Skip to main content

Advertisement

Log in

Increased polyethylene wear after cementless ABG I total hip arthroplasty

  • Orthopaedic Surgery
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Introduction

The cementless, hydroxyapatite-coated Anatomique Benoist Giraud-I (ABG-I) hip endoprosthesis represented a modern implant in the 1990s. The aim of the current retrospective study was to evaluate the clinical and radiological results of this prosthesis. In addition, an analysis of the complications and retrieved implants was conducted.

Materials/methods

The medium-term results (follow-up 5.23 years) of 193 hip joints are presented. Of 158 total cohorts, 81.9% was able to undergo follow-up performed with standardized clinical and radiological investigations. Physical characteristics of the patients and the underlying disease prompting the need for total hip arthroplasty, as well as a clinical score (Merle d’Aubigné) were recorded. At the time of follow-up, a radiologic examination of all patients with a standardized evaluation was performed. In addition, the migration of the acetabular cup and femoral head as well as polyethylene wear could be determined digitally in 118 cases (61.1%) using one-picture Roentgen analysis.

Results

Clinical results, as measured with a Merle d’Aubigné Score increase from 8.4 to 16.2, were very good. Radiographs demonstrated successful osseous integration of the anatomically molded shaft. Within the period of the investigation, no revision procedures of the femoral shaft were necessary. However, the rate of polyethylene abrasion of 0.23 mm/year was markedly high. 13.9% of hips (n = 27) required acetabular cup revision due to wear. This calculates to a prosthesis 7-year survival probability of 63%. Intraoperative findings during the revision cases showed extensive periacetabular osteolysis with foreign body granulation tissue. Analysis of data from the total patient cohort versus data from cases requiring revision showed a significantly increased frequency of high polyethylene wear in young active patients as well as in cases where an unfavorable acetabular cup to femoral head relation existed in correspondence with polyethylene thickness. There is evidence, however, that suggests that multifactorial causes for the increased wear are significant in regards to the principal material and technical features of the prosthesis.

Conclusion

On the basis of these results, it is strongly recommended that all patients treated with an ABG-I hip endoprosthesis should receive close clinical and most importantly close radiologic follow-up.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9

Similar content being viewed by others

References

  1. Araujo CG, Gonzalez JF, Tonino A (1998) Rheumatoid arthritis and hydroxyapatite coated hip prostheses: five year results from the International ABG Study Group. J Arthroplasty 13:660

    Article  Google Scholar 

  2. Ballard WT, Callaghan JJ, Sullivan PM, Johnston RC (1994) The results of improved cementing techniques for total hip arthroplasty in patients less than fifty years old. J Bone Joint Surg 76A:959–964

    Google Scholar 

  3. Brooker AF, Bowerman JW, Robinson RA, Riley LH (1973) Ectopic ossification following total hip replacement. J Bone Joint Surg 55A:1629–1632

    Google Scholar 

  4. DeLee JG, Charnley J (1976) Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop 121:20–32

    PubMed  Google Scholar 

  5. Duffy P, Sher JL, Partington PF (2004) Premature wear and osteolysis in an HA-coated, uncemented total hip arthroplasty. J Bone Joint Surg Br 86B(1):34–38

    Google Scholar 

  6. Eckardt A, Aberman HM, Cantwell HD, Heine J (2003) Biological fixation of hydroxyapatite-coated versus grit-blasted titanium hip stems: a canine study. Arch Orthop Trauma Surg 123(1):28–35

    PubMed  Google Scholar 

  7. Eckardt A, Delank K-St, Drees P, Menzel N, Schwitalle M, Kriegsmann J, Heine J (2001) 5-Jahres-Ergebnisse nach zementfreier Hüftendoprothetik mit der ABG-Prothese. Abrieb und acetabuläre Osteolysen. Orthopädische Praxis 32

  8. Engh CA, Gloss FE, Bobyn JD (1985) Biologic fixation arthroplasty in the treatment of osteonecrosis. Orthop Clin North Am 16:771–787

    PubMed  CAS  Google Scholar 

  9. Giannikas KA, Din R, Sadiq S, Dunningham TH (2002) Medium-term results of the ABG total hip arthroplasty in young patients. J Arthroplasty 17(2):184–188

    Article  PubMed  Google Scholar 

  10. Gruen TA, McNiece JM, Amstutz HC (1979) Modes of failure of cemented stem type femoral components: a radiographic analysis of loosening. Clin Orthop 141:17–29

    PubMed  Google Scholar 

  11. Hansen T, Otto M, Buchhorn GH, Scharnweber D, Gaumann A, Delank K-S, Eckardt A, Willert HG, Kriegsmann J, Kirkpatrick CJ (2002) New aspects in the histological examination of polyethylene wear particles in failed total joint replacements. Acta Histochem 104(3):263–269

    Article  PubMed  Google Scholar 

  12. Krismer M, Bauer R, Tschupik J, Mayerhofer P (1995) EBRA: a method to measure migration of acetabular components. J Biomech 28:1225

    Article  PubMed  CAS  Google Scholar 

  13. Merle d’Aubigné R, Postel M (1954) Functional results of hip arthroplasty with acrylic prosthesis. J Bone Joint Surg Am 36:451–475

    Google Scholar 

  14. Nourissat C, Afrey J, Berteaux D, et al. (1998) ABG Scientific Group: clinical results at seven years. In: Meeting of the International ABG Scientific Group

  15. Oosterbos CJM, Rahmy AIA, Tonino AJ (2001) Hydroxyapatite coated hip prosthesis followed up for 5 years. Int Orthop 25:17–21

    Article  PubMed  CAS  Google Scholar 

  16. Rogers A, Kulkarni R, Downes EM (2003) The ABG hydroxyapatite-coated hip prosthesis: one hundred consecutive operations with average 6-year follow-up. J Arthroplasty 18(5):619–625

    Article  PubMed  CAS  Google Scholar 

  17. Rossi P, Sibelli P, Fumero S, Crua E (1995) Short-term results of hydroxyapatite-coated primary total hip arthroplasty. Clin Orthop 310:98–102

    PubMed  Google Scholar 

  18. Russe W (1988) Röntgenphotogrammetrie der künstlichen Hüftgelenkspfanne: Aktuelle Probleme. In: Saegresser M, Burri C, Harder F, Bauer R (eds) Chirurgie und Orthopädie. Hans Huber, Bern, pp 32

    Google Scholar 

  19. Schmalzried TP, Shepherd EF, Dorey FJ, Jackson WO, dela Rosa M, Fa’vae F, McKellop HA, McClung CD, Martell J, Moreland JR, Amstutz HC (2000) Wear is a function of use, not time. Clin Orthop 381:36–46

    PubMed  Google Scholar 

  20. Sullivan PM, MacKenzie JR, Callaghan JJ, Johnston RC (1994) Total hip arthroplasty with cement in patients who are less than fifty years old. J Bone Joint Surg 76A:863–869

    Google Scholar 

  21. Tonino AJ, Rahmy AI (2000) The hydroxyapatite-ABG hip system: 5- to 7-year results from an international multicentre study. The International ABG Study Group. J Arthroplasty 15(3):274–282

    Article  PubMed  CAS  Google Scholar 

  22. Tonino AJ, Romanini L, Rossi P, Borroni M, Greco F, Garcia-Araujo C, Garcia-Dihinx L, Murcia-Mazon A, Hein W, Anderson J (1995) Hydroxyapatite coated hip prostheses: early results from an international study. Clin Orthop 312:211–225

    PubMed  Google Scholar 

  23. Zweymüller RA, Lintner FR, Selmitsch MF (1988) Biologic fixation of a press-fit titanium hip joint endoprosthesis. Clin Orthop 235:195–206

    PubMed  Google Scholar 

Download references

Acknowledgements

The authors thank Prof. Dr. H.-G. Willert and Dipl.-Ing. G.H. Buchhorn of Goettingen University, and Dr. R.M. Streicher at Stryker SA for the support and consultation for the joint claims analysis.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Karl-Stefan Delank.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Delank, KS., Drees, P., Menzel, N. et al. Increased polyethylene wear after cementless ABG I total hip arthroplasty. Arch Orthop Trauma Surg 126, 509–516 (2006). https://doi.org/10.1007/s00402-006-0168-8

Download citation

  • Received:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-006-0168-8

Keywords

Navigation