Device for accurately siting the acetabular component during total hip replacement

  • Technical Report
  • Published:
Archives of Orthopaedic and Trauma Surgery Aims and scope Submit manuscript

Abstract

Introduction. In this paper we describe a device which allows consistent acetabular cup placement to reduce post-arthroplasty dislocation.

Patients and methods. The development of this surgical appliance is traced from the initial concept to the construction of a device suitable for clinical trials. The system of use is explained. The constraints imposed by the working environment and the methods employed to meet the demands imposed by surgical practice are elucidated. A clinical trial is described in which the device was used to assist in siting acetabular cup placement in ten patients.

Results. The accuracy of the ten cup placements done using the device was then compared by radiography with the accuracy of placement of 17 acetabular cups performed in the standard manner. Comparison of the results show a positive benefit arising from using the device.

Conclusions. Our initial good results on cup positioning in laboratory studies are maintained in a clinical setting, but further modifications could lend to even better results.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Electronic Publication

Rights and permissions

Reprints and permissions

About this article

Cite this article

Doyle, .J., Farmer, .M. & Shepard, .G. Device for accurately siting the acetabular component during total hip replacement. Arch Orthop Trauma Surg 122, 554–556 (2002). https://doi.org/10.1007/s00402-002-0405-8

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00402-002-0405-8

Navigation