000) Contrast-enhanced, three-dimensional power Doppler sonography for differentiation of adnexal masses. Obstet Gynecol 96:452–458

    Article  CAS  PubMed  Google Scholar 

  21. Kupesic S, Bekavac I, Bjelos D, Kurjak A (2001) Assessment of endometrial receptivity by transvaginal color Doppler and three-dimensional power Doppler ultrasonography in patients undergoing in vitro fertilization procedures. J Ultrasound Med 20:125–134

    CAS  PubMed  Google Scholar 

  22. Kurjak A, Kupesic S, Sparac V, Bekavac I (2001) Preoperative evaluation of pelvic tumors by Doppler and three-dimensional sonography. J Ultrasound Med 20:829–840

    CAS  PubMed  Google Scholar 

  23. Kyei-Mensah A, Maconochie N, Zaidi J, Pittrof R, Campbell S, Tan SL (1996) Transvaginal three-dimensional ultrasound: reproducibility of ovarian and endometrial volume measurements. Fertil Steril 66:718–722

    Google Scholar 

  24. Kyei-Mensah A, Zaidi J, Pittrof R, Shaker A, Campbell S, Tan SL (1996) Transvaginal three-dimensional ultrasound: accuracy of follicular volume measurements. Fertil Steril 65:371–376

    CAS  PubMed  Google Scholar 

  25. Pan H A, Wu MH, Cheng YC, Li CH, Chang FM (2002) Quantification of Doppler signal in polycystic ovary syndrome using three-dimensional power Doppler ultrasonography: a possible new marker for diagnosis. Hum Reprod 17:201–206

    Article  PubMed  Google Scholar 

  26. Pellicer A, Ardiles G, Neuspiller F, Remohi J, Simon C, Bonilla-Musoles F (1998) Evaluation of the ovarian reserve in young low responders with normal basal levels of follicle-stimulating hormone using three-dimensional ultrasonography. Fertil Steril 70:671–675

    Google Scholar 

  27. Salle B, Bied-Damon V, Benchaib M, Desperes S, Gaucherand P, Rudigoz RC (1998) Preliminary report of an ultrasonography and color Doppler uterine score to predict uterine receptivity in an in-vitro fertilization programme. Hum Reprod 13:1669–1673

    Article  CAS  PubMed  Google Scholar 

  28. Schild RL, Indefrei D, Eschweiler S, Van der Ven H, Fimmers R, Hansmann M (1999) Three-dimensional endometrial volume calculation and pregnancy rate in an in-vitro fertilization programme. Hum Reprod 14:1255–1258

    Article  CAS  PubMed  Google Scholar 

  29. Schild RL, Holthaus S, d'Alquen J, Fimmers R, Dorn C, van Der Ven H, Hansmann M (2000) Quantitative assessment of subendometrial blood flow by three-dimensional ultrasound is an important predictive factor of implantation in an in-vitro fertilization programme. Hum Reprod 15:89–94

    Article  CAS  PubMed  Google Scholar 

  30. Schild RL, Knobloch C, Dorn C, Fimmers R, van der Ven H, Hansmann M (2001) Endometrial receptivity in an in vitro fertilization program as assessed by spiral artery blood flow, endometrial thickness, endometrial volume, and uterine artery blood flow. Fertil Steril 75:361–366

    Article  CAS  PubMed  Google Scholar 

  31. Schild RL, Knobloch C, Dorn C, Fimmers R, van der Ven H, Hansmann M (2001) The role of ovarian volume in an in vitro fertilization programme as assessed by 3D ultrasound. Arch Gynecol Obstet 265:67–72

    Article  CAS  PubMed  Google Scholar 

  32. Schlief R (1991) Ultrasound contrast agents. Curr Opin Radiol 3:198–207

    CAS  PubMed  Google Scholar 

  33. Schlief R, Deichert U (1991) Hysterosalpingo-contrast sonography of the uterus and fallopian tubes: results of a clinical trial of a new contrast medium in 120 patients. Radiology 178:213–215

    CAS  PubMed  Google Scholar 

  34. Schwartz LB, Chiu AS, Courtney M, Krey L, Schmidt-Sarosi C (1997) The embryo versus endometrium controversy revisited as it relates to predicting pregnancy outcome in in-vitro fertilization-embryo transfer cycles. Hum Reprod 12:45–50

    Article  CAS  PubMed  Google Scholar 

  35. Sedelaar JP, Aarnink RG, van Leenders GJ, Beerlage HP, Debruyne FM, Wijkstra H, de La Rosette JJ (2000) The application of three-dimensional contrast-enhanced ultrasound to measure volume of affected tissue after HIFU treatment for localized prostate cancer. Eur Urol 37:59–68

    Article  Google Scholar 

  36. Sparac V, Kupesic S, Kurjak A (2000) What do contrast media add to three-dimensional power Doppler evaluation of adnexal masses? Croat Med J 41:257–261

    Google Scholar 

  37. Ugolini P, Delouche A, Herment A, Diebold B (2001) In vitro quantification of flow using continuous infusion of Levovist and pairs of harmonic power Doppler images. Ultrasound Med Biol 27:637–642

    Article  CAS  PubMed  Google Scholar 

  38. Wu CH, Shih JC, Chang YL, Lee SY, Hsieh FJ (1998) Two-dimensional and three-dimensional power Doppler sonographic classification of vascular patterns in cervical lymphadenopathies. J Ultrasound Med 17:459–464

    CAS  PubMed  Google Scholar 

  39. Yao J, Takeuchi M, Teupe C, Sheahan M, Connolly R, Walovitch RC, Fetterman RC, Church CC, Udelson JE, Pandian NG (2002) Evaluation of a new ultrasound contrast agent (AI-700) using two-dimensional and three-dimensional imaging during acute ischemia. J Am Soc Echocardiogr 15:686–694

    Article  PubMed  Google Scholar 

  40. Yuval Y, Lipitz S, Dor J, Achiron R (1999) The relationships between endometrial thickness, and blood flow and pregnancy rates in in-vitro fertilization. Hum Reprod 14:1067–1071

    Article  CAS  PubMed  Google Scholar 

  41. Zaidi J, Campbell S, Pittrof R, Tan SL (1995) Endometrial thickness, morphology, vascular penetration and velocimetry in predicting implantation in an in vitro fertilization program. Ultrasound Obstet Gynecol 6:191–198

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to C. Dorn.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Dorn, C., Reinsberg, J., Willeke, C. et al. Three-dimensional power Doppler ultrasound of the subendometrial blood flow under the administration of a contrast agent (Levovist). Arch Gynecol Obstet 270, 94–98 (2004). https://doi.org/10.1007/s00404-003-0491-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00404-003-0491-z

Keywords

Navigation