Abstract
We reported a rare revision case necessitated by massive metallosis of a metal head due to a fragmented ceramic screw which had been extraarticularily used in the primary THA here and reviewed previous literature about metallosis of metal heads associated with ceramic materials. The ceramic screw had been used for the fixation of a bone graft and reattachment of the greater trochanter in the primary THA. Scanning electron microscope analysis revealed large alumina ceramic particles embedded on the bearing surface of the polyethylene. It has previously been pointed out that metallosis of an inner head occurred in bipolar case due to a fracture of the ceramic screws in contact with a proximally migrated outer head. However, this case showed that if a ceramic screw fragmented even though it was used outside of the hip joint, the fragments could migrate into bearing surface and cause serious metallosis of the metal head. Thus, we recommend that surgeons should pay special attention to the radiographic signs of ceramic failure or metallosis such as a change of the metal head contour, if an aseptic loosening or severe osteolysis is observed in a THA in which ceramic materials were used near the hip joint. And, when a surgical revision is planned, the surgeon should be prepared to exchange with a new ceramic head.
Similar content being viewed by others
References
Allain J, Roudot-Thoraval F, Delecrin J, Anract P, Migaud H, Goutallier D (2003) Revision total hip arthroplasty performed after fracture of a ceramic femoral head. A multicenter survivorship study. J Bone Joint Surg Am 85:825
Besong AA, Hailey JL, Ingham E, Stone M, Wroblewski BM, Fisher J (1997) A study of the combined effects of shelf ageing following irradiation in air and counterface roughness on the wear of UHMWPE. Biomed Mater Eng 7:59
Bohler M, Mochida Y, Bauer TW, Plenk H Jr, Salzer M (2000) Wear debris from two different alumina-on-alumina total hip arthroplasties. J Bone Joint Surg Br 82:901
Davidson JA, Poggie RA, Mishra AK (1994) Abrasive wear of ceramic, metal, and UHMWPE bearing surfaces from third-body bone, PMMA bone cement, and titanium debris. Biomed Mater Eng 4:213
Iizuka T, Fujimoto H, Ono T (1987) A new material (single crystal sapphire screw) for internal fixation of the mandibular ramus. J Craniomaxillofac Surg 15:24
Ishizuki M, Furuya K (1991) Clinical application of sapphire pins as an internal fixation device for the upper extremity. J Hand Surg [Am] 16:922
Kempf I, Semlitsch M (1994) Massive wear of a steel ball head by ceramic fragments in the polyethylene acetabular cup after revision of a total hip prosthesis with fractured ceramic ball. Med Prog Technol 20:185
Matsuda Y, Yamamuro T (1990) Metallosis due to abnormal abrasion of the femoral head in bipolar hip prosthesis. Implant retrieval and analysis in six cases. Arch Orthop Trauma Surg 109:284
Matziolis G, Perka C, Disch A (2003) Massive metallosis after revision of a fractured ceramic head onto a metal head. Arch Orthop Trauma Surg 123:48
McNie CM, Barton DC, Fisher J, Stone MH (2000) Modeling of damage to articulating surfaces by third body particles in total joint replacements. J Mater Sci: Mater Med 11:569
Ooi Y, Suka T, Satoh Y, Shibuya K, Ohki I, Masubuchi M, Mikanagi K (1982) Clinical experience of Al2O3 ceramics as a surgical implant. Int Orthop 6:93
Shikita T, Inoue A, Hamaguchi T, Oonishi H (1980) Experience of peripheral osteosynthesis using single crystal alumina ceramics screws. Z Orthop Ihre Grenzgeb 118:975
Su EP, Callander PW, Salvati EA (2003) The bubble sign: a new radiographic sign in total hip arthroplasty. J Arthroplasty 18:110
Tipper JL, Hatton A, Nevelos JE, Ingham E, Doyle C, Streicher R, Nevelos AB, Fisher J (2002) Alumina–alumina artificial hip joints. Part II; characterisation of the wear debris from in vitro hip joint simulations. Biomaterials 23:3441
Watanabe M, Okumura H, Kihara Y, Shibata T (1993) Metallosis caused by alumina ceramic screw fixation of grafted bone in a bipolar hip endoprosthesis. Arch Orthop Trauma Surg 112:296