Abstract
Introduction
The use of rotating-hinge total knee arthroplasties (TKA), despite several developments in prosthetic design, remains controversial. Results as well as indications of these devices are still discussed in primary intention and for young patients. The aim was to analyze early complications and survival rate of rotating-hinge TKA in primary intention and for revisions.
Methods
A retrospective study included all the patients operated for primary or revision TKA procedure using a rotating-hinge TKA between 2015 and 2018. Clinical and radiological data were collected before surgery and then at a minimum follow-up of 1 year. The primary endpoint was the aseptic revision-free survival rate. Secondary endpoints were the overall survival rate, IKS scores, range of motion and patellar complications.
Results
Forty patients were included at an average follow-up of 18 months. Primary implantation was performed for 12 patients (30%), and revision for 28 cases (70%). At a mean follow-up of 18 months, only one implant was removed for a septic cause. The cumulative survival rate at 24 months was 95%. At final review, eight knees (20%) had been revised, five (12.5%) due to infection, two (5%) because of extensor mechanism failure, two (5%) for global stiffness. The objective and subjective IKS were significantly higher postoperatively in both primary and revision groups (p < 0.0001). Patellar height was significantly smaller after revision (p = 0.04), while ROM significantly improved in this group (p = 0.02). At final endpoint, one implant was removed for a septic cause.
Conclusion
This rotating-hinge TKA provides satisfying clinical and functional outcomes in primary intentions and in revision cases. There was no implant-associated complication. The complication rate remains high for revision surgery cases, mostly due to previous joint infections and poor soft tissue quality causing extensor mechanism failure. A longer-term study should be conducted to confirm this trend.
Similar content being viewed by others
References
Robertsson O, Ranstam J, Sundberg M, W-Dahl A, Lidgren L (2014) The Swedish knee arthroplasty register: a review. Bone Joint Res 3(7):217–222. https://doi.org/10.1302/2046-3758.37.2000289
Parratte S, Pagnano MW (2008) Instability after total knee arthroplasty. Instr Course Lect 57:295–304
Vince KG, Abdeen A, Sugimori T (2006) The unstable total knee arthroplasty: causes and cures. J Arthroplasty 21(4 Suppl 1):44–49. https://doi.org/10.1016/j.arth.2006.02.101
Kouk S, Rathod PA, Maheshwari AV, Deshmukh AJ (2018) Rotating hinge prosthesis for complex revision total knee arthroplasty: a review of the literature. J Clin Orthop Trauma 9(1):29–33. https://doi.org/10.1016/j.jcot.2017.11.020
Aubriot JH, Deburge A, Genet JP, Group G (2014) GUEPAR hinge knee prosthesis. Orthop Traumatol Surg Res 100(1):27–32. https://doi.org/10.1016/j.otsr.2013.12.012
Barrack RL (2001) Evolution of the rotating hinge for complex total knee arthroplasty. Clin Orthop Relat Res 392(392):292–299
Jones EC, Insall JN, Inglis AE, Ranawat CS (1979) GUEPAR knee arthroplasty results and late complications. Clin Orthop Relat Res 140:145–152
Barrack RL (2002) Rise of the rotating hinge in revision total knee arthroplasty. Orthopedics 25(10):1020–1058
Kester MA, Cook SD, Harding AF, Rodriguez RP, Pipkin CS (1988) An evaluation of the mechanical failure modalities of a rotating hinge knee prosthesis. Clin Orthop Relat Res 228:156–163
Draganich LF, Whitehurst JB, Chou LS, Piotrowski GA, Pottenger LA, Finn HA (1999) The effects of the rotating-hinge total knee replacement on gait and stair stepping. J Arthroplasty 14(6):743–755
Guenoun B, Latargez L, Freslon M, Defossez G, Salas N, Gayet LE (2009) Complications following rotating hinge Endo-Modell (Link) knee arthroplasty. Orthop Traumatol Surg Res 95(7):529–536. https://doi.org/10.1016/j.otsr.2009.07.013
Pour AE, Parvizi J, Slenker N, Purtill JJ, Sharkey PF (2007) Rotating hinged total knee replacement: use with caution. J Bone Joint Surg Am 89(8):1735–1741. https://doi.org/10.2106/JBJS.F.00893
Petrou G, Petrou H, Tilkeridis C, Stavrakis T, Kapetsis T, Kremmidas N et al (2004) Medium-term results with a primary cemented rotating-hinge total knee replacement. A 7- to 15-year follow-up. J Bone Joint Surg Br 86(6):813–817
Gehrke T, Kendoff D, Haasper C (2014) The role of hinges in primary total knee replacement. Bone Joint J 96-B(11 Supple A):93–95. https://doi.org/10.1302/0301-620X.96B11.34143
Kearns SM, Culp BM, Bohl DD, Sporer SM, Della Valle CJ, Levine BR (2018) Rotating hinge implants for complex primary and revision total knee arthroplasty. J Arthroplasty 33(3):766–770. https://doi.org/10.1016/j.arth.2017.10.009
Debette C, Parratte S, Maucort-Boulch D, Blanc G, Pauly V, Lustig S et al (2014) French adaptation of the new knee society scoring system for total knee arthroplasty. Orthop Traumatol Surg Res 100(5):531–534. https://doi.org/10.1016/j.otsr.2014.03.025
Gaillard R, Bankhead C, Budhiparama N, Batailler C, Servien E, Lustig S (2018) Influence of patella height on total knee arthroplasty: outcomes and survival. J Arthroplasty. https://doi.org/10.1016/j.arth.2018.10.037
Cottino U, Abdel MP, Perry KI, Mara KC, Lewallen DG, Hanssen AD (2017) Long-term results after total knee arthroplasty with contemporary rotating-hinge prostheses. J Bone Joint Surg Am 99(4):324–330. https://doi.org/10.2106/JBJS.16.00307
Farid YR, Thakral R, Finn HA (2015) Intermediate-term results of 142 single-design, rotating-hinge implants: frequent complications may not preclude salvage of severely affected knees. J Arthroplasty 30(12):2173–2180. https://doi.org/10.1016/j.arth.2015.06.033
Smith TH, Gad BV, Klika AK, Styron JF, Joyce TA, Barsoum WK (2013) Comparison of mechanical and nonmechanical failure rates associated with rotating hinged total knee arthroplasty in nontumor patients. J Arthroplasty 28(1):62–67e61. https://doi.org/10.1016/j.arth.2012.05.008
Springer BD, Hanssen AD, Sim FH, Lewallen DG (2001) The kinematic rotating hinge prosthesis for complex knee arthroplasty. Clin Orthop Relat Res 392(392):283–291
Westrich GH, Mollano AV, Sculco TP, Buly RL, Laskin RS, Windsor R (2000) Rotating hinge total knee arthroplasty in severly affected knees. Clin Orthop Relat Res 379(379):195–208
Brown LR, Clement ND, MacDonald DJ, Breusch SJ (2019) The survivorship of the link endo-rotational hinge total knee arthroplasty: 5-12-year follow-up of 100 patients. Arch Orthop Trauma Surg 139(1):107–112. https://doi.org/10.1007/s00402-018-3064-0
Zanirato A, Cavagnaro L, Basso M, Divano S, Felli L, Formica M (2018) Metaphyseal sleeves in total knee arthroplasty revision: complications, clinical and radiological results. A systematic review of the literature. Arch Orthop Trauma Surg 138(7):993–1001. https://doi.org/10.1007/s00402-018-2967-0
Gathen M, Wimmer MD, Ploeger MM, Weinhold L, Schmid M, Wirtz DC et al (2018) Comparison of two-stage revision arthroplasty and intramedullary arthrodesis in patients with failed infected knee arthroplasty. Arch Orthop Trauma Surg 138(10):1443–1452. https://doi.org/10.1007/s00402-018-3007-9
Deehan DJ, Murray J, Birdsall PD, Holland JP, Pinder IM (2008) The role of the rotating hinge prosthesis in the salvage arthroplasty setting. J Arthroplast 23(5):683–688. https://doi.org/10.1016/j.arth.2007.05.055
Funding
There is no funding source.
Author information
Authors and Affiliations
Contributions
LR contributed to study design, data collection, statistical analysis, literature review and manuscript writing CB contributed to study design, statistical analysis, literature review and manuscript editing. BM contributed to study design, literature review and manuscript editing. PN and ES contributed to study design and manuscript editing. SL contributed to study design, supervision, literature review and manuscript editing. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Conflict of interest
CB, LR and BM declare that they have no conflict of interest. PN declare the following conflicts of interest: consultant for Smith & Nephew, royalties from Tornier-Wright, institutional research support to Tornier-Wright and Amplitude. ES declare she is consultant for Corin. SL declare the following conflicts of interest: consultant for Smith & Nephew, institutional research support to Corin and Amplitude. No benefits in any form have been received or will be received from a commercial party related directly or indirectly to the subject of this article.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Rouquette, L., Batailler, C., Muller, B. et al. Early complications and causes of revision after rotating-hinge TKA. Arch Orthop Trauma Surg 140, 109–119 (2020). https://doi.org/10.1007/s00402-019-03290-9
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-019-03290-9