Long-term results of HA coated threaded versus HA coated hemispheric press fit cups: 287 hips followed for 11 to 16 years

  • Orthopaedic Surgery
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Introduction

There are many studies on the short and medium term results of hydroxyapatite (HA) coated acetabular components, but information on survival in the longer run is sparse. In this paper, we report 11 to 16 years follow up results of HA coated threaded versus HA coated hemispherical press fit cups.

Patients and methods

During the years 1988–1993, we performed 323 primary total hip replacements in 276 patients with an HA coated acetabular component. In 128 cases we used a HA coated threaded cup, and in 195 cases a HA coated hemispherical press fit cup was inserted. There were 189 women and 87 men aged 15–79 (mean 48) years. During follow-up, 14 patients died and 21 patients would not come for the follow-up examination as they were doing quite well with their prostheses. These patients were censored at their last control examination. This left 241 patients (287 hips) followed for 11 to 16 (mean 13) years after the operation.

Results

During the follow-up period 48 press fit cups and 9 threaded cups were revised, none because of infection. Survival analyses indicated a cumulative survival of the threaded HA coated cup of 0.91 (95% CI: 0.86–0.97) at 16 years with one patient at risk. The cumulative survival of the HA coated press fit cup was 0.74 (0.69–0.80) at 15 years with no patient at risk. The difference is significant (P = 0.0002). Wear and age was significantly associated with revision (P < 0.0001 and 0.0002, respectively), and wear was significantly (P < 0.0001) associated with osteolysis.

Conclusion

These results indicate that HA coated threaded cups provide satisfactory bony interlock to resist force loads in the long run whereas the HA coated hemispherical press fit cups do not.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Aldinger PR, Thomsen M Lukoschek M, Mau H, Ewerbeck V, Breusch SJ (2004) Long-term fate of uncemented, threaded acetabular components with smooth surface treatment: minimum 10-year follow-up of two different designs. Arch Orthop Trauma Surg 124:469–475

    Article  PubMed  Google Scholar 

  2. Bohm P, Bosche R (1998) Survival analysis of the Harrris-Galante cup. J Bone Joint Surg Br 80:396–403

    Article  PubMed  CAS  Google Scholar 

  3. Bruijn JD, Seelen JH, Feenstra RM, Hansen BE, Bernoski FP (1995) Failure of the Mecring screw-ring acetabular component in total hip arthroplasty. A three to seven-year follow-up study. J Bone Joint Surg Am 77:760–766

    PubMed  CAS  Google Scholar 

  4. Capello WN, Colyer RA, Kernek CB,Carnahan JV, Hess JJ (1993) Failure of the Mecron screw-in ring. J Bone Joint Surg Br 75:835–836

    PubMed  CAS  Google Scholar 

  5. Capello WN, D`Antonio JA, Manley M T, Feinberg JR (1998) Hydroxyaapatite in total hip arthroplasty. Clinical results and critical issues. Clin Orthop 355:200–211

    Article  PubMed  Google Scholar 

  6. Clohishy JC, Harris WH (1999). The Harris-Galante porous-coated acetabular component with screw fixation: an avearge ten-year follow-up study. J Bone Joint Surg Am 8:66–73

    Google Scholar 

  7. Cook SD, Thomas KA, Kay JF, Jarcho M (1988) Hydroxyapatitet-coated titanium for orthopedic implant application. Clin Orthop 232:225–243

    PubMed  CAS  Google Scholar 

  8. Crowther JD, Lachiewicz PF (2002). Survival and polyethylene wear of porous-coated acetabular components in patients less than 50 years old: results at nine to fourteen years. J Bone Joint Surg Am 84:729–735

    Article  PubMed  Google Scholar 

  9. D’Antonio JA, Capello WN, Manley MT, Feinberg J (1997) Hydroxyapatite coated implants: total hip arthroplasty in young patients and patients with avascular necrosis. Clin Orthop 344:124–138

    Article  PubMed  Google Scholar 

  10. DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip repleacement. Clin Orthop 121:20–32

    PubMed  Google Scholar 

  11. Fox GM, McBeath AA; Heiner JP (1994) Hip replacement with a threaded acetabular cup. A follow-up study. J Bone Joint Surg Am 76:195–201

    PubMed  CAS  Google Scholar 

  12. Garcia-Cimbrelo E, Cruz-Pardos A, Cordero J, Sanchez-Sotelo J (2000) Cementless total hip arthroplasty in patients younger than 40 years old: 20–25 year results. J Arthroplasty 15: 825–832

    Article  PubMed  CAS  Google Scholar 

  13. Geesink RGT, Hoefnagel NHM (1995). Six-year results of hydroxyapatite-coated total hip replacement. J Bone Joint Surg Br 77:534–547

    PubMed  CAS  Google Scholar 

  14. Giannikas KA, Din R,Sadiq S, Dunningham TH (2002) Mediumterm results of the ABG total hip arthroplasty in young patients. J Arthroplasty 17:184–188

    Article  PubMed  Google Scholar 

  15. Grubl A, Chiari C, Gruber M, Kaider A, Gottsauner-Wolf F (2002) Cementless total hip arhroplasty with a tapered, rectangular titanium stem and a threaded cup: a minimum 10 year follow-up. J Bone Joint Surg Am 84:425–431

    PubMed  Google Scholar 

  16. Johnsson R, Franzen H, Nilsson LT (1994) Combined survivorship and multivariate analysis of revision in 799 hip prostheses. A 10–20 year review of mechanical loosening. J Bone Joint Surg 76:439–443

    CAS  Google Scholar 

  17. Kroon PO, Freeman MAR (1992) Hydroxyapatite coating of hip prostheses. Effect on migration into the femur. J Bone Joint Surg Br 74-B:518–22

    Google Scholar 

  18. KuboT, Inoue S Maeda T, Arai Y, Hirakawa K, Wu Y, Suehara H, Ogura T, Hirasawa Y (2001) Cementless lord total hip arthroplasty: cup loosening common after a minimum 10-year follow-up of 103 hips. Acta Orthop Scand 72:585–590

    Article  Google Scholar 

  19. Lai KA, Shen WJ, Chen CH, Yang CY, Hu WP, Chang GL (2002) Failure of hydroxyapatite-coated acetabular cups. Ten-year follow-up of 85 Landos Atoll arthroplasties. J Bone Joint Surg Br 84:641–646

    Article  PubMed  Google Scholar 

  20. Livermore J, Ilstrup D, Morrey B (1990) Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am 72:518–28

    PubMed  CAS  Google Scholar 

  21. Maloney WJ, Peters P, Engh CA, Chandler H (1993): Severe osteolysis of the pelvis in association with acetabular replacement without cement. J Bone Joint Surg Am 75:1627–1635

    PubMed  CAS  Google Scholar 

  22. Manley MT, Capello WN, D`Antonio JA, Edidin AA, Geesink RGT (1998) Fixation of acetabular cups without cement in total hip arthroplasty: a comparison of three different implant surfaces with a minimum duration of follow-up of 5 years. J Bone Joint Surg Am 80:1175–1185

    PubMed  CAS  Google Scholar 

  23. Massin P, Schmidt L, Engh CA (1989) Evaluation of cementless acetabular component migration: an experiemental study. J Arthroplasty 4:245–251

    PubMed  CAS  Google Scholar 

  24. Morscher EW (1991) Hydroxyapatite coating of prostheses. J Bone Joint Surg Br 73: 705–706

    PubMed  CAS  Google Scholar 

  25. Morscher EW (1992) Current status of acetabular fixation in primary total hip arthroplasty. Clin Orthop 274:172–193

    PubMed  Google Scholar 

  26. Onsten I, Carlsson ÅS, Sanzén L, Besjakov J (1996) Migration and wear of a hydroxyapatite-coated hip prosthesis. J Bone Joint Surg Br 78:85–91

    PubMed  CAS  Google Scholar 

  27. Reikerås O, Gundessson RB (2002) Failure of HA coating on a gritblasted acetabular cup: 155 patients followed for 7–10 years. Acta Orthop Scand 73:104–108

    Article  PubMed  Google Scholar 

  28. Rossi P, Sibelli P, Fumero S, Crua E (1995) Short-term results of hydroxyapatite-coated primary total hip arthroplasty. Clin Orthop 310:98–102

    PubMed  Google Scholar 

  29. Røkkum M, Brandt M, Bye K, Hetland KR, Waage S, Reigstad A (1999) Polyethylene wear, osteolysis and acetabular loosening with a HA-coated prosthesis. A follow-up of 94 consecutive arthroplasties. J Bone Joint Surg Br 81:582–589

    Article  PubMed  Google Scholar 

  30. Overgaard S, Soballe K, Lind M, Bunger C (1997) Resorption of hydroxyapatite and fluoroapatite coatings in man; an experimental study in trabecular bone. J Bone Joint Surg Br 79:654–665

    Article  PubMed  CAS  Google Scholar 

  31. Udomkiat P, Dorr LD, Wan Z (20002). Cement less hemispheric porous coated sockets implanted with press-fit technique without screws: average ten-year follow-up. J Bone Joint Surg Am 84:1195–1200

    Google Scholar 

  32. Zicat B, Engh CA, Gocken E (1995) Patterns of osteolysis around total hip components inserted with and without cement. J Bone Joint Surg Am 77:432–439

    PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Olav Reikerås.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Reikerås, O., Gunderson, R.B. Long-term results of HA coated threaded versus HA coated hemispheric press fit cups: 287 hips followed for 11 to 16 years. Arch Orthop Trauma Surg 126, 503–508 (2006). https://doi.org/10.1007/s00402-006-0176-8

Download citation

  • Received:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-006-0176-8

Keywords

Navigation