Skip to main content
Log in

Reliability of 2D and 3D ultrasound for infant hip dysplasia in the hands of novice users

  • Ultrasound
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract

Purpose

Developmental dysplasia of the hip (DDH) diagnosis by two-dimensional ultrasound (2DUS) can have poor inter-rater reliability. 3D ultrasound (3DUS) may be more reliably performed, particularly by novice users. We compared intra- and inter-rater reliability between expert and novice operators performing 2DUS and 3DUS for DDH.

Materials and methods

Infants with suspected DDH were assessed with 2DUS and 3DUS. Novice operators had 1.5 h of training and Experts had 5–15 years’ experience. Images included two 2DUS static and two 3DUS sweep images per operator. Image quality was assessed by 5-point system (yes/no: full femoral head; full acetabular roof; horizontal iliac wing; os ischium; absent motion/artifact). 2DUS indices (alpha angle, coverage) were measured centrally by a blinded reader with 2 years DDH US experience. 3DUS was post-processed by semi-automated custom software generating acetabular surface models, indices and estimated probability of DDH. Gold-standard diagnosis of each hip as normal, borderline or dysplastic was based on radiologist review of expert 2DUS.

Results

Thirty infants, mean age 10.8 weeks were enrolled. Quality scores were 2.7±1.2 Novice versus 4.9±0.3 Expert for 2DUS (p = 0.04), and 4.2±1.0 Novice versus 4.9±0.3 Expert for 3DUS (p = 0.99). Inter-rater reliability was poor for 2DUS (ICC=0.10 for alpha angle, 0.04 for acetabular coverage) and moderate to high for 3DUS (ICC=0.73-0.83 for alpha angle, 0.55 for acetabular coverage). Intra-rater reliability and diagnostic accuracy was higher for 3DUS than 2DUS.

Conclusion

Novice operators can perform 3DUS for DDH with reliability and accuracy approaching expert sonographers. Novices perform 2DUS with poor reliability and accuracy.

Key Points

• Novice/expert inter-rater reliability improved from poor with 2DUS to moderate/high with 3DUS.

• Novice operators using 3DUS correctly classified 57/58 (98%) of infant hips.

• DDH can be reliably assessed by novice operators using 3DUS.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

Abbreviations

2DUS:

Two-dimensional ultrasound

2Dα:

2D Alpha Angle

3DUS:

Three-dimensional ultrasound

ACR:

American College of Radiology

DDH:

Developmental dysplasia of the hip

ICC:

Intraclass correlation coefficients

POCUS:

Point-of-care ultrasound

References

  1. Shorter D, Hong T, Osborn DA (2013) Cochrane review: screening programmes for developmental dysplasia of the hip in newborn infants. Evid Based Child Health 8:11–54. https://doi.org/10.1002/ebch.1891

    Article  PubMed  Google Scholar 

  2. Ömeroğlu H, Köse N, Akceylan A (2015) Success of Pavlik Harness Treatment Decreases in Patients ≥ 4 Months and in Ultrasonographically Dislocated Hips in Developmental Dysplasia of the Hip. Clin Orthop Relat Res 474(5):1146–1152. https://doi.org/10.1007/s11999-015-4388-5

    Article  PubMed Central  Google Scholar 

  3. Furnes O, Lie SA, Espehaug B, Vollset SE, Engesaeter LB, Havelin LI (2001) Hip disease and the prognosis of total hip replacements. A review of 53,698 primary total hip replacements reported to the norwegian arthroplasty register 1987-99. J Bone Joint Surg Br 83(4):579–586. https://doi.org/10.1302/0301-620x.83b4.0830579

    Article  CAS  PubMed  Google Scholar 

  4. Paton RW (2017) Screening in Developmental Dysplasia of the Hip (DDH). Surgeon 15(5):290–296. https://doi.org/10.1016/j.surge.2017.05.002

    Article  PubMed  Google Scholar 

  5. Pollet V, Percy V, Prior HJ (2017) Relative Risk and Incidence for Developmental Dysplasia of the Hip. J Pediatr 181:202–207. https://doi.org/10.1016/j.jpeds.2016.10.017

    Article  PubMed  Google Scholar 

  6. Mostert AK, Tulp NJ, Castelein RM (2000) Results of pavlik harness treatment for neonatal hip dislocation as related to graf's sonographic classification. J Pediatr Orthop 20(3):306–310. https://doi.org/10.1097/01241398-200005000-00007

    Article  CAS  PubMed  Google Scholar 

  7. Orak M, Onay T, Çağırmaz T, Elibol C, Elibol F, Centel T (2015) The reliability of ultrasonography in developmental dysplasia of the hip: How reliable is it in different hands? Indian J Orthop 49(6):610. https://doi.org/10.4103/0019-5413.168753

    Article  PubMed  PubMed Central  Google Scholar 

  8. Jaremko JL, Mabee M, Swami VG, Jamieson L, Chow K, Thompson RB (2014) Potential for Change in US Diagnosis of Hip Dysplasia Solely Caused by Changes in Probe Orientation: Patterns of Alpha-angle Variation Revealed by Using Three-dimensional US. Radiology 273(3):870–878. https://doi.org/10.1148/radiol.14140451

    Article  PubMed  Google Scholar 

  9. Hünerbein M, Raschke M, Khodadadyan C, Hohenberger P, Haas NP, Schlag PM (2001) Three-dimensional ultrasonography of bone and soft tissue lesions. Eur J Ultrasound 13(1):17–23. https://doi.org/10.1016/s0929-8266(01)00117-3

    Article  PubMed  Google Scholar 

  10. American Institute of Ultrasound in Medicine (2013) AIUM practice guideline for the performance of an ultrasound examination for detection and assessment of developmental dysplasia of the hip. J Ultrasound Med 32(7), 1307–1317. https://doi.org/10.7863/ultra.32.7.1307

  11. Zonoobi D, Hareendranathan A, Mostofi E et al (2018). Developmental Hip Dysplasia Diagnosis at Three-dimensional US: A Multicenter Study. Radiology 287(3):1003–1015. https://doi.org/10.1148/radiol.2018172592

  12. Noordin S, Umer M, Hafeez K, Nawaz H (2010) Developmental dysplasia of the hip. Orthop Rev (Pavia) 2(2):e19. https://doi.org/10.4081/or.2010.e19

    Article  Google Scholar 

  13. Mahan ST, Katz JN, Kim YJ (2009) To Screen or Not to Screen? A Decision Analysis of the Utility of Screening for Developmental Dysplasia of the Hip. J Bone Joint Surg Am 91(7):1705–1719. https://doi.org/10.2106/JBJS.H.00122

    Article  PubMed  PubMed Central  Google Scholar 

  14. Patel H, Canadian Task Force on Preventive Health Care (2001) Preventive health care, 2001 update: screening and management of developmental dysplasia of the hip in newborns. CMAJ 164(12):1669–1677

    CAS  PubMed  PubMed Central  Google Scholar 

  15. Shipman SA, Helfand M, Moyer VA, Yawn BP (2006) Screening for Developmental Dysplasia of the Hip: A Systematic Literature Review for the US Preventive Services Task Force. Pediatrics, 117(3). https://doi.org/10.1542/peds.2005-1597

  16. Buerger AM, Clark KR (2017) Point-of-Care Ultrasound: A Trend in Health Care. Radiol Technol 89(2):127–138

    PubMed  Google Scholar 

  17. Waterbrook AL, Adhikari S, Stolz U, Adrion C (2013) The accuracy of point-of-care ultrasound to diagnose long bone fractures in the ED. Am J Emerg Med 31(9):1352–1356. https://doi.org/10.1016/j.ajem.2013.06.006

    Article  PubMed  Google Scholar 

  18. Sippel S, Muruganandan K, Levine A, Shah S (2011) Review article: Use of ultrasound in the developing world. Int J Emerg Med 4:72. https://doi.org/10.1186/1865-1380-4-72

    Article  PubMed  PubMed Central  Google Scholar 

  19. Papali A (2015) Providing health care in rural and remote areas: Lessons from the international space station. Bull World Health Organ 94(1):73–74. https://doi.org/10.2471/blt.15.162628

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Funding

This study has received funding by: Capital Health Endowed Chair in Diagnostic Imaging; CIHR-IHDCYH New Investigator Research Grant; RSNA Research Seed Grant; AI-HS Postdoctoral Fellowship Award; CMA Joule Innovation Grant. Dr. Jaremko is supported by Medical Imaging Consultants. Mr. Mostofi was partly supported by the M.E. Ledingham Memorial Award.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jacob L. Jaremko.

Ethics declarations

Guarantor

The scientific guarantor of this publication is Dr. Jacob Jaremko, M.D., Ph.D, FRCPC.

Conflict of interest

The authors of this manuscript declare no relationships with any companies whose products or services may be related to the subject matter of the article.

Statistics and biometry

One of the authors has significant statistical expertise.

Informed consent

Written informed consent was obtained from all subjects (patients) in this study.

Ethical approval

Institutional Review Board approval was obtained.

Methodology

• Retrospective

• Cross-sectional study

• Performed at one institution

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Mostofi, E., Chahal, B., Zonoobi, D. et al. Reliability of 2D and 3D ultrasound for infant hip dysplasia in the hands of novice users. Eur Radiol 29, 1489–1495 (2019). https://doi.org/10.1007/s00330-018-5699-1

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00330-018-5699-1

Keywords

Navigation