Abstract
We have performed a comparison of laser (LsOPM) and lamp (LpOPM) pumped cesium vapor magnetometers. Although the LsOPM operated 50% above its shot-noise limit we found an intrinsic sensitivity of 15 fT/√Hz and 25 fT/√Hz for the LsOPM and the LpOPM, respectively. Two modes of operation, viz. the phase-stabilized and the self-oscillating modes, were investigated and found to yield a similar performance. We have compared the performance of the LsOPM and the LpOPM directly by simultaneous measurements of field fluctuations of a 2-μT magnetic field inside a multilayer magnetic shield and have used one of the magnetometers for an active field stabilization. In the stabilized mode we found a gradient instability of 25 fT within an integration time of 100 s, which represents an upper limit of the long-term stability of the magnetometers. Our research is motivated by the need for an improved control of magnetic fields and gradients in a planned neutron electric dipole experiment.
Similar content being viewed by others
References
K. Green, P.G. Harris, P. Iaydjiev, D.J.R. May, J.M. Pendlebury, K.F. Smith, M. van der Grinten, P. Geltenbort, S. Ivanov, Nuvl. Instr. and Meth. A 404, 381 (1998)
I.S. Altarev, Y.V. Borisov, N.V. Borovikova, A.I. Egorov, S.N. Ivanov, E.A. Kolomensky, M.S. Lasakov, V.M. Lobashev, V.A. Nazarenko, A.N. Pirozhkov, A.P. Serebrov, Y.V. Sobolev, E.V. Shulgina, Phys. At. Nuclei 59, 1152 (1996)
Y. Borisov, W. Heil, M. Leduc, V. Lobashev, E. Otten, Y. Sobolev, Nucl. Instrum. Methods A 440, 48 (2000)
S. Groeger, J.-L. Schenker, R. Wynands, A. Weis, Phys. Rev. A, submitted [arXiv:physics/0406105 v2 19 August 2004]
O. Moreau, B. Cheron, H. Gilles, J. Hamel, E. Noel, J. Phys. III 7, 99 (1997)
H. Gilles, J. Hamel, B. Cheron, Rev. Sci. Instrum. 72, 2253 (2001)
A.P. Kazantsev, V.S. Smirnow, A.M. Tumaikin, I.A. Yagofarov, Opt. Spectrosc. 57, 189 (1984)
E.B. Alexandrov, M.V. Balabas, D. Budker, D. English, D.F. Kimball, C.-H. Li, V.V. Yashchuk, Phys. Rev. A 66, 042903 (2002)
G. Breit, I.I. Rabi, Phys. Rev. 38, 2082 (1931)
V.V. Yashchuk, D. Budker, J.R. Davis, Rev. Sci. Instrum. 71, 341 (2000)
G. Bison, R. Wynands, A. Weis, J. Opt. Soc. Am. B 22, 77 (2005)
E.B. Aleksandrov, M.V. Balabas, A.K. Vershovskii, A.E. Ivanov, N.N. Yakobson, V.L. Velichanskii, N.V. Senkov, Opt. Spectrosc. 78, 325 (1995)
J.A. Barnes, A.R. Chi, L.S. Cutler, D.J. Healey, D.B. Leeson, T.E. McGunigal, J.A. Mullen, Jr., W.L. Smith, R.L. Sydnor, R.F.C. Vessot, G.M.R. Winkler, IEEE Trans. Instrum. Meas. 20, 105 (1971)
D. Nettels, R. Müller-Siebert, S. Ulzega, A. Weis, Appl. Phys. B 77, 563 (2003)
E.B. Alexandrov, A.S. Pazgalev, J.L. Rasson, Opt. Spectrosc. 82, 14 (1997)
I.K. Kominis, T.W. Kornack, J.C. Allred, M.V. Romalis, Nature 422, 596 (2003)
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Groeger, S., Pazgalev, A.S. & Weis, A. Comparison of discharge lamp and laser pumped cesium magnetometers. Appl. Phys. B 80, 645–654 (2005). https://doi.org/10.1007/s00340-005-1773-x
Received:
Revised:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00340-005-1773-x