Skip to main content
Log in

Is the determination of the defibrillation threshold in patients with an implantable cardioverter–defibrillator still required?

Wird die Bestimmung der Defibrillationsschwelle bei Patienten mit implantierbarem Kardioverter/Defibrillator noch benötigt?

  • Original Contribution
  • Published:
Herzschrittmachertherapie + Elektrophysiologie Aims and scope Submit manuscript

Abstract

Background

Intraoperative testing of implantable cardioverter–defibrillators (ICDs) is time consuming and associated with risks. In the present study, we elucidated whether the initial implantation of an ICD with high energy output makes intraoperative defibrillation threshold testing (DFTT) unnecessary even though antiarrhythmic (AA) therapy is needed in the future.

Methods

A total of 111 patients (94 men, 17 women) receiving an ICD with subsequent AA therapy (mexiletine, amiodarone, sotalol, flecainide) were analyzed retrospectively. DFT was performed during ICD implantation and after AA drug therapy. In a second step, DFT results from the study cohort were analyzed for implantation of virtual ICDs with either low (≤ 30 J, LOD), intermediate (34 J, IOD), or high energy output (36 J, HOD).

Results

In the study cohort, all patients reached the safety margin (SM) of 10 J between DFT and maximal shock energy of the ICD. After loading of AA agents, 6 patients (12%) with a LOD, 3 patients (11%) with an IOD, and 3 (13%) patients with a HOD failed the 10 J SM. Using virtual ICDs, 6 (5.5%) patients with a LOD, 1 patient (1%) with an IOD, and no patients with a HOD would have failed the 10 J SM. After loading of AA agents, 18 patients (16%) with a virtual LOD, 12 patients (10.8%) with an IOD, and still 9 patients (8%) with a HOD would have failed the 10 J SM.

Conclusion

Our results demonstrate that the 10 J SM would have been achieved intraoperatively in all patients with virtual HOD ICDs. Thus, determination of the DFT during implantation does not seem to be obligatory. However, in patients receiving AA agents, DFT testing is still required.

Zusammenfassung

Hintergrund

Die intraoperative Testung von implantierbaren Kardioverter/Defibrillatoren (ICD) ist personalintensiv und mit Komplikationen vergesellschaftet. In der vorliegenden Studie untersuchten wir, ob bei ICD-Implantationen mit unterschiedlicher maximaler Energieabgabe eine intraoperative Testung inklusive der Induktion von Kammerflimmern (DFT) notwendig gewesen wäre.

Material und Methoden

In einer retrospektiven Studie erhielten 111 Patienten (94 männlich, 17 weiblich) einen ICD. Nach Auftreten von ventrikulären Tachyarrhythmien wurden im weiteren Verlauf alle Patienten mit antiarrhythmischen Substanzen aufgesättigt (Mexiletin, Amiodaron, Sotalol, Flecainid). DFT-Testungen wurden intraoperativ und nach Aufsättigung mit antiarrhythmischen Substanzen durchgeführt. In einem zweiten Schritt wurden die DFT-Resultate der Studienkohorte analysiert hinsichtlich virtueller ICDs mit unterschiedlicher maximaler Energieabgabe ≤ 30 J (LOD), 34 J (IOD) oder 36 J (HOD).

Ergebnisse

Während Implantation erreichten alle Patienten die Sicherheitsmarge von 10 J (10-J-SM) zwischen maximaler Energieabgabe und DFT. Nach Aufsättigung mit antiarrhythmischen Substanzen verpassten 6 Patienten (12%) mit LOD, 3 Patienten (11%) mit IOD und 3 Patienten (13%) mit HOD die 10-J-Sicherheitsmarge. Hätte man für die Kohorte virtuelle ICDs mit LOD, IOD oder HOD implantiert, so hätten sechs (5,5%), ein (1%) bzw. kein Patient(en) (0%) die 10-J-SM nicht erreicht. Im Falle einer Aufsättigung der Studienkohorte mit antiarrhythmischen Substanzen hätten 18 (16%), 12 (10,8%) bzw. 9 (8%) der Patienten mit virtuellen ICDs (LOD/IOD/HOD) die 10-J-SM verpasst.

Schlussfolgerung

Die vorliegende Studie zeigt, dass intraoperativ bei allen Patienten mit virtuellem HOD-ICD die 10-J-SM erreicht worden wäre, jedoch nicht nach Aufsättigung mit antiarrhythmischen Substanzen. Eine intraoperative Testung aller Patienten mit HOD scheint entsprechend verzichtbar, solange keine Aufsättigung mit einem Antiarrhythmikum erfolgt.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Bardy GH, Lee KL, Mark DB et al (2005) Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) Investigators. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 352:225–237

    Article  PubMed  CAS  Google Scholar 

  2. Bhandari AK, Isber N, Estioko M et al (1998) Efficacy of low-energy T wave shocks for induction of ventricular fibrillation in patients with implantable cardioverter defibrillators. J Electrocardiol 3:31–37

    Article  Google Scholar 

  3. Moss AJ, Zareba W, Hall WJ et al (2002) Multicenter Automatic Defibrillator Implantation Trial II Investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 346:877–883

    Article  PubMed  Google Scholar 

  4. Moss AJ, Hall WJ, Cannom DS et al (1996) Improved survival with an implantable defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. N Engl J Med 335:1933–1940

    Article  PubMed  CAS  Google Scholar 

  5. Gauder D, Vohl MC, Perron P et al (1997) The Antiarrhythmics Versus Implantable Defibrillator Investigators. A comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from near-fatal arrhythmias. N Engl J Med 337:1576–1583

    Article  Google Scholar 

  6. Blatt JA, Poole JE, Johnson GW et al (2008) No benefit from defibrillation threshold testing in the SCD-HeFT (Sudden Cardiac Death in Heart Failure Trial). J Am Coll Cardiol 52:551–556

    Article  PubMed  Google Scholar 

  7. Bakker P, Viens E, Vries J de, Bredee J (1994) Impact of defibrillation threshold testing on cerebral circulation and oxygenation. PACE 17:802

    Google Scholar 

  8. Benedini G, Marchini A, Curnis A et al (1995) Implantable defibrillator and thromboembolic event. PACE, 18:199–202

  9. Tokano T, Bach D, Chang J et al (1998) The effect of ventricular shock strength on caridac hemodynamics. J Cardiovasc Electrophysiol 9:791–797

    Article  PubMed  CAS  Google Scholar 

  10. Hall B, Jeevanantham V, Levine E et al (2007) Comparison of outcomes in patients undergoing defibrillation threshold testing at the time of implantable cardioverter-defibrillator implantation versus no defibrillation threshold testing. Cardiol J 14:463–469

    PubMed  Google Scholar 

  11. Russo AM, Sauer W, Gerstenfeld EP et al (2005) Defibrillation threshold testing: is it really necessary at the time of implantable cardioverterdefibrillator insertion? Heart Rhythm 2:456–461

    Article  PubMed  Google Scholar 

  12. Epstein AE, Ellenbogen KA, Kirk KA et al (1992) Clinical characteristics of patients with high defibrillation thresholds. Circulation 86:1206–1216

    PubMed  CAS  Google Scholar 

  13. Theuns DA, Szili-Torok T, Jordaens LJ (2005) Defibrillation efficacy testing: long-term follow-up and mortality. Europace 7:509–515

    Article  PubMed  Google Scholar 

  14. DeSouza C, Koehler J, DeGroot P (1998) Failing defibrillator implant criterion on the initial attempt is a predictive factor for sudden cardiac death. Pacing Clin Electrophysiol 21:801 (abstract)

    Google Scholar 

  15. Shukla HH, Flaker GC, Jayam V, Roberts D (2003) High defibrillation thresholds in transvenous biphasic implantable defibrillators: Clinical predictors and prognostic implications. Pacing Clin Electrophysiol 26:44–48

    Article  PubMed  Google Scholar 

  16. Gula LJ, Massel D, Krahn AD, Yee R et al (2008) Is defibrillation testing still necessary? A decision analysis and Markov model. J Cardiovasc Electrophysiol 19:400–405

    Article  PubMed  Google Scholar 

  17. Gold MR, Higgins S, Klein R et al (2002) Efficacy and temporal stability of reduced safety margins for ventricular defibrillation: Primary results from the Low Energy Safety Study (LESS). Circulation 105(17):2043–2048

    Article  PubMed  Google Scholar 

  18. Manz M, Jung W, Luderitz B (1994) Interactions between drugs and devices: experimental and clinical studies. Am Heart J 127(4 Pt 2):978–984

    Article  PubMed  CAS  Google Scholar 

  19. Carnes CA, Mehdirad AA, Nelson SD (1998) Drug and defibrillator interactions. Pharmacotherapy 18(3):516–525

    PubMed  CAS  Google Scholar 

  20. Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators (1997) A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from nearfatal ventricular arrhythmias. The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. N Engl J Med 337(22):1576–1583

    Article  Google Scholar 

  21. Pelosi F Jr, Oral H, Kim MH et al (2000) Effect of chronic amiodarone therapy on defibrillation energy requirements in humans. J Cardiovasc Electrophysiol 11(7):736–740

    Article  PubMed  Google Scholar 

  22. Hohnloser SH, Dorian P, Roberts R et al (2006) Effect of amiodarone and sotalol on ventricular defibrillation threshold: the optimal pharmacological therapy in cardioverter defibrillator patients (OPTIC) trial. Circulation 114(2):104–109

    Article  PubMed  CAS  Google Scholar 

  23. Zhou L, Chen BP, Kluger J et al (1998) Effects of amiodarone and its active metabolite desethylamiodarone on the ventricular defibrillation threshold. J Am Coll Cardiol 31(7):1672–1678

    Article  PubMed  CAS  Google Scholar 

  24. Dorian P (2000) Amiodarone and defibrillation thresholds: a clinical conundrum. J Cardiovasc Electrophysiol 11(7):741–743

    Article  PubMed  CAS  Google Scholar 

  25. Kuhlkamp V, Mewis C, Suchalla R (1999) Effect of amiodarone and sotalol on the defibrillation threshold in comparison to patients without antiarrhythmic drug treatment. Int J Cardiol 69(3):271–279

    Article  PubMed  CAS  Google Scholar 

  26. Jung W, Manz M, Pizzulli L et al (1992) Effects of chronic amiodarone therapy on defibrillation threshold. Am J Cardiol 70(11):1023–1027

    Article  PubMed  CAS  Google Scholar 

  27. Swerdlow CD, Russo AM, Degroot PJ (2007) The dilemma of ICD implant testing. Pacing Clin Electrophysiol 30:675–700

    Article  PubMed  Google Scholar 

  28. Goldschlager N, Epstein A, Friedman P et al (2001) Environmental and drug effects on patients with pacemakers and implantable cardioverter/defibrillators: a practical guide to patient treatment. Arch Intern Med 161(5):649–655

    Article  PubMed  CAS  Google Scholar 

  29. Rajawat YS, Dias D, Gerstenfeld EP et al (2002) Interactions of antiarrhythmic drugs and implantable devices in controlling ventricular tachycardia and fibrillation. Curr Cardiol Rep 4(5):434–440

    Article  PubMed  Google Scholar 

  30. Topham SL, Cha YM, Peters BB, Chen PS (1992) Effects of lidocaine on relation between defibrillation threshold and upper limit of vulnerability in open-chest dogs. Circulation 85(3):1146–1151

    PubMed  CAS  Google Scholar 

  31. Peters RW, Gilbert TB, Johns-Walton S, Gold MR (1997) Lidocaine related increase in defibrillation threshold. Anesth Analg 85(2):299–300

    PubMed  CAS  Google Scholar 

  32. Ellenbogen KA, Kay GN, Lau CP, Wilkoff BL (eds) (2007) Clinical cardiac pacing, defibrillation, and resynchronization therapy, 3rd edn. Saunders, Philadelphia, PA

  33. Gold MR, Olsovsky MR, Pelini MA et al (1998) Comparison of single- and dual-coil acive pectoral defibrillation lead systems. J Am Coll Cardiol 31:1391–1394

    Article  PubMed  CAS  Google Scholar 

  34. Birnie D, Tung S, Simpson D et al (2008) Complications associated with defibrillation threshold testing: the Canadian experience. Heart Rhythm 5:387–390

    Article  PubMed  Google Scholar 

  35. Grubb BP, Durzinsky D, Brewster P et al (1997) Sudden cerebral vasoconstriction during induced polymorphic ventricular tachycardia and fibrillation: further observations of a paradoxic response. Pacing Clin Electrophysiol 20:1667–1672

    Article  PubMed  CAS  Google Scholar 

  36. Meyer J, Mollhoff T, Seifert T et al (1996) Cardiac output is not affected during intraoperative testing of the automatic implantable cardioverter defibrillator. J Cardiovasc Electrophysiol 7:211–216

    Article  PubMed  CAS  Google Scholar 

  37. Murkin JM, Baird DL, Martzke JS et al (1997) Cognitive dysfunction after ventricular fibrillation during implantable cardiovertor/defibrillator procedures is related to duration of the reperfusion interval. Anesth Analg 84:1186–1192

    PubMed  CAS  Google Scholar 

Download references

Conflict of interest

The corresponding author states that there are no conflicts of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to A. Bauer.

Additional information

*The first two authors contributed equally to this work.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Licka, M., Jahn, L., Kelemen, K. et al. Is the determination of the defibrillation threshold in patients with an implantable cardioverter–defibrillator still required?. Herzschr. Elektrophys. 22, 209–213 (2011). https://doi.org/10.1007/s00399-011-0150-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00399-011-0150-1

Keywords

Schlüsselwörter

Navigation