Abstract
Introduction
In our previous studies, we were able to identify anatomical differences as a predictor for aseptic loosening following primary and revision surgery with the use of rotating hinge prosthesis. This study was performed to answer following question: can a novel radiological classification system of the distal femur be identified?
Materials and methods
A total of 200 patients who received standardized anteroposterior (AP) and lateral views of the knee joint were included in this study. On AP radiographs, we measured the distance between inner diameter of the femur at 20 cm proximally from the knee and at a point 2 cm proximally from the adductor tubercle. The ratio of the inner diameter of the femoral canal at 20 cm proximal of knee joint to the inner diameter of medullary canal at 2 cm proximal of adductor tubercle was used as a novel index ratio. Two observers blindly and independently reviewed the anteroposterior radiographs twice.
Results
Three groups of anatomical classification can be constructed for each sex according to the 25th and 75th percentiles. A higher distribution of Type C was found in female patients. The median intra-observer reliability for rater 1 was 0.995 (IQR 0.994–0.997). We had also a high inter-observer reliability with ICC of 0.997 (95% CI 0.996–0.998).
Conclusions
The novel classification presents three different types of the knee joint for male and female patients. Type C has a wider inner diaphyseal diameter compared to Type A with a narrow inner diaphyseal diameter.
Similar content being viewed by others
Availability of data and materials
Data and materials are available.
References
Rodriguez-Merchan EC (2019) Total knee arthroplasty using hinge joints: Indications and results. EFORT Open Rev 4(4):121–132. https://doi.org/10.1302/2058-5241.4.180056
Abdulkarim A, Keane A, Hu SY, Glen L, Murphy DJ (2019) Rotating-hinge knee prosthesis as a viable option in primary surgery: literature review & meta-analysis. Orthop Traumatol Surg Res 105(7):1351–1359. https://doi.org/10.1016/j.otsr.2019.08.012
Kouk S, Rathod PA, Maheshwari AV, Deshmukh AJ (2018) Rotating hinge prosthesis for complex revision total knee arthroplasty: a review of the literature. J Clin Orthop Trauma 9(1):29–33. https://doi.org/10.1016/j.jcot.2017.11.020
Morgan-Jones R, Oussedik SI, Graichen H, Haddad FS (2015) Zonal fixation in revision total knee arthroplasty. Bone Jt J 97-B(2):147–149. https://doi.org/10.1302/0301-620X.97B2.34144
Chang A, Hochberg M, Song J, Dunlop D, Chmiel JS, Nevitt M, Hayes K, Eaton C, Bathon J, Jackson R, Kwoh CK, Sharma L (2010) Frequency of varus and valgus thrust and factors associated with thrust presence in persons with or at higher risk of developing knee osteoarthritis. Arthritis Rheum 62(5):1403–1411. https://doi.org/10.1002/art.27377
Cheng FB, Ji XF, Lai Y, Feng JC, Zheng WX, Sun YF, Fu YW, Li YQ (2009) Three dimensional morphometry of the knee to design the total knee arthroplasty for Chinese population. Knee 16(5):341–347. https://doi.org/10.1016/j.knee.2008.12.019
Gillespie RJ, Levine A, Fitzgerald SJ, Kolaczko J, DeMaio M, Marcus RE, Cooperman DR (2011) Gender differences in the anatomy of the distal femur. J Bone Jt Surg Br 93(3):357–363. https://doi.org/10.1302/0301-620X.93B3.24708
Hitt K, Shurman JR II, Greene K, McCarthy J, Moskal J, Hoeman T, Mont MA (2003) Anthropometric measurements of the human knee: correlation to the sizing of current knee arthroplasty systems. J Bone Joint Surg Am 85-A(Suppl 4):115–122
Walensky NA (1965) A study of anterior femoral curvature in man. Anat Rec 151:559–570. https://doi.org/10.1002/ar.1091510406
Ho WP, Cheng CK, Liau JJ (2006) Morphometrical measurements of resected surface of femurs in Chinese knees: correlation to the sizing of current femoral implants. Knee 13(1):12–14. https://doi.org/10.1016/j.knee.2005.05.002
Urabe K, Mahoney OM, Mabuchi K, Itoman M (2008) Morphologic differences of the distal femur between Caucasian and Japanese women. J Orthop Surg (Hong Kong) 16(3):312–315. https://doi.org/10.1177/230949900801600309
Urabe K, Miura H, Kuwano T, Matsuda S, Nagamine R, Sakai S, Masuda K, Iwamoto Y (2003) Comparison between the shape of resected femoral sections and femoral prostheses used in total knee arthroplasty in Japanese patients: simulation using three-dimensional computed tomography. J Knee Surg 16(1):27–33
Yue B, Varadarajan KM, Ai S, Tang T, Rubash HE, Li G (2011) Differences of knee anthropometry between Chinese and white men and women. J Arthroplasty 26(1):124–130. https://doi.org/10.1016/j.arth.2009.11.020
Everhart JS, Chaudhari AM, Flanigan DC (2016) Creation of a simple distal femur morphology classification system. J Orthop Res 34(6):924–931. https://doi.org/10.1002/jor.23102
Levent A, Suero EM, Gehrke TMC (2020) Risk factors for aseptic loosening after total knee arthroplasty with a rotating hinge implant. A case-control study. J Bone Jt Surg Am. https://doi.org/10.2106/JBJS.20.00788
Levent A, Suero EM, Godarzi BI, Gehrke T, Citak M (2021) Risk factors for aseptic loosening in complex revision total knee arthroplasty using rotating hinge implants. Int Orthop 45(1):125–132
Koh YG, Nam JH, Chung HS, Kim HJ, Lee HY, Kang KT (2020) Gender differences exist in rotational anatomy of the distal femur in osteoarthritic knees using MRI. Knee Surg Sports Traumatol Arthrosc 28(9):2990–2997. https://doi.org/10.1007/s00167-019-05730-w
Schmutz B, Kmiec S Jr, Wullschleger ME, Altmann M, Schuetz M (2017) 3D Computer graphical anatomy study of the femur: a basis for a new nail design. Arch Orthop Trauma Surg 137(3):321–331. https://doi.org/10.1007/s00402-016-2621-7
Funding
There is no funding source.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest regarding this manuscript. Independent from the submitted manuscript, received royalties/payments of all authors are listed: payments for presentations from Waldemar Link, Hamburg, Germany, Zimmer Biomet Inc., Warsaw, USA. Heraeus, Ceramtec Co.
Ethical approval
Ethical approval was performed and has following number: WF-086/20.
Consent to participate
Consent to participate is given.
Consent to publish
Consent to publish is given.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Citak, M., Levent, A., Suero, E.M. et al. A novel radiological classification system of the distal femur. Arch Orthop Trauma Surg 142, 315–322 (2022). https://doi.org/10.1007/s00402-021-03828-w
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-021-03828-w