Abstract
Objectives
To evaluate and compare the image quality of propagation-based phase-contrast computed tomography (PB-CT) using synchrotron radiation and conventional cone-beam breast computed tomography (CBBCT) based on various radiological image quality criteria.
Methods
Eight excised breast tissue samples of various sizes and containing different lesion types were scanned using PB-CT at a synchrotron facility and using CBBCT at a university-affiliated breast imaging centre. PB-CT scans were performed at two different mean glandular dose (MGD) levels: standard (5.8 mGy) and low (1.5 mGy), for comparison with CBBCT scans at the standard MGD (5.8 mGy). Image quality assessment was carried out using six quality criteria and six independent medical imaging experts in a reading room with mammography workstations. The interobserver agreement between readers was evaluated using intraclass correlation coefficient (ICC), and image quality was compared between the two breast imaging modalities using the area under the visual grading characteristic curve (AUCVGC).
Results
Interobserver agreement between the readers showed moderate reliability for five image criteria (ICC: ranging from 0.488 to 0.633) and low reliability for one criterion (image noise) (ICC 0.307). For five image quality criteria (overall quality, perceptible contrast, lesion sharpness, normal tissue interfaces, and calcification visibility), both standard-dose PB-CT images (AUCVGC 0.958 to 1, p ≤ .05) and low dose PB-CT images (AUCVGC 0.785 to 0.834, p ≤ .05) were of significantly higher image quality than standard-dose CBBCT images.
Conclusions
Synchrotron-based PB-CT can achieve a significantly higher radiological image quality at a substantially lower radiation dose compared with conventional CBBCT.
Key Points
• PB-CT using synchrotron radiation results in higher image quality than conventional CBBCT for breast imaging.
• PB-CT using synchrotron radiation requires a lower radiation dose than conventional CBBCT for breast imaging.
• PB-CT can help clinicians diagnose patients with breast cancer.
Similar content being viewed by others
Abbreviations
- AUCVGC :
-
Area under the visual grading characteristic curve
- CBBCT:
-
Cone-beam breast CT
- DBT:
-
Digital breast tomosynthesis
- DCIS:
-
Ductal carcinoma in situ
- ICC:
-
Intraclass correlation coefficient
- iFBP:
-
Iterative filtered back projection
- MG:
-
Mammography
- MGD:
-
Mean glandular dose
- MRMC:
-
Multiple-reader, multiple-case
- PB-CT:
-
Propagation-based phase-contrast CT
- ROC:
-
Receiver operating characteristic
- SIRT:
-
Simultaneous iterative reconstruction technique
- TIE-Hom:
-
Homogeneous transport of intensity equation
- VGC:
-
Visual grading characteristics
References
Australian Institute of Health and Welfare (2017) Australian Cancer Incidence and Mortality (ACIM) books. AIHW, Canberra
Bray F, Ferlay J, Soerjomataram I, Siegel RL, Torre LA, Jemal A (2018) Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin 68:394–424
Tavakoli Taba S, Brennan PC, Lewis S (2019) Dynamics of breast imaging research: a global scoping review and Sino-Australian comparison case study. PLoS One 14:e0210256
Independent UK Panel on Breast Cancer Screening (2012) The benefits and harms of breast cancer screening: an independent review. Lancet 380:1778–1786
Ciatto S, Houssami N, Bernardi D et al (2013) Integration of 3D digital mammography with tomosynthesis for population breast-cancer screening (STORM): a prospective comparison study. Lancet Oncol 14:583–589
Pisano ED, Gatsonis C, Hendrick E et al (2005) Diagnostic performance of digital versus film mammography for breast-cancer screening. N Engl J Med 353:1773–1783
Mandelson MT, Oestreicher N, Porter PL et al (2000) Breast density as a predictor of mammographic detection: comparison of interval- and screen-detected cancers. J Natl Cancer Inst 92:1081–1087
European Reference Organisation for Quality Assured Breast Screening and Diagnostic Services (EUREF) (2008) European guidelines for quality assurance in breast cancer screening and diagnosis. Fourth Edition. Available via https://www.euref.org/europeanguidelines/4th-edition
Australian Institute of Health and Welfare (2018) BreastScreen Australia monitoring report 2018. AIHW, Canberra
Alakhras M, Bourne R, Rickard M, Ng KH, Pietrzyk M, Brennan PC (2013) Digital tomosynthesis: a new future for breast imaging? Clin Radiol 68:e225–e236
Skaane P, Bandos AI, Gullien R et al (2013) Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology 267:47–56
Durand MA, Haas BM, Yao X et al (2015) Early clinical experience with digital breast tomosynthesis for screening mammography. Radiology 274:85–92
Taba ST, Gureyev TE, Alakhras M, Lewis S, Lockie D, Brennan PC (2018) X-ray phase-contrast technology in breast imaging: principles, options, and clinical application. AJR Am J Roentgenol 211:133–145
Dullum JR, Lewis EC, Mayer JA (2000) Rates and correlates of discomfort associated with mammography. Radiology 214:547–552
Wienbeck S, Lotz J, Fischer U (2017) Review of clinical studies and first clinical experiences with a commercially available cone-beam breast CT in Europe. Clin Imaging 42:50–59
Uhlig J, Uhlig A, Biggemann L, Fischer U, Lotz J, Wienbeck S (2019) Diagnostic accuracy of cone-beam breast computed tomography: a systematic review and diagnostic meta-analysis. Eur Radiol 29:1194–1202
Seifert P, Conover D, Zhang Y et al (2014) Evaluation of malignant breast lesions in the diagnostic setting with cone beam breast computed tomography (breast CT): feasibility study. Breast J 20:364–374
He N, Wu YP, Kong Y et al (2016) The utility of breast cone-beam computed tomography, ultrasound, and digital mammography for detecting malignant breast tumors: a prospective study with 212 patients. Eur J Radiol 85:392–403
Lindfors KK, Boone JM, Nelson TR, Yang K, Kwan AL, Miller DF (2008) Dedicated breast CT: initial clinical experience. Radiology 246:725–733
O'Connell A, Conover DL, Zhang Y et al (2010) Cone-beam CT for breast imaging: radiation dose, breast coverage, and image quality. AJR Am J Roentgenol 195:496–509
Wilkins S, Gureyev T, Gao D, Pogany A, Stevenson A (1996) Phase-contrast imaging using polychromatic hard X-rays. Nature 384:335–338
Arfelli F, Assante M, Bonvicini V et al (1998) Low-dose phase contrast x-ray medical imaging. Phys Med Biol 43:2845
Arboleda C, Wang Z, Jefimovs K et al (2019) Towards clinical grating-interferometry mammography. Eur Radiol. https://doi.org/10.1007/s00330-019-06362-x
Nesterets YI, Gureyev TE (2014) Noise propagation in x-ray phase-contrast imaging and computed tomography. J Phys D Appl Phys 47:105402
Gureyev TE, Nesterets YI, Baran PM et al (2019) Propagation-based X-ray phase-contrast tomography of mastectomy samples using synchrotron radiation. Med Phys. https://doi.org/10.1002/mp.13842
Longo R, Arfelli F, Bellazzini R et al (2016) Towards breast tomography with synchrotron radiation at Elettra: first images. Phys Med Biol 61:1634–1649
Longo R, Arfelli F, Bonazza D et al (2019) Advancements towards the implementation of clinical phase-contrast breast computed tomography at Elettra. J Synchrotron Radiat 26:1343–1353
Tavakoli Taba S, Baran P, Lewis S et al (2019) Toward improving breast cancer imaging: radiological assessment of propagation-based phase-contrast CT technology. Acad Radiol 26:e79–e89
Baran P, Pacile S, Nesterets YI et al (2017) Optimization of propagation-based x-ray phase-contrast tomography for breast cancer imaging. Phys Med Biol 62:2315–2332
Gureyev T, Mayo S, Nesterets YI et al (2014) Investigation of the imaging quality of synchrotron-based phase-contrast mammographic tomography. J Phys D Appl Phys 47:365401
Pacile S, Brun F, Dullin C et al (2015) Clinical application of low-dose phase contrast breast CT: methods for the optimization of the reconstruction workflow. Biomed Opt Express 6:3099–3112
Nesterets YI, Gureyev TE, Mayo SC et al (2015) A feasibility study of X-ray phase-contrast mammographic tomography at the Imaging and Medical beamline of the Australian Synchrotron. J Synchrotron Radiat 22:1509–1523
Brombal L, Donato S, Dreossi D et al (2018) Phase-contrast breast CT: the effect of propagation distance. Phys Med Biol 63:24 nt03
Brombal L, Golosio B, Arfelli F et al (2019) Monochromatic breast computed tomography with synchrotron radiation: phase-contrast and phase-retrieved image comparison and full-volume reconstruction. J Med Imaging (Bellingham) 6:031402
Boss A (2018) Editorial comment: cone-beam and phase contrast CT: new horizons in breast imaging? Eur Radiol 28:3729–3730
Koning Corporation (2015) User’s manual for the Koning Breast CT System User’s Manual for the Koning Breast CT System: https://www.accessdata.fda.gov/cdrh_docs/pdf13/P130025c.pdf,
Castelli E, Tonutti M, Arfelli F et al (2011) Mammography with synchrotron radiation: first clinical experience with phase-detection technique. Radiology 259:684–694
Fedon C, Longo F, Mettivier G, Longo R (2015) GEANT4 for breast dosimetry: parameters optimization study. Phys Med Biol 60:N311–N323
Mettivier G, Fedon C, Di Lillo F et al (2016) Glandular dose in breast computed tomography with synchrotron radiation. Phys Med Biol 61:569–587
Gureyev TE, Nesterets Y, Ternovski D et al (2011) Toolbox for advanced X-ray image processing. SPIE Optical Engineering + Applications. International Society for Optics and Photonics, pp 81410B-81410B-81414
Brun F, Pacilè S, Accardo A et al (2015) Enhanced and flexible software tools for X-ray computed tomography at the Italian synchrotron radiation facility Elettra. Fundam Inform 141:233–243
Paganin D, Mayo SC, Gureyev TE, Miller PR, Wilkins SW (2002) Simultaneous phase and amplitude extraction from a single defocused image of a homogeneous object. J Microsc 206:33–40
Raupach R, Flohr T (2012) Performance evaluation of x-ray differential phase contrast computed tomography (PCT) with respect to medical imaging. Med Phys 39:4761–4774
TS-Imaging Website (2016) http://ts-imaging.science.unimelb.edu.au/Services/Simple/ICUtilXdata.aspx (Accessed: 12 May 2019)
Myers GR, Thomas CDL, Paganin DM, Gureyev TE, Clement JG (2010) A general few-projection method for tomographic reconstruction of samples consisting of several distinct materials. Appl Phys Lett 96:021105
Van der Sluis A, van der Vorst HA (1990) SIRT-and CG-type methods for the iterative solution of sparse linear least-squares problems. Linear Algebra Appl 130:257–303
Båth M, Månsson LG (2007) Visual grading characteristics (VGC) analysis: a non-parametric rank-invariant statistical method for image quality evaluation. Br J Radiol 80:169–176
Båth M, Hansson J (2016) VGC Analyzer – a software for statistical analysis of multiple-reader multiple-case visual grading characteristics (VGC) studies. Radiat Prot Dosimetry 169:46–53
Shrout PE, Fleiss JL (1979) Intraclass correlations: uses in assessing rater reliability. Psychol Bull 86:420–428
Lai CJ, Shaw CC, Chen L et al (2007) Visibility of microcalcification in cone beam breast CT: effects of X-ray tube voltage and radiation dose. Med Phys 34:2995–3004
Rafferty EA, Durand MA, Conant EF et al (2016) Breast cancer screening using tomosynthesis and digital mammography in dense and nondense breasts. JAMA 315:1784–1786
Wienbeck S, Fischer U, Luftner-Nagel S, Lotz J, Uhlig J (2018) Contrast-enhanced cone-beam breast-CT (CBBCT): clinical performance compared to mammography and MRI. Eur Radiol 28:3731–3741
Zhao Y, Brun E, Coan P et al (2012) High-resolution, low-dose phase contrast X-ray tomography for 3D diagnosis of human breast cancers. Proc Natl Acad Sci U S A 109:18290–18294
Longo R, Tonutti M, Rigon L et al (2014) Clinical study in phase-contrast mammography: image-quality analysis. Philos Trans A Math Phys Eng Sci 372:20130025
Lewis SJ, Gureyev TE, Baran P et al (2018) Towards clinic-friendly solutions for patient trials in breast cancer phase contrast imaging. Proc. SPIE 10718, 14th International Workshop on Breast Imaging (IWBI 2018), 107181P
Tang R, Xi Y, Chai WM et al (2011) Microbubble-based synchrotron radiation phase contrast imaging: basic study and angiography applications. Phys Med Biol 56:3503–3512
Millard TP, Endrizzi M, Everdell N et al (2015) Evaluation of microbubble contrast agents for dynamic imaging with x-ray phase contrast. Sci Rep 5:12509
Wu D, Wong MD, Yang K et al (2018) Using microbubble as contrast agent for high-energy X-ray in-line phase contrast imaging: demonstration and comparison study. IEEE Trans Biomed Eng 65:1117–1123
Lång K, Arboleda C, Forte S et al (2019) Microbubbles as a contrast agent in grating interferometry mammography: an ex vivo proof-of-mechanism study. Eur Radiol Exp 3:19
Funding
This study was funded by the National Health and Medical Research Council, Australia (#APP1138283). Travel funding was also provided by the International Synchrotron Access Program (ISAP) managed by the ANSTO - Australian Synchrotron and funded by the Australian Government.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Guarantor
The scientific guarantor of this publication is Prof. Patrick Brennan.
Conflict of interest
The authors of this manuscript declare no relationships with any companies whose products or services may be related to the subject matter of the article.
Statistics and biometry
Two of the authors have significant statistical expertise.
Informed consent
Written informed consent was obtained from all subjects (patients) in this study.
Ethical approval
Institutional Review Board approval was obtained.
Methodology
• Prospective
• Experimental
• Multicentre study
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Tavakoli Taba, S., Baran, P., Nesterets, Y.I. et al. Comparison of propagation-based CT using synchrotron radiation and conventional cone-beam CT for breast imaging. Eur Radiol 30, 2740–2750 (2020). https://doi.org/10.1007/s00330-019-06567-0
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-019-06567-0