Complete acetabular cup revision versus isolated liner exchange for polyethylene wear and osteolysis without loosening in cementless total hip arthroplasty

  • Hip Arthroplasty
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Introduction

Revision surgery in patients showing polyethylene wear and acetabular osteolysis without visible acetabular cup loosening involves the difficult decision of whether to revise only the liner or both the cup and the liner. The purpose of this study is to compare the outcomes of complete acetabular revision and isolated liner exchange in patients showing wear and osteolysis without loosening.

Materials and methods

We evaluated 80 cases of revision surgery for polyethylene wear and osteolysis without cup loosening performed between October 1997 and December 2008. The cup revision group consisted of 45 patients who underwent a complete acetabular cup replacement, and the cup retention group consisted of 35 patients who underwent either an isolated liner exchange or a liner cementing procedure. Comparisons between the two groups were performed.

Results

There were differences in femoral stem revision, estimated blood loss, and hospital stay. Other variables including complications, osteolysis progression, re-revision rate, clinical score, and satisfaction showed no differences between the two groups. There was one case of early loosening and subsequent re-revision surgery in the cup revision group, as well as one case of wear progression and liner dislodgement leading to complete re-revision of the acetabular component and femoral stem in the cup retention group.

Conclusion

We found no differences in acetabular osteolysis progression, fixation failure, or complication between the cup revision and retention groups. Therefore, isolated liner exchange without cup extraction in cases of osteolysis that includes a well-fixed and well-positioned shell could be considered as a viable treatment option.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Beaule PE, Campbell P, Mirra J, Hooper JC, Schmalzried TP (2001) Osteolysis in a cementless, second generation metal-on-metal hip replacement. Clin Orthop Relat Res 386:159–165

    Google Scholar 

  2. Berry DJ (1999) Management of osteolysis around total hip arthroplasty. Orthopedics 22:805–808

    PubMed  CAS  Google Scholar 

  3. Blaha JD (2002) Well-fixed acetabular component retention or replacement: the whys and the wherefores. J Arthroplasty 17:157–161

    Article  PubMed  Google Scholar 

  4. Blom AW, Astle L, Loveridge J, Learmonth ID (2005) Revision of an acetabular liner has a high risk of dislocation. J Bone Joint Surg Br 87:1636–1638

    Article  PubMed  CAS  Google Scholar 

  5. Boucher HR, Lynch C, Young AM, Engh CA Jr, Engh C Sr (2003) Dislocation after polyethylene liner exchange in total hip arthroplasty. J Arthroplasty 18:654–657

    Article  PubMed  Google Scholar 

  6. Chiang PP, Burke DW, Freiberg AA, Rubash HE (2003) Osteolysis of the pelvis: evaluation and treatment. Clin Orthop Relat Res 417:164–174

    Google Scholar 

  7. Claus AM, Walde TA, Leung SB, Wolf RL, Engh CA Sr (2003) Management of patients with acetabular socket wear and pelvic osteolysis. J Arthroplasty 18:112–117

    Article  PubMed  Google Scholar 

  8. Della Valle CJ, Shuaipaj T, Berger RA, Rosenberg AG, Shott S, Jacobs JJ, Galante JO (2005) Revision of the acetabular component without cement after total hip arthroplasty. A concise follow-up, at fifteen to nineteen years, of a previous report. J Bone Joint Surg Am 87:1795–1800

    Article  PubMed  Google Scholar 

  9. Gonzalez della Valle A, Ruzo PS, Li S, Pellicci P, Sculco TP, Salvati EA (2001) Dislodgment of polyethylene liners in first and second-generation Harris–Galante acetabular components. A report of eighteen cases. J Bone Joint Surg Am 83-A:553–559

    PubMed  CAS  Google Scholar 

  10. Griffin WL, Fehring TK, Mason JB, McCoy TH, Odum S, Terefenko CS (2004) Early morbidity of modular exchange for polyethylene wear and osteolysis. J Arthroplasty 19:61–66

    Article  PubMed  Google Scholar 

  11. Haft GF, Heiner AD, Callaghan JJ, Dorr LD, Wan Z, Long W, Longjohn DB, Brown TD (2002) Polyethylene liner cementation into fixed acetabular shells. J Arthroplasty 17:167–170

    Article  PubMed  Google Scholar 

  12. He C, Feng JM, Yang QM, Wang Y, Liu ZH (2009) Results of selective hip arthroplasty revision in isolated acetabular failure. J Surg Res 164:228–233

    Article  PubMed  Google Scholar 

  13. Hodgkinson JP, Shelley P, Wroblewski BM (1988) The correlation between the roentgenographic appearance and operative findings at the bone-cement junction of the socket in Charnley low friction arthroplasties. Clin Orthop Relat Res 228:105–109

    Google Scholar 

  14. Hozack WJ, Bicalho PS, Eng K (1996) Treatment of femoral osteolysis with cementless total hip revision. J Arthroplasty 11:668–672

    Article  PubMed  CAS  Google Scholar 

  15. Lachiewicz PF, Soileau ES (2005) Changing indications for revision total hip arthroplasty. J Surg Orthop Adv 14:82–84

    PubMed  Google Scholar 

  16. Lachiewicz PF, Soileau ES (2006) Polyethylene liner exchange of the Harris–Galante porous I and II acetabular components without cement: results and complications. J Arthroplasty 21:992–997

    Article  PubMed  Google Scholar 

  17. Leung S, Naudie D, Kitamura N, Walde T, Engh CA (2005) Computed tomography in the assessment of periacetabular osteolysis. J Bone Joint Surg Am 87:592–597

    Article  PubMed  Google Scholar 

  18. Lie SA, Hallan G, Furnes O, Havelin LI, Engesaeter LB (2007) Isolated acetabular liner exchange compared with complete acetabular component revision in revision of primary uncemented acetabular components: a study of 1649 revisions from the Norwegian Arthroplasty Register. J Bone Joint Surg Br 89:591–594

    Article  PubMed  CAS  Google Scholar 

  19. Makela KT, Eskelinen A, Pulkkinen P, Paavolainen P, Remes V (2008) Total hip arthroplasty for primary osteoarthritis in patients fifty-five years of age or older. An analysis of the Finnish arthroplasty registry. J Bone Joint Surg Am 90:2160–2170

    Article  PubMed  Google Scholar 

  20. Maloney WJ, Herzwurm P, Paprosky W, Rubash HE, Engh CA (1997) Treatment of pelvic osteolysis associated with a stable acetabular component inserted without cement as part of a total hip replacement. J Bone Joint Surg Am 79:1628–1634

    PubMed  CAS  Google Scholar 

  21. Maloney WJ, Paprosky W, Engh CA, Rubash H (2001) Surgical treatment of pelvic osteolysis. Clin Orthop Relat Res 393:78–84

    Google Scholar 

  22. Meldrum RD, Hollis JM (2001) The strength of a cement acetabular locking mechanism. J Arthroplasty 16:748–752

    Article  PubMed  CAS  Google Scholar 

  23. Mitchell PA, Masri BA, Garbuz DS, Greidanus NV, Wilson D, Duncan CP (2003) Removal of well-fixed, cementless, acetabular components in revision hip arthroplasty. J Bone Joint Surg Br 85:949–952

    Article  PubMed  CAS  Google Scholar 

  24. Naudie DD, Engh CA Sr (2004) Surgical management of polyethylene wear and pelvic osteolysis with modular uncemented acetabular components. J Arthroplasty 19:124–129

    Article  PubMed  Google Scholar 

  25. O’Brien JJ, Burnett RS, McCalden RW, MacDonald SJ, Bourne RB, Rorabeck CH (2004) Isolated liner exchange in revision total hip arthroplasty: clinical results using the direct lateral surgical approach. J Arthroplasty 19:414–423

    Article  PubMed  Google Scholar 

  26. Paprosky W, Sporer S, O’Rourke MR (2006) The treatment of pelvic discontinuity with acetabular cages. Clin Orthop Relat Res 453:183–187

    Article  PubMed  Google Scholar 

  27. Restrepo C, Ghanem E, Houssock C, Austin M, Parvizi J, Hozack WJ (2009) Isolated polyethylene exchange versus acetabular revision for polyethylene wear. Clin Orthop Relat Res 467:194–198

    Article  PubMed  Google Scholar 

  28. Schmalzried TP, Fowble VA, Amstutz HC (1998) The fate of pelvic osteolysis after reoperation. No recurrence with lesional treatment. Clin Orthop Relat Res 350:128–137

    Google Scholar 

  29. Schmalzried TP, Guttmann D, Grecula M, Amstutz HC (1994) The relationship between the design, position, and articular wear of acetabular components inserted without cement and the development of pelvic osteolysis. J Bone Joint Surg Am 76:677–688

    PubMed  CAS  Google Scholar 

  30. Springer BD, Hanssen AD, Lewallen DG (2003) Cementation of an acetabular liner into a well-fixed acetabular shell during revision total hip arthroplasty. J Arthroplasty 18:126–130

    Article  PubMed  Google Scholar 

  31. Stulberg BN, Della Valle AG (2008) What are the guidelines for the surgical and nonsurgical treatment of periprosthetic osteolysis? J Am Acad Orthop Surg 16(Suppl 1):S20–S25

    PubMed  Google Scholar 

  32. Su EP (2010) Lysis in the well-fixed shell: “hold ‘em” or “fold ‘em”. Orthopedics 33:650

    PubMed  Google Scholar 

  33. Talmo CT, Kwon YM, Freiberg AA, Rubash HE, Malchau H (2010) Management of polyethylene wear associated with a well-fixed modular cementless shell during revision total hip arthroplasty. J Arthroplasty 26(4):576–581

    Article  PubMed  Google Scholar 

  34. Wade FA, Rapuri VR, Parvizi J, Hozack WJ (2004) Isolated acetabular polyethylene exchange through the anterolateral approach. J Arthroplasty 19:498–500

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

This study was partially supported by a grant of Korea Health Technology R&D Project, Ministry for Health, Welfare & Family Affairs, Republic of Korea (A090876).

Conflict of interest

The authors declare that they have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Youn-Soo Park.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Koh, K.H., Moon, YW., Lim, SJ. et al. Complete acetabular cup revision versus isolated liner exchange for polyethylene wear and osteolysis without loosening in cementless total hip arthroplasty. Arch Orthop Trauma Surg 131, 1591–1600 (2011). https://doi.org/10.1007/s00402-011-1338-x

Download citation

  • Received:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-011-1338-x

Keywords

Navigation