Abstract
Introduction
The concept of custom total knee arthroplasty (TKA) is explored with specific attention to current limitations. Arguments in favor of custom TKA are the anatomic and functional variability we encounter in our patients. The biggest conceptual challenge is to marry the need for correction of deformity with the ambition to stay as close as possible to original anatomy.
Materials and methods
A Pubmed search was performed on the following terms: ‘patient specific implant’, ‘custom made implant’, ‘custom implant’, ‘total knee arthroplasty’ and ‘total knee replacement’. These studies were evaluated for the following intra- and post-operative variables: blood loss, hospital stay, range of motion, patient-reported outcome measures, limb and implant alignment, implant fit, tibiofemoral kinematics, complications and revision rates.
Results
Out of 1117 studies found with the initial search, a total of 17 articles were included in the final analysis. In eight out of the 17 (47%) studies, either the research was commercially funded or one of the authors had a conflict of interest related to the work. 11 out of 17 studies included a control group in their study setup. Of those studies that included a control group, both superior and inferior results compared to off-the-shelf implants have been reported.
Conclusion
Custom knee implants are the next step in matching the geometric features of the prosthesis to the anatomy of the individual patient, after several iterations that added asymmetry and sizes in the existing implants. Several companies have proven that it is feasible to produce these implants in a safe way. An overview of current literature reveals the lack of strong methodological studies that prove the value of this new technology. Custom knee implants face conceptual and practical difficulties, some of which might be overcome with technological advances, such as robotics and artificial intelligence.
Similar content being viewed by others
References
Richter P (2016) Official launch of the International Consortium for Personalised Medicine. http://www.icpermed.eu/. Accessed 19 April 2021
Sañudo JR, Vázquez R, Puerta J (2021) Meaning and clinical interest of the anatomical variations in the 21st century. Eur J Anat 7:1–3
Vesalius A, van Calcar JS, Oporinus J (1543). De humani corporis fabrica libri septem. Basilae: ex Ioannis Oporini
Straus WL, Temkin O (1943) Vesalius and the problem of variability. Bull Hist Med 14:609–633
LaPrade RF, Engebretsen AH, Ly TV et al (2007) The anatomy of the medial part of the knee. JBJS 89:2000–2010
Victor JMK, Bassens D, Bellemans J et al (2014) Constitutional varus does not affect joint line orientation in the coronal plane. Clin Orthop Relat Res 472:98–104
Thienpont E, Schwab P-E, Cornu O et al (2017) Bone morphotypes of the varus and valgus knee. Arch Orthop Trauma Surg 137:393–400
Bonnin MP, Saffarini M, Bossard N et al (2016) Morphometric analysis of the distal femur in total knee arthroplasty and native knees. Bone Jt J 98:49–57
Bellemans J, Carpentier K, Vandenneucker H et al (2010) The John Insall award: both morphotype and gender influence the shape of the knee in patients undergoing TKA. Clin Orthop Relat Res 468:29–36. https://doi.org/10.1007/s11999-009-1016-2
Engh GA, Lounici S, Rao AR, Collier MB (2001) In vivo deterioration of tibial baseplate locking mechanisms in contemporary modular total knee components. JBJS 83:1660–1665
Ivie CB, Probst PJ, Bal AK et al (2014) Improved radiographic outcomes with patient-specific total knee arthroplasty. J Arthroplast 29:2100–2103. https://doi.org/10.1016/j.arth.2014.06.024
Kay AB, Kurtz WB, Martin GM et al (2018) Manipulation rate is not increased after customized total knee arthroplasty. Reconstr Rev. https://doi.org/10.15438/rr.8.1.210
Steinert AF, Sefrin L, Jansen B et al (2020) Patient-specific cruciate-retaining total knee replacement with individualized implants and instruments (iTotalTM CR G2). Oper Orthop Traumatol. https://doi.org/10.1007/s00064-020-00690-8
Wheatley B, Nappo K, Fisch J et al (2019) Early outcomes of patient-specific posterior stabilized total knee arthroplasty implants. J Orthop 16:14–18. https://doi.org/10.1016/j.jor.2018.11.003
White PB, Ranawat AS (2016) Patient-specific total knees demonstrate a higher manipulation rate compared to “off-the-shelf implants.” J Arthroplast 31:107–111. https://doi.org/10.1016/j.arth.2015.07.041
Zeller IM, Sharma A, Kurtz WB et al (2017) Customized versus patient-sized cruciate-retaining total knee arthroplasty: an in vivo kinematics study using mobile fluoroscopy. J Arthroplast 32:1344–1350. https://doi.org/10.1016/j.arth.2016.09.034
Arnholdt J, Kamawal Y, Horas K et al (2020) Accurate implant fit and leg alignment after cruciate-retaining patient-specific total knee arthroplasty. BMC Musculoskelet Disord 21:699. https://doi.org/10.1186/s12891-020-03707-2
Bonnin MP, Beckers L, Leon A et al (2020) Custom total knee arthroplasty facilitates restoration of constitutional coronal alignment. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-020-06153-8
Culler SD, Martin GM, Swearingen A (2017) Comparison of adverse events rates and hospital cost between customized individually made implants and standard off-the-shelf implants for total knee arthroplasty. Arthroplast Today 3:257–263. https://doi.org/10.1016/j.artd.2017.05.001
Kumar P, Elfrink J, Daniels JP et al (2020) Higher component malposition rates with patient-specific cruciate retaining TKA than contemporary posterior stabilized TKA. J Knee Surg. https://doi.org/10.1055/s-0040-1701453
Levengood GA, Dupee J (2018) Accuracy of coronal plane mechanical alignment in a customized, individually made total knee replacement with patient-specific instrumentation. J Knee Surg 31:792–796. https://doi.org/10.1055/s-0037-1608946
Meheux CJ, Park KJ, Clyburn TA (2019) A retrospective study comparing a patient-specific design total knee arthroplasty with an off-the-shelf design: unexpected catastrophic failure seen in the early patient-specific design. J Am Acad Orthop Surg Glob Res Rev. https://doi.org/10.5435/JAAOSGlobal-D-19-00143
Neginhal V, Kurtz W, Schroeder L (2020) Patient satisfaction, functional outcomes, and survivorship in patients with a customized posterior-stabilized total knee replacement. JBJS Rev 8:e1900104. https://doi.org/10.2106/JBJS.RVW.19.00104
Patil S, Bunn A, Bugbee WD et al (2015) Patient-specific implants with custom cutting blocks better approximate natural knee kinematics than standard TKA without custom cutting blocks. Knee 22:624–629. https://doi.org/10.1016/j.knee.2015.08.002
Reimann P, Brucker M, Arbab D, Lüring C (2019) Patient satisfaction—a comparison between patient-specific implants and conventional total knee arthroplasty. J Orthop 16:273–277. https://doi.org/10.1016/j.jor.2019.03.020
Schroeder L, Martin G (2019) In vivo tibial fit and rotational analysis of a customized, patient-specific tka versus off-the-shelf TKA. J Knee Surg 32:499–505. https://doi.org/10.1055/s-0038-1653966
Schwarzkopf R, Brodsky M, Garcia GA, Gomoll AH (2015) Surgical and functional outcomes in patients undergoing total knee replacement with patient-specific implants compared with “off-the-shelf” implants. Orthop J Sport Med 3:2325967115590379. https://doi.org/10.1177/2325967115590379
Meier M, Zingde S, Steinert A et al (2019) What Is the possible impact of high variability of distal femoral geometry on TKA? A CT data analysis of 24,042 knees. Clin Orthop Relat Res 477:561–570. https://doi.org/10.1097/CORR.0000000000000611
Bonnin MP, de Kok A, Verstraete M et al (2017) Popliteus impingement after TKA may occur with well-sized prostheses. Knee Surg Sports Traumatol Arthrosc 25:1720–1730. https://doi.org/10.1007/s00167-016-4330-8
Rivière C, Iranpour F, Auvinet E et al (2017) Alignment options for total knee arthroplasty: a systematic review. Orthop Traumatol Surg Res 103:1047–1056. https://doi.org/10.1016/j.otsr.2017.07.010
Bellemans J, Vandenneucker H, Vanlauwe J, Victor J (2010) The influence of coronal plane deformity on mediolateral ligament status: an observational study in varus knees. Knee Surg Sport Traumatol Arthrosc 18:152–156. https://doi.org/10.1007/s00167-009-0903-0
Bellemans J, Vandenneucker H, Victor J, Vanlauwe J (2006) Flexion contracture in total knee arthroplasty. Clin Orthop Relat Res 452:78–82. https://doi.org/10.1097/01.blo.0000238791.36725.c5
Meere PA, Schneider SM, Walker PS (2016) Accuracy of balancing at total knee surgery using an instrumented tibial trial. J Arthroplast 31:1938–1942. https://doi.org/10.1016/j.arth.2016.02.050
Zhang J, Ndou WS, Ng N et al (2021) Robotic-arm assisted total knee arthroplasty is associated with improved accuracy and patient reported outcomes: a systematic review and meta-analysis. Knee Surg Sports Traumatol Arthrosc. https://doi.org/10.1007/s00167-021-06464-4
Pondal M, del Ser T (2008) Normative data and determinants for the timed “up and go” test in a population-based sample of elderly individuals without gait disturbances. J Geriatr Phys Ther 31:57–63. https://doi.org/10.1519/00139143-200831020-00004
Hortobágyi T, Westerkamp L, Beam S et al (2005) Altered hamstring-quadriceps muscle balance in patients with knee osteoarthritis. Clin Biomech (Bristol, Avon) 20:97–104. https://doi.org/10.1016/j.clinbiomech.2004.08.004
Funding
No funding was received for conducting this study.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
Each author certifies that he or she has no commercial associations (e.g., consultancies, stock ownership, equity interest, patent/licensing arrangements, etc.) that might pose a conflict of interest in connection with the submitted article. The authors did not receive support from any organization for the submitted work. No funding was received to assist with the preparation of this manuscript. No funding was received for conducting this study. No funds, grants, or other support was received.
Ethical approval
As no patients are included in this study, no ethical approval was necessary.
Informed consent
As no patients are included in this study, no informed consents were necessary.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Victor, J., Vermue, H. Custom TKA: what to expect and where do we stand today?. Arch Orthop Trauma Surg 141, 2195–2203 (2021). https://doi.org/10.1007/s00402-021-04038-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-021-04038-0