Comparison of Pura-Vario and Palmaz-Schatz stents following implantation using normal and high pressure in pigs: immediate and late results assessed by 3-dimensional IVUS

  • STENTS/PTCA
  • Published:
Zeitschrift für Kardiologie Aims and scope Submit manuscript

Zusammenfassung

Einleitung: Der Pura-Vario-Stent weist 2 neuartige “Brückensegmente” zur Erhöhung der longitudinalen Flexibilität auf. Hierdurch soll die Plazierung und Implantation des Stents vereinfacht und sicherer werden.

Versuchsziel und Methoden: Ziel der vorliegenden tierexperimentellen Untersuchung war es, die Stentexpansion des Pura-Vario-Stents (PV) zu analysieren, und sie mit dem Palmaz-Schatz-Stent (PS) zu vergleichen. Außerdem wurde die Stentimplantation mit “Hochdruck” (18 atm) (HP) mit der “Normaldruckimplantation” (12 atm)(NP) verglichen. Insgesamt 16 Stents wurden randomisiert in die “left anterior descending artery” (LAD) und in den “Ramus circumflexus” (CX) von 8 Schweinen implantiert. Stentflächen, endoluminale Flächen und Stentasymmetrie wurden mittels dreidimensionalem intravaskulärem Ultraschall (3D-IVUS) zu folgenden Zeitpunkten vermessen: unmittelbar nach Stentimplantation (Tag 0) und nach 14 Tagen (Tag 14).

Resultate: Die Stentexpansion im Längsschnitt war nicht uniform: die Brückensegmente beider Stentmodelle waren größer als die dazwischengelegten “Rautensegmente” am Tag 0. Diese Differenz war nach 14 Tagen jedoch aufgehoben. Trotz der höheren Flexibilität des Pura-Vario-Stents war die Stentexpansion beider Modelle gleich, sowohl unmittelbar nach Implantation (Lumenfläche: 9,75 ± 0,28 mm2 (PV) vs. 9,82 ± 0,34 mm2 (PS)) als nach 14 Tagen (7,44 ± 0,16 mm2 vs. 7,45 ± 0,22 mm2). Die Pura-Vario-Stents waren jedoch im Querschnitt symmetrischer. “Hochdruckimplantation” führte lediglich am Tag 0 zu einer größeren und symmetrischeren Stentexpansion (Lumenfläche: 9,54 ± 0,39 mm2 (HP) vs. 8,77 ± 0,33 mm2 (NP) (p < 0,05)); dieser Vorteil ging jedoch bis zur Nachuntersuchung vollständig verloren, was vor allem auf den höheren Stent-Recoil in dieser Gruppe zurückzuführen war.

Schlußfolgerung: Trotz der höheren Flexibilität des Pura-Vario-Stents waren die Expansionscharakteristika beider Modelle vergleichbar. “Hochdruckimplantation” erbrachte bessere Ergebnisse lediglich an Tag 0, nach 14 Tagen fand sich jedoch kein Unterschied mehr zu den mit “Normaldruck” implantierten Stents.

Summary

Introduction: The Pura-Vario stent features 2 newly designed “bridging segments” for enhanced longitudinal flexibility, in order to allow easier and safer stent implantation.

Methods: The aim of the present experimental investigation was to analyze the expansion characteristics of the Pura-Vario-Stent (PV), and to compare it with the Palmaz-Schatz stent (PS). Furthermore, stent implantation using “high pressure” (18 atm) (HP) was compared with “normal pressure” (12 atm) (NP). Stents (n = 16) were implanted into the left anterior descending artery (LAD) and the circumflex artery (CX) of 8 pigs. Stent area, lumen area and stent asymmetry were measured by means of three-dimensional intravascular ultrasound (3D-IVUS): (1) immediately after implantation, and (2) at 14-days follow-up.

Results: Stent expansion was found not to be uniform: the “bridging segments” were significantly larger than the “diamond segments” in either stent model at day 0; this difference, however, disappeared at 14-days follow-up. Despite higher flexibility of the Pura-Vario stent, no difference in stent expansion was found between both stent models, neither immediately after implantation (mean lumen area: 9.75 ± 0.28 mm2 (PV) vs. 9.82 ± 0.34 mm2 (PS)), nor at 14-days follow-up (7.44 ± 0.16 mm2 (PV) vs. 7.45 ± 0.22 mm2 (PS)). Pura-Vario stents, however, were less asymmetric in the cross-sectional view. Implantation using “high pressure” resulted in larger and less asymmetric stent expansion only at day 0 (lumen area: 9.54 ± 0.39 mm2 (HP) vs. 8.77 ± 0.33 mm2 (NP) (p < 0.05)); this difference, however, disappeared after 14 days due to higher stent recoil in the “high pressure” group.

Conclusion: Despite higher flexibility of the Pura-Vario stent, expansion characteristics of both stent models were comparable. “High pressure” implantation compared favorably with “normal pressure” implantation only at day 0, whereas no difference could be found between both techniques at 14-days follow-up.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Author information

Authors and Affiliations

Authors

Additional information

Eingegangen: 18. Februar 1999, Akzeptiert: 4. Juni 1999

Rights and permissions

Reprints and permissions

About this article

Cite this article

von Degenfeld, G., Heinrich, D., Giehrl, W. et al. Comparison of Pura-Vario and Palmaz-Schatz stents following implantation using normal and high pressure in pigs: immediate and late results assessed by 3-dimensional IVUS. Z Kardiol 88, 906–913 (1999). https://doi.org/10.1007/s003920050368

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s003920050368

Navigation