Abstract
Purpose
Fixation of cranial bones in pediatric craniofacial surgery with biodegradable materials has developed into an accepted method. However, placing the fixation material on the outer surface of the cranial bone at the frontal cranium in infants can result in suboptimal cosmetic outcomes, as the plates and screws can be palpable. The placement of resorbable fixation devices on the inner surface of the skull would allow for less obvious fixation beneath the skin with a potentially superior cosmetic result. The authors report the use of such resorbable fixation devices on the inner or endocranial aspect of the cranium which appears to be novel.
Materials and methods
Ten patients with the mean age of 14 months (range 7–35 months) were treated with cranial remodeling using poly(lactide-co-glycolide) (PLGA) biodegradable fixation on the inner surface of the cranial bones. Five patients had trigonocephaly, four plagiocephaly, and one brachycephaly. All patients had 3D computed tomography (CT) and clinical photographs done preoperatively and postoperatively at follow-up. The outcome was judged at follow-up by clinical evaluation, photographs, 3D CT, and interview of the parents. The mean follow-up time was 3.5 years (range 0.5–6 years).
Results
The primary recovery was uneventful in all cases. No wound infection occurred, but one patient had a minor skin necrosis which required a single revision operation to correct. The cosmetic outcome was scored as excellent, good, fair, or poor. There was no case with delayed union. Seven out of ten cases were judged as excellent and three as good, none as fair or poor.
Conclusions
The use of resorbable PLGA fixation devices on the inner aspect of the skull appears to provide a satisfactory cosmetic result in this small preliminary group of pediatric patients. Further long-term study of these materials in this specific location in a larger patient group is needed.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.References
Losee JE, Mason AC (2005) Deformational plagiocephaly: diagnosis, prevention, and treatment. Clin Plast Surg 32(1):53–64
Serlo W, Ashammakhi N, Törmälä P, Waris T (2000) Case report: a new technique for correction of trigonocephaly in an infant: application of an absorbable endocranial plate. Childs Nerv Syst 16:595–597
Fiala T, Novelline R, Yaremchuk M (1993) Comparison of CT imaging artefacts from craniomaxillofacial internal fixation devices. Plast Reconstr Surg 92(7):1227–1232
Fiala T, Paige K, Davis T, Campbell T, Rosen B, Yaremchuk M (1994) Comparison of artefact from craniomaxillofacial internal fixation devices: magnetic resonance imaging. Plast Reconstr Surg 93(4):725–731
Eppley B, Sparks C, Herman E, Edwards M, McCarty M, Sadove M (1993) Effects of skeletal fixation on craniofacial imaging. J Craniofac Surg 4(2):67–73
Eppley B, Sadove M (1995) A comparison of resorbable and metallic fixation in healing of calvarial bone grafts. Plast Reconstr Surg 96(2):316–322
Habal M, Pietrzak W (1999) Key points in the fixation of the craniofacial skeleton with absorbable biomaterial. J Craniofac Surg 10(6):491–499
Eppley B, Reilly M (1997) Degradation characteristics of PLLA-PGA bone fixation devices. J Craniofac Surg 8(2):116–120
Eppley B, Prevel C (1997) Nonmetallic fixation in traumatic midfacial fractures. J Craniofac Surg 8(2):103–109
Tiainen J, Soini Y, Törmälä P, Waris T, Ashammakhi N (2004) Self-reinforced polylactide-polyglycolide 80/20 screws take more than 1 1/2 years to resorb in rabbit cranial bone. J Biomed Mater Res (Appl Biomat) 70B(1):49–55
Ylikontiola LP, Ashammaki N, Törmälä P, Sundquist K, Sàndor GKB (2004) Self-reinforced bio-absorbable polylactide (SR-PLD/LA) 70/30 miniplates and miniscrews are reliable for fixation of anterior mandibular fractures. Oral Surg Oral Med Oral Pathol Oral Radiol Endodont 97(3):312–317
Weisberger E, Eppley B (1997) Resorbable fixation plates in head and neck surgery. Laryngoscope 107:716–719
Wiley RC, Sarver DR (1993) Method for heating biocompatible implants in a thermal packaging line. US Patent 5,263,991, 23 Nov 1993
Tschakaloff A (1994) Surgical system. US Patent 5,290,281, 01 Mar 1994
Piettrzak W, Sarver D, Bianchini S, D’Alessio K (1997) Effect of simulated intraoperative heating and shaping on mechanical properties of a bioabsorbable fracture plate material. J Biomed Mater Res (Appl Biomater) 38:17–24
Eppley B, Sadove M (1992) Effects of resorbable fixation on craniofacial skeletal growth: A pilot experimental study. J Craniofac Surg 3(4):190–196
Eppley B, Sadove M (1994) Effects of resorbable fixation on craniofacial skeletal growth: modifications in plate size. J Craniofac Surg 5(2):110–114
Montag M, Morales L, Daane S (1997) Bioabsorbables: their use in pediatric craniofacial surgery. J Craniofac Surg 8(2):100–102
Goldstein J, Quereshy F, Cohen A (1997) Early experience with biodegradable fixation for congenital pediatric craniofacial surgery. J Craniofac Surg 8(2):110–115
Kumar A, Staffenberg D, Petronio J, Wood R (1997) Bioabsorbable plates and screws in pediatric craniofacial surgery: a review of 22 cases. J Craniofac Surg 8:97–99
Enislidis G, Pichorner S, Kainberger F, Ewers R. (1997) Lactosorb panel and screws for repair of large orbital floor defects. J Craniomaxillofac Surg 25(6):316–321
Rinehart G, Pittman T (1998) Growing skull fractures: strategies for repair and reconstruction. J Craniofac Surg 9(1):65–72
Tharanon W, Sinn D, Hobar C, Sklar F, Salomon J (1998) Surgical outcomes using bioabsorbable plating systems in pediatric craniofacial surgery. J Craniofac Surg 9(5):441–444
Goldstein JA (2001) The use of bioresorbable material in craniofacial surgery. Clin Plast Surg 28(4):653–659
Imola MJ, Hamlar DD, Shao W, Chowdhury K, Tatum S (2001) Resorbable plate fixation in pediatric craniofacial surgery: long-term outcome. Arch Facial Plast Surg 3(2):79–90
Cohen SR, Holmes RE, Meltzer HS, Nakaji P.(2002) Immediate cranial vault reconstruction with bioresorbable plates following endoscopically assisted sagittal synostectomy. J Craniofac Surg 13(4):578–582
Serlo W, Ashammakhi N, Lansman S, Törmälää P, Waris T (2003) A new technique for correction of trigonocephaly using bioabsorbable osteofixation tacks and plates and a novel tack-shooter. J Craniofac Surg 14(1):92–96
Spanio S, Ashammakhi N, Ilomaki J, Valimaa T, Rainer C, Waris T, Törmälä P, Ninkovic M (2002) Use of new bioabsorbable tacks and a tackshooter in cranial bone osteofixation saves operative time. J Craniofac Surg 13(5):693–696
Iera D, Haddad AJ, Sàndor GKB, Ashammakhi N (2005) Les matériels d’osteosynthèse résorbales (Bioresorebable osteofixation devices). Annales de Chirgurie Plastiques Esthétique 50:723–732
Antikainen T, Pernu H, Törmälä P et al (1994) Development of the neurocranium after transsutural fixing by new, resorbable poly-l-lactide miniplates. A comparison to fixing with the common titanium miniplates. Acta Neurochir (Wien) 128:26–31
Waris E, Ashammakhi N, Kelly CP, Andrus L, Waris T, Jackson IT. (2005) Transphyseal bioabsorbable screws cause temporary growth retardation in rabbit femur. J Pediatr Orthop 25(3):342–345
Ashammakhi N, Renier D, Arnaud E, Marchac D, Ninkovic M, Donaway D, Jones B, Serlo W, Laurikainen K, Törmälä P, Waris T (2004) Successful use of biosorb osteofixation devices in 165 cranial and maxillofacial cases: a multicenter report. J Craniofac Surg 15(4):692–701
Vesala AL, Kallioinen M, Törmälä P, Kellomaki M, Waris T, Ashammakhi N (2002) Bone tissue engineering: treatment of cranial bone defects in rabbits using self-reinforced poly-l,d-lactide 96/4 sheets. J Craniofac Surg 13(5):607–613
Ashammakhi N, Makela EA, Vihtonen K, Rokkanen P, Törmälä P (1995) Repair of bone defects with absorbable membranes. A study on rabbits. Ann Chir Gynaecol 84(3):309–315
Haddad AJ, Peel SA, Clokie CML, Sándor GKB (2006) Closure of rabbit calvarial critical-sized defects using protective composite allogeneic and alloplastic bone substitutes. J Craniofac Surg 17(5):926–934
Serlo W, Kaarela OI, Peltoniemi HH, Merikanto J, Ashammakhi NA, Lassila K, Pohjonen T, Törmälä P, Waris TH (2001) Use of self-reinforced polylactide osteosynthesis devices in craniofacial surgery: a long-term follow-up study. Scand J Plast Reconstr Surg Hand Surg 35(3):285–292
Ashammakhi N, Moreira A, Gonzalez AM, Törmälä P, Jackson IT (2004) Review: new resorbable bone fixation biomaterials in craniomaxillofacial surgery: Present and future. Eur J Plast Surg 26(8):383–390
Ashammakhi N, Peltoniemi H, Waris E, Suuronen R, Serlo W, Kellomaki M, Törmälä P, Waris T. (2001) Developments in craniomaxillofacial surgery: use of self-reinforced bioabsorbable osteofixation devices. Plast Reconstr Surg 108(1):167–180
Mackool R, Yim J, McCarthy J (2006) Delayed degradation in a resorbable plating system. J Craniofac Surg 17(1):194–197
Bergsma JE, De Bruijn WC, Rozema FR et al (1995) Late degradation tissue response to poly(l-lactide) bone plates and screws. Biomaterials 16:25–31
Acknowledgments
The authors wish to acknowledge the invaluable efforts of Ms Marianne Forsberg in the preparation of this manuscript.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Serlo, W.S., Ylikontiola, L.P., Vesala, AL. et al. Effective correction of frontal cranial deformities using biodegradable fixation on the inner surface of the cranial bones during infancy. Childs Nerv Syst 23, 1439–1445 (2007). https://doi.org/10.1007/s00381-007-0470-y
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00381-007-0470-y