Inter-observer validation study of quantitative CT-osteodensitometry in total knee arthroplasty

  • Basic Science
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Introduction

Proximal tibial bone quality is an important factor in implant mechanical stability following total knee arthroplasty. Quantitative computed tomography (CT) osteodensitometry has been used to measure the change in bone density post-operatively. We sought to validate the inter-observer reliability of these measurements.

Methods

Sixteen CT scans were obtained on cadaveric lower limbs. Two observers independently obtained cortical and cancellous bone mineral density measurements from these scans. Eight CT scans were performed with prosthesis in situ and eight without. Results were compared using correlation coefficient and Bland–Altman analysis.

Results

The study showed a high correlation coefficient of greater than 0.990 for all slices. The mean difference between observers falls within two standard deviations of the mean as seen on a Bland–Altman plot.

Interpretation

Quantitative CT-osteodensitometry is a reproducible method to measure bone mineral density changes about the knee and shows little variation between observers.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Ming GL, Nilsson KG (2000) The effect of the preoperative bone quality on the fixation of the tibial component in total knee arthroplasty. J Arthroplasty 115(6):744–753

    Google Scholar 

  2. Peterson MM, Nielson PT, Lauritzen JB, Lund B (1995) Changes in bone mineral density of the proximal tibia after uncemented total knee arthroplasty. Acta Orthop Scand 66(6):513–516

    Article  Google Scholar 

  3. Peterson MM, Nielson PT, Lebech A, Toksvig-Larsen S, Lund B (1999)Preoperative bone mineral density of the proximal tibia and migration of thetibial component after uncemented total knee arthroplasty. J Arthroplasty 14(1):77–81

    Article  Google Scholar 

  4. Karbowski A, Schwitalle M, Eckardt A, Heine J (1999) Periprosthetic bone remodeling after total knee arthroplasty: early assessment by dual energy X-ray absorptiometry. Arch Orthop Trauma Surg 119:324–326

    Article  PubMed  CAS  Google Scholar 

  5. Trevisan C, Bigoni M, Denti M, Marinoni EC, Ortolani S (1998) Bone assessment after total knee arthroplasty by dual energy X-ray absorptiometry: analysis protocol and reproducibility. Calcif Tissue Int 62:359–361

    Article  PubMed  CAS  Google Scholar 

  6. Schmidt R, Pitto RP, Kress A et al (2005) Inter- and intraobserver assessment of periacetabular osteodensitometry after cemented and uncemented total hip arthroplasty using computed tomography. Arch Orthop Trauma Surg 125:291–297

    Article  PubMed  Google Scholar 

  7. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet Feb:307–310

    Google Scholar 

  8. Zannoni C, Viceconti M, Pierotti L, Cappello A (1998) Analysis of titanium induced CT artifacts in the development of biomechanical finite element models. Med Eng Phys 20:653–659

    Article  PubMed  CAS  Google Scholar 

  9. Pitto RP, Pandit S, Munro J. (2005) Bone remodelling of the proximal tibia after fixed versus rotating platform total knee arthroplasty. Proceedings of the 23rd world congress of SIROT, Istanbul, Turkey, pp 245

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jacob Munro.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Reilly, K., Munro, J., Pandit, S. et al. Inter-observer validation study of quantitative CT-osteodensitometry in total knee arthroplasty. Arch Orthop Trauma Surg 127, 729–731 (2007). https://doi.org/10.1007/s00402-007-0351-6

Download citation

  • Received:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-007-0351-6

Keywords

Navigation