Abstract.
A method of determination of the true temporal response function for pump-probe type of experiments with femtosecond time resolution is presented. An analytical formula allowing calculation of group velocity dispersion (GVD) modified and sample thickness-sensitive pump-probe cross correlation function is provided. The reliability of the formula is further experimentally verified with measurements of ultrafast stimulated Raman scattering and transient absorption signals.
Similar content being viewed by others
Author information
Authors and Affiliations
Additional information
Received: 31 October 2000 / Revised version: 6 February 2001 / Published online: 27 April 2001
Rights and permissions
About this article
Cite this article
Ziólek, M., Lorenc, M. & Naskrecki, R. Determination of the temporal response function in femtosecond pump-probe systems . Appl Phys B 72, 843–847 (2001). https://doi.org/10.1007/s003400100587