Abstract
Introduction
Superior cut-out of a lag screw remains a serious complication in the treatment of intertrochanteric fractures. It is related to the stability of fracture reduction. We describe the application of a trochanter supporting plate (TSP) to restore the fracture stability after early cut-out of a lag screw in unstable intertrochanteric fractures.
Materials and methods
A total of 11 consecutive patients with superior cut-out of the lag screw of a dynamic hip screw (DHS) or a gamma nail in an unstable intertrochanteric fracture occurring within 6 months after surgery were included in the present study. They underwent repeat surgery for placement of a DHS and a laterally mounted TSP of our design. All patients were monitored for at least 6 months (median 15 months; range 6–28 months).
Results
There was no repeated cut-out of a lag screw, and 10 patients (91%) achieved bony union within 5 months. At the last follow-up, all patients could walk with or without aids.
Conclusion
It reveals that a TSP, as an adjuvant to a lag screw placed inferiorly, is an easy and safe solution for the treatment of early cut-out of a lag screw in unstable intertrochanteric fractures.
Similar content being viewed by others
References
Augat P, Rapp S, Claes L (2002) A modified hip screw incorporating injected cement for the fixation of osteoporotic trochanteric fractures. J Orthop Trauma 16:311–316
Aune AK, Ekeland A, Odegaard B, Grogaard B (1994) Alho A. Gamma nail vs compression screw for trochanteric femoral fractures. Acta Orthop Scand 65:127–130
Babst R, Martinet O, Renner N, Rosso R, Bodoky A, Heberer M, Regazzoni P (1993) Die DHS Abstützplatte für die Versorgung der instabilen proximalen Femurfrakturen. Schweizer Med Wochenschr 123:566–568
Babst R, Renner N, Biedermann M, Rosso R, Heberer M, Harder F, Regazzoni P (1998) Clinical results using the trochanter stabilizing plate (TSP): the modular extension of the dynamic hip screw (DHS) for internal fixation of selected unstable intertrochanteric fractures. J Orthop Trauma 12:392–399
Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM (1995) The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am 77:1058–1064
Cheng CL, Chow SP, Pun WK, Leong JC (1989) Long-term results and complications of cement augmentation in the treatment of unstable trochanteric fractures. Injury 20:134–138
David A, Hufner T, Lewandrowski KU, Pape D, Muhr G (1996) Dynamische Hüftschraube (DHS) mit Abstützplatte—eine sichere Osteosynthese für hochinstabile “reverse” trochantere Frakturen. Chirurg 67:1166–1173
Davis TR, Sher JL, Horsman A, Simpson M, Porter BB, Checketts RG (1990) Intertrochanteric femoral fractures: mechanical failure after internal fixation. J Bone Joint Surg Br 72:26–31
DeLee JC, Heckman JD, Lewis AG (1981) Partial fibulectomy for ununited fractures of the tibia. J Bone Joint Surg Am 63:1390–1395
Desjardins AL, Roy A, Paiement G, Newman N, Pedlow F, Desloges D, Turcotte RE (1993) Unstable intertrochanteric fracture of the femur: a prospective randomized study comparing anatomical reduction and medial displacement osteotomies. J Bone Joint Surg Br 75:445–447
Jensen JS (1980) Classification of trochanteric fractures. Acta Orthop Scand 51:803–810
Leung KS, So WS, Shen WY, Hui PW (1992) Gamma nails and dynamic hip screws for peritrochanteric fractures—a randomised prospective study in elderly patients. J Bone Joint Surg Br 74:345–351
Madsen JE, Naess L, Aune AK, Alho A, Ekeland A, Stromsoe K (1998) Dynamic hip screw with trochanteric stabilizing plate in the treatment of unstable proximal femoral fractures: a comparative study with the gamma nail and compression hip screw. J Orthop Trauma 12:241–248
Mainds CC, Newman RJ (1989) Implant failures in patients with proximal fractures of the femur treated with a sliding screw device. Injury 20:98–100
Mariani EM, Rand JA (1987) Nonunion of intertrochanteric fractures of the femur following open reduction and internal fixation: results of second attempts to gain union. Clin Orthop 218:81–89
Seinsheimer F (1978) Subtrochanteric fractures of the femur. J Bone Joint Surg Am 60:300–306
Wu CC, Shih CH, Chen WJ, Tai CL (1998) Treatment of cutout of a lag screw of a dynamic hip screw in an intertrochanteric fracture. Acrh Orthop Trauma Surg 117:193–196
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Lee, PC., Yu, SW., Hsieh, PH. et al. Treatment of early cut-out of a lag screw using a trochanter supporting plate: 11 consecutive patients with unstable intertrochanteric fractures. Arch Orthop Trauma Surg 124, 119–122 (2004). https://doi.org/10.1007/s00402-003-0607-8
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00402-003-0607-8