Clinical evaluation of a new stationary high strip density antiscatter grid in comparison with a conventional moving grid: influence on image quality and patient radiation dose

  • Physics
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract.

This study aimed to evaluate a new stationary 13/75 antiscatter grid in comparison with a moving 12/40 grid with regard to its influence on image quality and patient radiation dose. One hundred consecutive patients had plain films of the abdomen using a moving 12/40 grid and another 100 patients were examined with a stationary 13/75 grid (tube voltage 73 kV, film size 35 cm × 43 cm, film-screen system speed class 400). As an intraindividual comparison, radiographs of the abdomen with a film size of 18 cm × 43 cm were taken of 27 patients using both grid types for every patient. Patients' weight and dose–area product were recorded. Image quality was evaluated by a multireader study using standardized questionnaires and three- or five-level grading systems. Results showed that use of the 13/75 grid accords with the ALARA principle: measurement of the dose–area product demonstrated a dose saving of 24% for a film size of 35 cm × 43 cm and a saving of 34% for a film size of 18 cm × 43 cm without calling into question the diagnostic image quality. The mean scores of the 12/40 grid were lower for three out of six normal anatomic structures (P<0.05) among the 200 patients, indicating a loss of image quality with the 13/75 grid. Assessment of the intraindividual trial revealed reduced detectability of three out of four anatomical details when the 13/75 grid was employed. At the cost of a minor loss of image quality in comparison with the 12/40 grid, the new stationary grid provides a clear reduction of radiation dose under clinical circumstances.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Electronic Publication

Rights and permissions

Reprints and permissions

About this article

Cite this article

Wamser, G., Aichinger, H., Maier, W. et al. Clinical evaluation of a new stationary high strip density antiscatter grid in comparison with a conventional moving grid: influence on image quality and patient radiation dose. Eur Radiol 11, 1710–1719 (2001). https://doi.org/10.1007/s003300000730

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s003300000730

Navigation