Abstract
Introduction
The insertion of thoracic pedicle screws (T1–T10) is subject to a relevant rate of malplacement. The optimum implantation procedure is still a topic of controversial debate. Currently, a postoperative computed tomography is required to evaluate the screw positions. The present study was undertaken to clarify whether intraoperative 3D imaging is a reliable method of determining the position of thoracic pedicle screws.
Methods
This prospective study involved 40 consecutive patients with thoracic spinal injuries, with intraoperative 3D scans being performed to determine the positions of 240 pedicle screws in T1–T10. The results of the 3D scans were compared with the findings of postoperative CT scans, using a clinical classification system.
Results
The positions of 204 pedicle screws could be viewed by means of both 3D and CT scans and the results compared. The 3D scans achieved a sensitivity of 90.9 % and a specificity of 98.8 %. The rate of misclassification by the 3D scans was 2.5 %. Nine pedicle screws were classified as misplaced and their position corrected intraoperatively (3.8 %). No screws required postoperative revision.
Conclusions
Performing an intraoperative 3D scan enables the position of thoracic pedicle screws to be determined with sufficient accuracy. The rate of revision surgery was reduced to 0 %.
Similar content being viewed by others
References
Vaccaro AR, Rizzolo SJ, Balderston RA, Allardyce TJ, Garfin SR, Dolinskas C, An HS (1995) Placement of pedicle screws in the thoracic spine. Part II: an anatomical and radiographic assessment. J Bone Joint Surg Am 77:1200–1206
Bransford R, Bellabarba C, Thompson JH, Henley MB, Mirza SK, Chapman JR (2006) The safety of fluoroscopically-assisted thoracic pedicle screw instrumentation for spine trauma. J Trauma 60:1047–1052
Lekovic GP, Potts EA, Karahalios DG, Hall G (2007) A comparison of two techniques in image-guided thoracic pedicle screw placement: a retrospective study of 37 patients and 277 pedicle screws. J Neurosurg Spine 7:393–398
Schizas C, Theumann N, Kosmopoulos V (2007) Inserting pedicle screws in the upper thoracic spine without the use of fluoroscopy or image guidance. Is it safe? Eur Spine J 16:625–629
Kuntz C, Maher PC, Levine NB, Kurokawa R (2004) Prospective evaluation of thoracic pedicle screw placement using fluoroscopic imaging. J Spinal Disord 17:206–214
Upendra BN, Meena D, Chowdhury B, Ahmad A, Jayaswal A (2008) Outcome-based classification for assessment of thoracic pedicular screw placement. Spine 33:384–390
Zdichavsky M, Blauth M, Knop C, Lotz J, Krettek C, Bastian L (2004) Accuracy of pedicle screw placement in thoracic spine fractures. Part II: a retrospective analysis of 278 pedicle screws using computed tomographic scans. Eur J Trauma 30:241–247
Sarlak AY, Tosun B, Atmaca H, Sarisoy HT, Buluc L (2009) Evaluation of thoracic pedicle screw placement in adolescent idiopathic scoliosis. Eur Spine J 18:1892–1897
Kim YJ, Lenke LG, Cheh G, Riew KD (2005) Evaluation of pedicle screw placement in the deformed spine using intraoperative plain radiographs: a comparison with computerized tomography. Spine 30:2084–2088
Berlemann U, Heini P, Müller U, Stoupis C, Schwarzenbach O (1997) Reliability of pedicle screw assessment utilizing plain radiographs versus CT reconstruction. Eur Spine J 6:406–411
Weinstein JN, Spratt KF, Spengler D, Brick C, Reid S (1988) Spinal pedicle fixation: reliability and validity of roentgenogram-based assessment and surgical factors on successful screw placement. Spine 13:1012–1018
Zausinger S, Scheder B, Uhl E, Heigl T, Morhard D, Tonn JC (2009) Intraoperative computed tomography with integrated navigation system in spinal stabilizations. Spine 34:2919–2926
Rock C, Kotsianos D, Linsenmaier U, Fischer T, Brandl R, Vill F, Wirth S, Kaltschmidt R, Euler E, Pfeifer KJ, Reiser M (2002) Studies on image quality, high contrast resolution and dose for the axial skeleton and limbs with a new, dedicated CT system (ISO-C-3 D). Rofo 174:170–176
Beck M, Mittlmeier T, Gierer P, Harms C, Gradl G (2009) Benefit and accuracy of intraoperative 3D-imaging after pedicle screw placement: a prospective study in stabilizing thoracolumbar fractures. Eur Spine J 18:1469–1477
Zdichavsky M, Blauth M, Knop C, Graessner M, Hermann H, Krettek C, Bastian L (2004) Accuracy of pedicle screw placement in thoracic spine fractures. Part I: inter- and intraobserver reliability of the scoring system. Eur J Trauma 30:234–240
Berlet GC, Boubez G, Gurr KR, Bailey SI (1999) The USS pedicle hook system: a morphometric analysis of its safety in the thoracic spine. J Spinal Disord 12:234–239
Kretzer RM, Chaput C, Sciubba DM, Garonzik IM, Jallo GI, McAfee PC, Cunningham BW, Tortolani PJ (2011) A computed tomography-based morphometric study of thoracic pedicle anatomy in a random United States trauma population. J Neurosurg Spine 14:235–243
Vaccaro AR, Rizzolo SJ, Allardyce TJ, Ramsey M, Salvo J, Balderston RA, Cotler JM (1995) Placement of pedicle screws in the thoracic spine. Part I: morphometric analysis of the thoracic vertebrae. J Bone Joint Surg Am 77:1193–1199
Lien SB, Liou NH, Wu SS (2007) Analysis of anatomic morphometry of the pedicles and the safe zone for through-pedicle procedures in the thoracic and lumbar spine. Eur Spine J 16:1215–1222
Parker SL, McGirt MJ, Farber SH, Amin AG, Rick AM, Suk I, Bydon A, Sciubba DM, Wolinsky JP, Gokaslan ZL, Witham TF (2011) Accuracy of free-hand pedicle screws in the thoracic and lumbar spine: analysis of 6816 consecutive screws. Neurosurgery 68:170–178
Fisher CG, Sahajpal V, Keynan O, Boyd M, Graeb D, Bailey C, Panagiotopoulos K, Dvorak F (2006) Accuracy and safety of pedicle screw fixation in thoracic spine trauma. J Neurosurg Spine 5:520–526
Kosmopoulos V, Schizas C (2007) Pedicle screw placement accuracy. A meta-analysis. Spine 32:E111–E120
Belmont JB, Klemme WR, Dhawan A, Polly DW (2001) In vivo accuracy of thoracic pedicle screws. Spine 26:2340–2346
Atesok K, Finkelstein J, Khoury A, Peyser A, Weil Y, Liebergall M, Mosheiff R (2007) The use of intraoperative three-dimensional imaging (ISO-C-3D) in fixation of intraarticular fractures. Injury 38:1163–1169
Verlaan JJ, van de Kraats EB, van Walsum T, Dhert WJA, Oner FC, Niessen WJ (2005) Three-dimensional rotational X-ray imaging for spine surgery. Spine 30:556–561
Wang MY, Kim KA, Liu CY, Kim P, Apuzzo ML (2004) Reliability of three-dimensional fluoroscopy for detecting pedicle screw violations in the thoracic and lumbar spine. Neurosurgery 54:1138–1142
Santos ER, Ledonio CG, Castro CA, Truong WH, Sembrano JN (2012) The accuracy of intraoperative O-arm images for the assessment of pedicle screw position. Spine 37:E119–E125
Samdani AF, Tantorski M, Cahill PJ, Ranade A, Koch S, Clements DH, Betz RR, Asghar J (2011) Triggered electromyography for placement of thoracic pedicle screws: is it reliable? Eur Spine J 20:869–874
de Blas G, Barrios C, Regidor I, Montes E, Burgos J, Pizá-Vallespir G, Hevia E (2012) Safe pedicle screw placement in thoracic scoliotic curves using t-EMG stimulation threshold variability at concavity and convexity in apex segments. Spine 37:E387–E395
Biswas D, Bible JE, Bohan M, Simpson AK, Whang PG, Grauer JN (2009) Radiation exposure from musculoskeletal computerized tomographic scans. J Bone Joint Surg Am 91:1882–1889
Richards PJ, George J, Metelko M, Brown M (2010) Spine computed tomography doses and cancer induction. Spine 35:430–433
Abul-Kasim K, Ohlin A, Strömbeck A, Maly P, Sundgren PC (2010) Radiological and clinical outcome of screw placement in adolescent idiopathic scoliosis: evaluation with low-dose computed tomography. Eur Spine J 19:96–104
Acknowledgments
The authors thank T. Wodetzki (University of Rostock) for designing the medical illustrations.
Conflict of interest
No funds were received in support of this work. No benefits in any form have been or will be received from a commercial party related directly or indirectly to the subject of the manuscript including National Institutes of Health (NIH), Wellcome Trust, Howard Hughes Medical Institute (HHMI), and others.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Beck, M., Rotter, R., Gradl, G. et al. Reliability and consequences of intraoperative 3D imaging to control positions of thoracic pedicle screws. Arch Orthop Trauma Surg 132, 1371–1377 (2012). https://doi.org/10.1007/s00402-012-1555-y
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-012-1555-y